Joseph Murphy-Boesch1, Kazu Yamamoto2, Ananda Kumar3, Peter L. Choyke4, and Murali Krishna Cherukuri2
1NINDS, National Institutes of Health, Bethesda, MD, United States, 2NCI, National Institutes of Health, Bethesda, MD, United States, 3Lambda-Z Technologies, Baltimore, MD, United States, 4NCI, NIH, Bethesda, MD, United States
Synopsis
Dissolution DNP with Hyperpolarized 13C
requires receive coils optimized for sensitivity and a transmit coil capable of
delivering strong B1 fields for the carbon frequency. Here, a dual-tuned, four-ring birdcage head
coil is described that can produce a transmit field of >90 microTesla and serve
as a receive coil for an MRI system with few receivers. Hyperpolarized 13C
spectra localized using 2D-CSI were obtained in 6 seconds from a human
pancreatic tumor xenograph grown in the flank of a nude mouse. SNR of 13C spectra obtained from a
tumor volume of <1 ml was suitable to measure the lactate/pyruvate ratio.
Introduction
Dissolution dynamic nuclear polarization (DNP)
techniques1 have made it possible to acquire sensitive 13C
metabolic images from prostate2, and brain3,4. The low gyromagnetic ratio of 13C
requires strong transmit B1 fields to excite 13C
spins. For head imaging, nested birdcage
designs with 8 struts per layer are often used5. The four-ring birdcage, however, utilizes a
single layer of conductors to provide superior axial B1 homogeneity at
both frequencies6. Here, a
16-leg four-ring birdcage is described that provides excellent baseline
performance for 1H imaging and 13C metabolic imaging at 3
Tesla. To provide a measure of the 13C
sensitivity that might be obtained from a tumor in human brain, localized 13C
spectra were acquired from a small (1 ml) human pancreatic
tumor MiaPaCa-2 xenograft grown on the flank of the hind leg of a nude mouse. Methods
A dual-tuned, four-ring birdcage coil6,
was designed as a transmit/receive coil for 1H imaging and 13C
spectroscopy of the head. The coil,
shown in Fig. 1a, was constructed with 16 legs, mounted concentrically inside
an RF shield, and configured to operate in quadrature mode at both
frequencies. The dimensions and ring
spacings for the coil, shown in Fig. 1b, were chosen to be proportionally
the same as for a dual tuned body coil reported earlier7. Four saddle-type inductive couplers were mounted
over outer meshes of the coil as elements of a four-port coupling network for
the 1H frequency. For the 13C
frequency, two quadrature feed cables were routed over the RF shield to matching
points on the inner legs of the coil. Traps for
the 1H frequency were inserted in series with these connections to
improve isolation. Tuning of the coil was
performed on the bench using a head-shaped phantom (Phantom Labs, Niskayuna,
NY) filled with PVP and saline (ε = 52, σ = 0.55). The RF currents and fields of the coil were
simulated at both frequencies using FEKO software (Altair Hyperworks, Troy,
MI). For the 1H field
simulations, Thevenin sources were connected across four outer mesh
capacitors. In Fig. 3, gradient echo 1H
images were acquired from the head phantom on a Philips Ingenia Elition 3T scanner, and 1H sensitivity was
compared with that of the standard head coil. To evaluate 13C sensitivity, 2-D
CSI spectra were acquired from an 18 cm spherical phantom filled with 100% ethylene
glycol and processed using R5.6.0 software on the scanner. Finally, localized hyperpolarized 13C
spectra were acquired from a pancreatic tumor xenograft grown in the flank of a
nude mouse. Bolus 96 mM [1-13C]pyruvate,
which was generated by SPINLAB (GE Healthcare, Milwaukee, Wisconsin), was intravenously injected to the animal 10
seconds after dissolution, and 2-D localized CSI spectra were acquired from the
