Marine A.C. Moussu1,2, Luisa Ciobanu3, Elodie Georget2, Stanislav Glybovski4, Andrew G. Webb5, Stefan Enoch1, and Redha Abdeddaim1
1Institut Fresnel, Aix Marseille Université, Marseille, France, 2Multiwave Imaging, Marseille, France, 3DRF/Joliot/Neurospin/UNIRS, Gif-sur-Yvette, France, 4ITMO University, Saint-Petersbourg, Russian Federation, 5Leiden Univ. Medical Center, Leiden, Netherlands
Synopsis
We
propose a theoretical model predicting the Signal-to-Noise Ratio (SNR) of
dielectric antennas for Magnetic Resonance Imaging. The SNR is estimated with a
semi-analytical approach describing the first transverse electric mode of a
high-permittivity ring resonator. When the ring is properly excited with a feeding
loop, it induces a strong magnetic field in the sample while dielectric losses
are limited due to low electric field intensity in this region. The proposed
model is validated with numerical simulations regarding microscopy applications
and its prediction confirmed experimentally at 17.2 T.
Introduction
Magnetic Resonance Microscopy (MRM) aims at
imaging small samples with very high spatial resolution (<< 100 μm) using B0 field
strengths between 7 T and 22 T1. At such fields, the conventional RF
coil is the solenoid. However, for lossy biological samples, its ultimate
Signal-to-Noise Ratio (SNR) is intrinsically limited by the significant
electric field within the sample2. Dielectric resonators are an alternative
probe design to tackle this issue3,4: high-permittivity rings
support eigenmodes whose frequency characteristics rely on the resonator
dimensions and the material properties5,6. The first transverse
electric mode (TE01δ) produces a strong B1
excitation field (Fig. 1), and a low E-field. Here we present a semi-analytical
model (SAM) to design dielectric ring probes with high SNR7. This
model is validated with numerical simulations and experimentally tested for MRM
at 17.2 T8.Methods
The noise is dominated by losses in the
dielectric ring (volume $$$V_\mathbf{ring}$$$, outer and inner radii $$$r_\mathbf{d}$$$ and $$$r_\mathbf{h}$$$, height $$$L$$$, permittivity $$$\epsilon_\mathbf{r}$$$) and in the sample (volume
$$$V_\mathbf{samp}$$$, radius $$$r_\mathbf{h}$$$, height $$$L$$$, electrical conductivity $$$\sigma_\mathbf{samp}$$$). The SNR is proportional to the SNR factor
(Eq. 1) with $$$H_\mathbf{samp}(0)$$$ the H-field
amplitude in the sample,
and (Eq. 2)9
the total dielectric
losses.
(Eq. 1) $$u_\mathbf{SNR}=H_\mathbf{samp}(0)/\sqrt{P_\mathbf{loss}}$$
(Eq. 2) $$P_\mathbf{loss}=\frac{1}{2}\omega\epsilon''\int_{V_\mathbf{ring}}|\mathrm{E}|^2dv+\frac{1}{2}\sigma_\mathbf{samp}\int_{V_\mathbf{samp}}|\mathrm{E}|^2dv$$
The TE01δ mode field is described by Bessel
and Hankel functions in cylindrical coordinates $$$\left(\rho,\theta,y\right)$$$. As this mode has a cylindrical
symmetry, $$$H_y$$$, $$$H_\rho$$$ and $$$E_\theta$$$ are the only
nonzero components10. Expressing $$$H_y$$$, as in Eqs. 3-4 with $$$\alpha_\Omega$$$ and $$$\beta_\Omega$$$ the wavenumbers
in region $$$\Omega$$$, enables to derive the other components11,12.
