Umberto Zanovello1 and Luca Zilberti1
1Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, Italy
Synopsis
The eddy
currents induced by the RF magnetic field B1
due to the presence of a conductive object, such as a metallic implant, may compromise both the RF coil transmit and receive sensitivity homogeneity; especially in the
periprosthetic areas. A further effect may be an
increase of the local SAR which can affect the overall safety of the MRI exam. In this study, the outcomes of an electromagnetic cloaking strategy based on the application of a suitable coating to elongated conductive implants are shown. The coating restores the coil sensitivities and SAR distribution to those obtained without the implant presence.
INTRODUCTION
The presence of metallic passive implants in a patient submitted to an
MRI scan may compromise both the quality of the MR image and the safety of the
exam itself. The interaction between the metallic object and the static magnetic field B0
has been recognised to be one of the most important cause for metal-induced artifacts1,2.
However, several artifacts mitigation sequences and techniques3
(e.g. SEMAC, MAVRIC, Filtered Back-Projection methods etc.) have been studied
to address the problem and some of them are already implemented in today
commercial scanners.
Another source of artifacts may be given by the eddy currents induced by the RF
magnetic field B1 due to the
conductive object presence4. These currents generate a scattered
magnetic field which can affect both the transmit sensitivity (B1+)
and the receive sensitivity (B1-) homogeneity in the
periprosthetic areas. To the best of the authors’ knowledge, only two possible
strategies can be adopted today to reduce such an issue: adiabatic pulses5
and multichannel transmission6. Whereas the former are generally
more difficult to be designed and implemented with respect to standard RF pulses
and only improve the transmit sensitivity, the latter requires a preliminary optimization,
based on the specific implant position inside the RF coil, to identify the proper
field polarization.
A further possible effect of the interaction between a conductive
implant and the RF field is represented by an increase of the local SAR which
can affect the overall safety of the MRI exam7. Whereas, even in
this case, multichannel transmission demonstrated beneficial effects8,
it still requires the preliminary study aimed to identify the optimum field
polarization.
This work shows, by means of numerical
simulations, the effects of a near-field electromagnetic cloaking aimed to
restore the magnetic and electric field (and, therefore, SAR) distributions to
those obtained without the detrimental effects of the implant9,10.METHODS
The
electromagnetic interaction between a generic birdcage coil and an elongated
metallic object placed inside a tissue-mimicking phantom has been approximately
described by means of a lumped element equivalent circuit (Fig. 1). The
capability of such a circuit to properly describe the occurring phenomena has
been tested by comparing the expectations from the circuit to the results
obtained through full-wave simulations. With the aid of the equivalent circuit,
a strategy has been identified to reduce the effects of the conductive implant
presence, namely to cover the object with a non-conductive dielectric coating
with suitable relative electric permittivity and thickness. A preliminary analysis, aimed to identify the possible relation between a
proper coating permittivity and thickness, has been performed at 128 MHz
considering a metallic cylinder. Thereafter, the study has been extended to 64
MHz and to a realistic model of a hip prosthesis stem. Both the B1+
magnitude homogeneity, evaluated over different slices by means of the its
standard deviation, the B1+ phase and the SAR
distribution have been taken into account to estimate the cloaking
effectiveness.RESULTS
As regards the investigated range of coating thickness
values (≤ 10 mm), the effectiveness
of the dielectric coating is influenced by the ratio between its thickness and
electric permittivity. The proper ratio between such parameters strongly
depends on the B1 frequency, being lower at higher frequency values.
This results, at 128 MHz, in a suitable coating relative permittivity lower
than one, requiring to design a proper metamaterial in order to realize the
coating itself.
The situation turned out to be different
focusing on a frequency equal to 64 MHz, where the coating relative
permittivity can be higher than one even considering a thin (2 mm) coating
thickness.
Figures 2 and 3 depict the magnitude and phase of the transmit sensitivity in a
coronal slice, whereas figure 4 shows the SAR distribution over the same slice.
For the sake of comparison, the empty phantom scenario is reported, together
with those relevant to the uncoated and coated prosthesis. As regards the
coated implant, a 2 mm Teflon® coating has been applied.DISCUSSION
The application of the 2 mm Teflon® coating restores
the original homogeneity of the B1+, leading to standard
deviations comparable to those obtained with the empty phantom. Analogous
beneficial effects are appreciable focusing on the B1+
phase, where the coating application strongly reduces the phase incoherencies
in the periprosthetic areas. Finally, the local SAR peaks, recorded when the
uncoated prosthesis is placed inside the phantom, are noticeably decreased
after the implant has been covered with the coating.CONCLUSION
Results highlight the beneficial outcomes obtained covering an
elongated implant, such as a hip prosthesis, with a non-conductive dielectric
coating with suitable thickness and electric permittivity. Positive effects are
observed both for B1+ and for SAR at 64 MHz employing a
2 mm coating made of an existing material such as Teflon®. Differently, a
dielectric coating with a comparable thickness should have a relative
permittivity lower than one at 128 MHz. In the last case, it should be realized
by means of a properly designed metamaterial.Acknowledgements
No acknowledgement found.References
- Hargreaves,
B. A. et al. Metal-Induced Artifacts
in MRI. Am J Roentgenol.2011;197:547-555.
- Schenck,
J. F. The role of magnetic susceptibility in magnetic resonance imaging: MRI
magnetic compatibility of the first and second kinds. Med Phys.1996;23:815-850.
- Koch,
K. M. et al. Magnetic Resonance
Imaging Near Metal Implants. J Magn Reson
Im.2010;32:773-787.
- Camacho C. R., Plewes, D. B. & Henkelman, R. M.
Nonsusceptibility Artifacts Due to Metallic Objects in Mr-Imaging. J Magn Reson Im.1995;5:75-88.
- Tannus A., Garwood M. Adiabatic pulses. NMR in Biomed.1997;10:423-434.
- T.J. Bachschmidt et al. Polarized multichannel transmit MRI to
reduce shading near metal implants. In: Magnetic
Resonance in Medicine.2016;75(1):217-226.
- Henry, S. Ho. Safety of metallic implants in magnetic resonance
imaging. J Magn Reson Im.2001;14:472-477.
- Nadja, A. 3-T MRI implant safety: heat induction with new
dual-channel radiofrequency transmission technology. European Radiology Experimental.2018;2:7.
- Zanovello, U. et al. An
ideal dielectric coat to avoid prosthesis RF-artefacts in Magnetic Resonance
Imaging. Sci Rep.2017;7:326.
- Zanovello, U. et al. A Near Field Cloaking Study to Reduce MRI
RF-Artefacts in Presence of elongated Prostheses. IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and
Biology.2018.