Ningrui Li1, Pooja Gaur2, and Kim Butts Pauly2
1Electrical Engineering, Stanford University, Stanford, CA, United States, 2Radiology, Stanford University, Stanford, CA, United States
Synopsis
MR acoustic radiation force imaging (MR-ARFI) can
be used to localize the focal spot for non-thermal transcranial ultrasound
therapies. The acoustic radiation force is proportional to the applied acoustic
intensity, meaning that tissue displacements measured with MR-ARFI can
potentially be used to estimate the acoustic intensity at the target. However, variable
brain stiffness is an obstacle to obtaining accurate acoustic intensity estimates.
Using gelatin phantoms with varying stiffnesses, we demonstrate that stiffness
information from MR elastography can be used in combination with MR-ARFI to improve
in situ estimates of acoustic intensity. This could enable safer and
more effective treatments.
Introduction
The skull presents a significant challenge to
focused ultrasound brain therapies. It attenuates the acoustic intensity at the
focal spot, and there can be up to a four-fold variation due to skull
heterogeneity across patients1. This is problematic for transcranial
therapies, such as blood-brain barrier opening, where modest increases in
acoustic intensity can cause tissue damage through inertial cavitation2.
MR acoustic radiation force imaging (MR-ARFI) methods have been developed for in vivo focal spot localization3,
and they have been applied successfully to procedures such as neuromodulation4
and blood-brain barrier opening5.
In addition to localization, MR-ARFI has the
potential to estimate the in situ acoustic
intensity6, since the acoustic radiation force is proportional to
the acoustic intensity at the focus7. However, determination of acoustic
intensity from MR-ARFI measurements of tissue displacement requires knowledge
of the tissue’s mechanical properties, which vary across patients, brain
regions8, and disease states such as Alzheimer’s9. Applying
the same ultrasound pulse to two patients: one with a thin skull (high
transmitted acoustic intensity) but a stiff brain and another with a thick
skull (low transmitted acoustic intensity) but a soft brain, might result in
the same tissue displacement even though the acoustic intensity at the focus is
different. In this study, we investigate whether stiffness information from MR
elastography (MRE) can be used to correct for this bias and improve in situ MR-ARFI-based estimates of
acoustic intensity.Methods
Phantoms were fabricated using 100-bloom gelatin
(Nitta Gelatin, Morrisville, NC) dissolved in a suspension of 50% deionized
water and 50% evaporated milk10. Phantom stiffness was adjusted by
varying gelatin concentration. To account for the influence of acoustic
properties on the applied acoustic radiation force7, the sound speed
and acoustic attenuation at 550 kHz was measured for each phantom using a
through-transmission method10 with a single-element planar
transducer and hydrophone (ONDA, Sunnyvale, CA).
MR-ARFI data were acquired at 3T (Signa
Excite, GE Healthcare, Milwaukee, WI) using a 2DFT spin-echo sequence (20 cm ×
20 cm field of view, 0.7 mm slice thickness, 256×128 acquisition matrix, 39 ms
echo time, 500 ms repetition time) with repeated bipolar gradients and 16 ms
ultrasound pulse length11. Complex phase difference images of
MR-ARFI data acquired with alternating motion encoding gradient polarities were
scaled to obtain displacement maps. Sonications were performed using applied
electric powers ranging from 50 W to 200 W at 550 kHz (ExAblate 2100, Insightec
Ltd., Haifa, Israel) with a 75 mm focal depth. Applied electric powers
corresponded to spatial-peak pulse-average acoustic intensities varying from
200 W/cm2 to 900 W/cm2 as measured in free water with a
fiber-optic hydrophone (Precision Acoustics, Dorset, UK), and they were
de-rated using acoustic attenuation and path lengths measured from each phantom
to acquire in situ acoustic intensity estimates.
MRE was performed using a single-shot,
spin-echo, echo-planar imaging pulse sequence12. Shear waves were
induced in the phantom by coupling it to a passive driver with a 60 Hz
actuation frequency, and wave images (128×128 acquisition matrix, 22 cm × 22 cm
field of view, 3.5 mm slice thickness, 2× acceleration factor, 58.6 ms echo
time, 2000 ms repetition time) were acquired with sixteen phase offsets sampled
over each period of motion. Sinusoidal motion-encoding gradients were applied
in the ±x, ±y, and ±z directions, and complex
shear modulus maps were reconstructed using a 3D direct inversion algorithm12.
The mean magnitude of the complex shear modulus at the focal spot was recorded
for each phantom.