(Eq. 3)$$H_\mathbf{y,samp}\left(\rho,y\right)\propto J_\mathbf{0}\left(\alpha_\mathbf{samp}\rho\right)\cos\left(\beta_\mathbf{samp}y\right)$$
(Eq. 4)$$H_\mathbf{y,ring}\left(\rho,y\right)\propto \left[H_\mathbf{0}^\mathbf{(1)}\left(\alpha_\mathbf{ring}\rho\right)+\xi H_\mathbf{0}^\mathbf{(2)}\left(\alpha_\mathbf{ring}\rho\right)\right]\cos\left(\beta_\mathbf{ring}y\right)$$
The
E-field, derived from the spatial derivative of $$$H_\mathbf{y}$$$,
is proportional to the H-field amplitude. In Eq. 3, this amplitude is $$$H_\mathbf{samp}(0)$$$. To analytically express dielectric losses, the
probe EM field is simplified to that of the corresponding disk: $$$\mathrm{E}\left(\overrightarrow{r}\right)_{|ring}=\mathrm{E}\left(\overrightarrow{r}\right)_{|disk}$$$, $$$\mathrm{E}\left(\overrightarrow{r}\right)_{|samp}=\tau\mathrm{E}\left(\overrightarrow{r}\right)_{|disk}$$$ and $$$\mathrm{E}\left(\overrightarrow{r}\right)_{|disk}\propto J_\mathbf{1}\left(\alpha_\mathbf{disk}\rho\right)$$$. $$$\tau$$$ reflects the
amplitude decrease from permittivity contrast between the sample and the ring. Eq. 1 becomes Eq. 5 with $$$P_\mathbf{loss,norm}^\mathbf{i}=P_\mathbf{loss}^\mathbf{i}/|H_\mathbf{disk}(0)|^2$$$.
(Eq. 5) $$u_\mathbf{SNR}^\mathbf{DR}=\tau/\sqrt{P_\mathbf{loss,norm}^\mathbf{ring}+\tau^2 P_\mathbf{loss,norm}^\mathbf{samp}}$$
A Frequency Domain
Solver (CST Microwave Studio) was used to evaluate the field distribution of the mode excited in the
dielectric probe by a 1 cm diameter non-resonant loop. In a “combined method”,
the SNR factor is computed with Eq. 5 applied to the field exported from
numerical simulations run in the CST Eigenmode Solver (mesh step: 0.2 mm isotropic).
The
experimental resonator was built with a ferroelectric ceramic8 with
the following parameters at 730 MHz (Larmor frequency at 17.2 T): $$$r_\mathbf{d}$$$ = 9 mm, $$$r_\mathbf{h}$$$ = 2.8 mm, $$$L$$$ = 10 mm, $$$\epsilon_\mathbf{r}$$$ = 536, $$$\tan\delta$$$ = $$$8\times 10^{-4}$$$. The phantom used was a
tube of diameter 4.5 mm and length 12 mm filled with a solution of permittivity
$$$\epsilon_\mathbf{r,test}$$$ = 50 and conductivity $$$\sigma_\mathbf{test}$$$ = 1 S/m. The corresponding optimal solenoid has 4 turns,
7 mm diameter, 12 mm length and is made with 1.5 diameter copper wire2.
Its SNR factor can
be found in the literature2. MRI acquisitions were performed on a
preclinical device (Bruker BioSpin, Ettlingen, Germany) equipped with a
triaxial gradient system of maximum strength 1 T/m. The biological sample studied
was a chemically fixed rat spinal cord. MR acquisitions were performed with
sequences described in Fig. 2.Results
The
SAM was used to estimate the SNR gain $$$u_\mathbf{SNR}^\mathbf{DR}/u_\mathbf{SNR}^\mathbf{sol}$$$,
plotted in Fig. 3 for the phantom as a function of the dielectric properties.
