Benjamin M Hardy1,2, Yurui Gao2,3, and Adam W Anderson2,3
1Department of Physics and Astronomy, Vanderbilt University, Nashville, TN, United States, 2Vanderbilt University Institute of Imaging Science, Nashville, TN, United States, 3Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, United States
Synopsis
Magnetic
and electric field values are simulated with up to 480 Radio Frequency
surface loops 1 cm in radius surrounding
realistic
head models. The work investigates the capability of B1+ shimming alone with
increased degrees of freedom. Transmit
efficiency,
power limitations, and target homogeneity across the brain volume inform design
choices for the array. Evidence of the
effects
of subject to subject variation can be seen in the optimized shim weights magnitude
and phase variation for a choice 256-element
geometry optimized over 20 unique head models oriented at 3 head angles.
Purpose
Conventional
RF coil designs fail to produce uniform excitation across the brain at 7 Tesla due
to the short wavelength of the B1+ field. Furthermore, coil
designs must account for field fluctuations due to subject to subject variation. Hardware adjustments
that allow the shaping and control of Radio Frequency (RF) waves are therefore needed to enable robust
imaging at 7 Tesla. RF shimming relies on modulating the amplitude and phase of
each coil element in order to homogenize excitation over a slice or volume.
Although previous work has shown RF shimming to be a promising technique for
homogeneous excitation 1–3, the efficacy of increased
degrees of freedom, or more coil elements, in RF shimming is still an open
question. With multi-channel systems and array power networks emerging in the
field, it is important to establish how RF shimming alone performs with
increased degrees of freedom over a range of realistic head models. Methods
Initially,
simulations using the finite difference time domain method were used to
calculate electric and magnetic fields across the male National Library of Medicine (NLM) Visible Human Project® head model with 1 mm 3 isotropic spatial resolution and 29 tissue compartments 4. Using a
transmission coil design of 1 cm radius copper loops placed on the surface of a 26.4 cm diameter cylinder, 480 (12 x 40 grid on the cylinder surface) individual
loop elements’ field distributions were calculated across the head model. This
grid of 480 was further sub-sampled row and column-wise into less dense arrays (see Fig. 1). The excitation profile
was optimized over phase and amplitude of each coil element with intent to
minimize the coefficient of variation of the B1+ magnitude across the brain’s volume while limiting power requirements. Starting
at zero-target-phase across the volume, Tikhonov regularized optimization was
performed iteratively, updating the target-phase pattern with the resulting
phase generated by the shim weights every iteration until the cost function
changed negligibly. The optimization simultaneously limits power and overcomes
the non-convex nature of the optimization problem 5. The global
specific absorption rate (SAR), local 10-gram average SAR, and transmit
efficiency values were then calculated using the optimized shim field values. A
separate geometry of 256-loops (16 x 16 staggered grid) centered on the
brain region returned feasible excitation homogeneity under reasonable power
constraints. This coil geometry’s shim weights were then optimized over 20 unique,
7 compartment, head models, at 3 (head angle) configurations within the coil to test
performance against subject variation.Results
Considering
a realistic power limitation on current amplitudes to each coil element (~12-20 amps), an array of 240-320 1 cm radius surface loops provides homogeneous
excitation across the NLM Visible Human Project® brain
volume with roughly 5% variation. A separate geometry of 256 elements centered
around the brain performed best with a maximum local SAR of .0132 W/kg and
global SAR of 8.6 µW/kg for a 90-degree, 0.1 µT target field (for a 1 µT hard
pulse local and global SAR would be 1.32 W/kg and global SAR of .86 mW/kg). Transmit
efficiency for the array ranged from 0.17 to 0.35 µT2/Watt. Coil geometries
with columns of loops stacked along z, had roughly a 0.2 µT2/Watt increase in
transmit efficiency in comparison to loops staggered on the surface of the
cylinder or limited to rows in the x-y plane (see Fig. 2). Over 60 unique
head configurations, the optimized shim weights confirm that the B1+ phase changes more drastically than the magnitudes in order to account for subject variability (see Fig. 3). Discussion
These
simulations are intended to direct the construction of a multi-surface-loop
coil array with RF shim capabilities. The coil array will be driven
by the Philips Multix8 Radio Frequency Power Amplifier (RFPA) system. A
distribution network utilizing micro-strip Wilkinson power splitters will be
combined with a butler matrix array connecting the coil array to the RFPAs. The
butler matrix array will potentially simplify the number of channels required
to achieve homogeneous excitation. Vector modulators will be used in order to
control the amplitude and phase of each coil.Conclusion
RF
shimming combined with array networks to numerous coil elements may facilitate
use of ultra-high field systems both for research and clinical applications.
Subject to subject variation may be accounted for given the correct shim
optimization. Theoretically, a 240 to 320 surface loop array provides enough
degrees of freedom to tailor the field given appropriate decoupling and
distribution networks.Acknowledgements
The authors would
like to acknowledge Bennett Landman and Benoit Dawant for preparing and providing
the registered head model data along with Will Grissom and Xinqiang Yan for
helpful discussions concerning coil design and shim optimization techniques.
Research reported in this publication was supported by the National Institute
of Health under award numbers R21 EB024311 and R01 NS095291.References
[1] Mao et al., "Exploring the limits of RF shimming for
high-field MRI of the human head". Magnetic Resonance in Medicine; 56:918–922, 2006
[2] Hoult et al., "Sensitivity and power deposition in a high-field imaging
experiment". Journal of Magnetic Resonance Imaging; 12:46–67, 2000
[3] Ibrahim et al., "Effect of RF coil excitation on field inhomogeneity at ultra
high fields: A field optimized TEM resonator". Magnetic Resonance Imaging; 19:1339–1347, 2001
[4] Courtesy
of the U.S. National Library of Medicine
[5] Grissom et al., "Small-tip-angle spokes pulse design using interleaved greedy and local
optimization methods". Magnetic Resonance in Medicine; 68(5): 1553–1562, 2012