Results
When sonicated using approximately equal
acoustic intensities, there was immense variability (2-3×) in MR-ARFI
displacements across phantoms with different gelatin concentrations (Fig. 1). Though
phantoms were found to have slightly dissimilar acoustic properties (Fig. 2),
meaning the applied acoustic radiation force differed across phantoms for the
same applied acoustic intensity, it was not enough to solely explain the
extensive discrepancy in MR-ARFI displacements (Fig. 1).
The inverse slope (W/cm2/μm) of
each MR-ARFI displacement vs. acoustic intensity line (from Fig. 1), was used
as a MR-ARFI-derived stiffness metric, or “apparent stiffness constant”. Fig. 3
shows that the phantom’s gelatin concentration was highly correlated with
stiffness estimated with both MR-ARFI and MRE. The “apparent stiffness constant” was
strongly correlated with MRE stiffness measurements (Fig. 4, top), and scaling
each phantom’s measured MR-ARFI displacements by their estimated MRE stiffness
led to superior agreement between estimated acoustic intensities across all the
phantoms (Fig. 4, bottom right).Discussion
This
simplistic correction scheme used in this study assumed that the phantoms were
purely elastic and did not account for the dampening effects of viscosity on
the measured MR-ARFI displacements13. More complex mechanical models
are necessary for better representing the viscoelastic nature of brain tissue.
Nonetheless, the results of this study emphasize the importance of
understanding the tissue’s viscoelastic properties when attempting to estimate
the acoustic intensity delivered to the treatment target. Our results
demonstrate the potential for MR-ARFI to provide both focal spot localization
and estimated acoustic intensity in combination with MRE stiffness estimates,
which would be crucial for enhancing the safety and efficacy of focused
ultrasound therapies. Future studies will focus on optimizing MR-ARFI and MRE
parameters as well as investigating mechanical models that account for
viscoelasticity.Acknowledgements
This work was supported by NIH T32 CA009695,
T32 EB009653, and NSF DGE 1656518. The authors would like to acknowledge
Richard L. Ehman and Kevin Glaser from the Mayo Clinic for providing the MR
elastography sequence and actuator device. We would also like to thank Patricia
Lan for her help acquiring MR elastography images.References
1. Vyas U., et al. Predicting variation in
subject thermal response during transcranial magnetic resonance guided focused
ultrasound surgery: Comparison in seventeen subject datasets. Medical Physics. 2016;43(9):5170-80.
2. Tung Y. et al. In vivo transcranial
cavitation threshold detection during ultrasound-induced blood–brain barrier
opening in mice. Physics in Medicine
& Biology. 2010;55(20):6141.
3. McDannold N. et al. Magnetic resonance
acoustic radiation force imaging. Medical
Physics. 2008;35(8):3748-58.
4. Gaur P., et al. In vivo MR-ARFI for
transcranial focused ultrasound in large animals. In 27th International Society for Magnetic Resonance in Medicine Annual
Meeting, Paris. 2018.
5. Magnin R., et al. Magnetic resonance-guided
motorized transcranial ultrasound system for blood-brain barrier
permeabilization along arbitrary trajectories in rodents. Journal of Therapeutic Ultrasound. 2015;3(1):22.
6. Gerstenmayer M., et al. Magnetic resonance
acoustic radiation force imaging for in vivo estimation of ultrasonic
transmission factor through rat skulls. In 16th
International Symposium on Therapeutic Ultrasound, Tel Aviv. 2016; p.
372-375.
7. Nightingale K., et al. Acoustic radiation
force impulse imaging: in vivo demonstration of clinical feasibility. Ultrasound in Medicine & Biology.
2002;28(2):227-35.
8. Arani A., et al. Measuring the effects of
aging and sex on regional brain stiffness with MR elastography in healthy older
adults. NeuroImage. 2015;111:59-64.
9. Murphy M., et al. Regional brain stiffness
changes across the Alzheimer's disease spectrum. NeuroImage: Clinical. 2016;10:283-90.
10. Farrer, A. I., et al. Characterization and
evaluation of tissue-mimicking gelatin phantoms for use with MRgFUS. Journal of Therapeutic Ultrasound, 2015 Dec;3(1):9.
11. Bitton R., et al. Toward MR-guided high
intensity focused ultrasound for presurgical localization: Focused ultrasound
lesions in cadaveric breast tissue. JMRI 2012;35:1089–1097.
12. Pepin K., et al. MR elastography analysis
of glioma stiffness and IDH1-mutation status. American Journal of Neuroradiology. 2018;39(1):31-6.
13. Bercoff J. et al. The role of viscosity in
the impulse diffraction field of elastic waves induced by the acoustic
radiation force. IEEE Transactions on
Ultrasonics, Ferroelectrics, and Frequency Control. 2004;51(11):1523-36.