For each permittivity, $$$r_\mathbf{d}$$$ was adapted to adjust the TE01δ mode to the Larmor frequency. When $$$\tan\delta \leq 2 \times 10^{-3}$$$ and $$$\epsilon_\mathbf{r}\geq 220$$$,
the semi-analytical method predicts a higher SNR with the dielectric probe than
with the optimal solenoid. From this dependency, the prototype properties were
selected, resulting in a theoretical SNR gain of 2.5. The prototype SNR factor was computed using the
SAM and compared to numerical simulations and a combined method as shown in
Fig. 4. The sample permittivities were 50 and 81 and its conductivity was
varied from 0 to 2.5 S/m. The relative error between the SAM and the commercial
software was always below 8%7. Finally,
experiments
performed on the test solution resulted in a measured SNR gain of 2.2 (versus
2.5 theoretically), as obtained with numerical simulations. Fig. 5 demonstrates
that the ceramic probe performs better than the optimal solenoid when imaging
the biological sample, with an SNR gain of 1.45.Discussion
The SAM estimates the SNR factor with an error
lower than 8% over a large range of electromagnetic properties. The SNR gain
estimation was validated in one case, in which the predicted value was within
15% of the experimentally measured value.Conclusion
This
study focused on the modelling of dielectric ring probes in which the first TE
mode is excited to generate the B1 field used in MR acquisitions.
The semi-analytical model predicts the performance of such probes with a good
accuracy and can be used as a design guideline. Our current and future work
includes the extension of dielectric probes modelling to clinical applications.
This includes resonators with human scale dimensions, as well as the study of
higher resonant modes (like HEM modes).Acknowledgements
This project received funding from the European
Union’s Horizon 2020 Research and Innovation programme (grant No 736937) and
from an ERC Advanced Grant (NOMA-MRI 670629). S.G.
acknowledges the support by the President of the Russian Federation
(МК-3620.2019.8).References
1. Ciobanu, L. (2017). Microscopic Magnetic
Resonance Imaging: A Practical Perspective. Pan Stanford.
2. Minard, K. R., & Wind, R. A. (2001).
Solenoidal microcoil design—Part II: Optimizing winding parameters for maximum
signal‐to‐noise performance. Concepts in Magnetic Resonance, 13(3), 190-210.
3. Haines, K., Neuberger, T., Lanagan, M.,
Semouchkina, E., & Webb, A. G. (2009). High Q calcium titanate cylindrical
dielectric resonators for magnetic resonance microimaging. Journal of Magnetic
Resonance, 200(2), 349-353.
4. Neuberger, T., Tyagi, V., Semouchkina, E.,
Lanagan, M., Baker, A., Haines, K., & Webb, A. G. (2008). Design of a
ceramic dielectric resonator for NMR microimaging at 14.1 tesla. Concepts in
Magnetic Resonance Part B: Magnetic Resonance Engineering: An Educational
Journal, 33(2), 109-114.
5. Mongia, R. K., & Bhartia, P. (1994).
Dielectric resonator antennas—A review and general design relations for
resonant frequency and bandwidth. International Journal of Microwave and
Millimeter‐Wave Computer‐Aided Engineering, 4(3), 230-247.
6. Kajfez, D., & Guillon, P. (1998).
Dielectric resonators (p. 547). Atlanta: Noble Publishing Corporation.
7. Moussu, M. A.C., Abdeddaim, R., Dubois, M.,
Georget, E., Webb, A. G., Nenasheva, E., Belov, P., Glybovski, S., Ciobanu, L.,
& Enoch, S. (2019). A Semi-Analytical Model Of High Permittivity Dielectric
Ring Resonators for Magnetic Resonance Imaging. Under review.
8. Moussu, M. A., Ciobanu, L., Kurdjumov, S.,
Nenasheva, E., Djemai, B., Dubois, M., ... & Glybovski, S. (2019).
Systematic Analysis of the Improvements in Magnetic Resonance Microscopy with
Ferroelectric Composite Ceramics. Advanced Materials, 1900912.
9. Petit, R. (1989). Ondes
électromagnétiques en radioélectricité et en optique (Vol. 2). Paris: Masson.
10. Pozar, D. M. (2009). Microwave engineering.
John Wiley & Sons.
11. Zhang, K., Li, D., Chang, K.,
Zhang, K., & Li, D. (1998). Electromagnetic
theory for microwaves and optoelectronics (pp. 274-283). Berlin: Springer.
12. Sheen, J. (2008). A dielectric resonator
method of measuring dielectric properties of low loss materials in the
microwave region. Measurement Science and Technology, 19(5), 055701.