entire mouse 30 sec after injection.Results
Tuning of the four-ring birdcage for 1H
and 13C was straight-forward, since none of the high-order modes of
the birdcage structures were overlapping. 1H port to 13C port isolations were greater than
-45 dB for both frequencies. RF field simulations
for each frequency are shown in Figs. 2b and 2d for coronal views through the
coil. In sagittal views,1H images
of the head phantom show good uniformity, with image intensity dominated by the
dielectric properties of the phantom. 1H sensitivities
measured at the center of a silicon oil phantom were 169 for the four-ring
birdcage and 173 for the standard head. For the 13C frequency,
simulations in Fig. 2b showed good coverage over the brain. Bench measurements
of a loaded coil using a calibrated loop yielded a B1 of 1.68 µT/ Watt1/2. The maximum available transmit power for 13C
was 3150 Watts, yielding a maximum B1 for the coil of 94 µT. This
was confirmed with tip angle measurements of the phantom. In Fig. 4, 2-D localized 13C
spectra acquired from the 13C phantom show baseline resolved triplets
from ethylene glycol. Spectroscopic
images in from the center and side peaks showed good uniformity, with image
intensities stronger near the coil. Nude
mice were injected with hyperpolarized [1-13C]pyruvate, and 2-D CSI
spectra were acquired in 6 seconds in a 7 x 14 array positioned over the animal
(Fig. 5). Spectra shown were obtained from
5mm x 5mm x 15mm voxels from the brain, liver, and tumor located in the hind
leg.Discussion
The four-ring birdcage provides excellent coverage
for the head with minor loss in 1H sensitivity as compared with a single-tuned
coil for an MRI system. The maximum transmit B1 field achieved for the 13C frequency was 94 µT, more
than double that of a “clam shell” coil provided with the 13C
accessory. Because the 13C accessory
provides only six receiver channels, a quadrature coil provides near
optimal sensitivity as compared with a limited array. Here, the four-ring birdcage provided enough sensitivity from a <0.5 ml tumor voxel to resolve and quantify the lactate-to-pyruvate ratio. Conclusion
The four-ring birdcage provides excellent 1H image uniformity and excellent transmit B1 capability for
hyperpolarized 13C experiments. 13C sensitivity and
coverage are adequate for metabolic imaging of the human brain. The four-ring birdcage provides near optimal performance as a quadrature coil and provides a competitive baseline with which to
compare head arrays3,4. Acknowledgements
This research was
supported (in part) by the Intramural Research Programs of the NIH, NINDS and
NCI.References
1. Ardenkjaer-Larsen
J-H, Fridlund B, Gram A, et al. Increase
in signal-to-noise ratio of >10,000 times in liquid-state NMR. PNAS 2003;100(18):10158.
2. Nelson
S, Kurhanewicz J, Vigneron D, et al. Metabolic Imaging of Patients with Prostate
Cancer Using Hyperpolarized [1-13C]Pyruvate. Sci Translational Med 2013;5(198):1.
3. Mareyam
A, Carvajal L, Xu D, et al. 31-Channel brain array for hyperpolarized 13C
imaging at 3T. Proc. ISMRM 2017;25:1225.
4. Ma J,
Hashoian R, Sun C, et al. Development of
1H/13C RF Head Coil for Hyperpolarized 13C Imaging of Human Brain. Proc. ISMRM 2019;27:568.
5. Fitzsimmons
J, Beck B, Brooker R. Double Resonant
Quadrature Birdcage. Magn Reson Med;1993;30:107-114.
6. Murphy-Boesch
J, Srinivasan R, Carvajal L, et al. Two Configurations of the Four-Ring
Birdcage Coil for 1H Imaging and 1H -Decoupled 31P
Spectroscopy of the Human Head. J Magn
Reson 1994;103B:103-114.
7. Boskamp E, Xie Z,
Taracila V, et al. A Dual-Tuned 70 cm
Whole-Body Resonator for 13C and Proton MRI/MRS at 3T. Proc. ISMRM 2018;26:1714.