Ming Lu1,2,3, William A. Grissom1,2,4, John C. Gore1,2,4, and Xinqiang Yan1,2
1Vanderbilt University Institute of Imaging Science, Nashville, TN, United States, 2Department of Radiology, Vanderbilt University Medical Center, Nashville, TN, United States, 3College of nuclear equipment and nuclear engineering, Yantai University, Yantai, China, 4Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, United States
Synopsis
Self-decoupled coils (SDCs) were recently described which can solve complex coupling issues in dense arrays. SDCs have a simple structure and represent a general approach. However, dipole mode in SDCs make it sensitive to the load. In this work, we introduced a differential mode to the design of SDCs to alter the necessary value of Cmode and a coils’ robustness to loading. This differential-mode SDC can be seen as a combination of differential-mode decoupling with self-decoupling. We found that suitable Cmode can choose to increase coil robustness as well as avoid added coil loss.
Purpose
Self-decoupled
coils (SDCs) were recently described which can solve the complex coupling
issues that arise in a dense array, and which re-distribute impedances to precisely
match electric couplings to cancel magnetic couplings [1]. SDCs have a simple
structure and represent a general approach applicable to both transmit and receive
arrays, as well as mixed arrays that combine loop coils with other elements
such as dipoles. However, the dipole mode in SDCs makes it sensitive to the
load. In this work, we introduce a differential-mode of operation for SDCs by
adding a passive loop. The passive loop makes the dipole mode in SDCs become micrstripline
mode. It also reduces the coil's self-inductance and thus increases the value
of Cmode.Theory
Figures
1a-d illustrate differential-mode SDCs using different types of passive closed
loops: Figure 1a uses circumferential shielding, Figure 1b uses a stacked
passive loop, and Figure 1c uses an inner passive loop. The differential mode
is generated by the inductive coupling between the SDC coil and the passive
loop, and thus the contribution of the differential mode depends on the distance
between the coil and the passive loop. It also should be noted that the
differential mode decreases the coil efficiency because it constrains the
magnetic field between the coil and the passive loops, so less magnetic flux goes
outside to the sample. However the differential mode is beneficial because it
reduces the self-inductance of the SDC, and thus increases the capacitance of
coil (mainly Cmode). There is therefore a tradeoff between coil efficiency and
the value of Cmode when choosing the distance between the SDC and the passive
loop.Methods
The three
differential-mode SDC types were modeled and simulated using Ansys software.
The coil diameter was set to 6cm and operated at 298MHz (7T). The
coil-to-passive loop distance was varied from a fraction of a mm to several mm:
0.25/0.5/1/2/3 mm for the inner passive loop case; 0.25/0.48/0.95/1.3/2.4 mm
for the stacked passive loop case; 0.205/0.305/0.605/1.405/2.155 mm for the
inner passive loop case, as shown in Figures 1d-f. For all cases, we simulated
the scattering parameter S21 between two closely spaced side-by-side loops with
various Cmode values (with coils well matched). A cuboid phantom (σ=0.6 S/m,εr=78,
40x15x15cm3) was placed 1.5cm below the coils as the load. Because
the coil-to-passive loop’s distance changes the coil efficiency, we also simulated
the B1 field of a single SDC with the optimal Cmode that ensures the best coil
isolation.
To evaluate the possible B1 efficiency decrease, we built a
6-cm-diameter non-optimized coil using semi-rigid coaxial cable (EZ86/CU/M17,
Huber+Suhnerouter). The outer conductor was used as the coil element, and the inner
conductor was the passive loop (they were 0.6 mm apart). For comparison, we
also built a conventional SDC of the same size. GRE images were acquired of an ex-vivo
squirrel monkey brain phantom using a
9.4T Varian system.Results
Figure
2 shows the values of the two coils’ transmission coefficients (S21) versus
Cmode. When the coil and passive loop are closely spaced, the differential mode
dominates and the EM fields are localized between them and do not go outside.
Therefore, they are highly isolated for different values of Cmode. For
instance, when the circumferential shielding is 1mm above the coils, their
isolation can always achieve <-15 dB (Figure 2a). The localized fields and
large Cmode increase the coil’s robustness versus loading. As shown in Figure 3,
the maximum frequency shift of a coil with a stacked loop 0.25-mm apart is
4.6MHz, while that of an original SDC is up to 17.3MHz. However, the coil’s
better robustness comes with a cost of lower efficiency. As shown in Figure 4b,
its B1 efficiency is only 49% compared to an original SDC. As the passive loops
are moved further away from the coil, the differential mode contributes less
and the coil performance becomes similar to that of an original SDC. Similar to
simulation results in Figure 3, the SNR from the non-optimized differential
mode SDC is only 56% compared to that of an original SDC.Discussion and Conclusion
We
introduced a differential mode to the design of SDCs to alter the necessary value
of Cmode and a coils’ robustness to loading. This differential-mode SDC can be
seen as a combination of differential-mode decoupling with self-decoupling.
Note that the differential-mode itself can greatly improve the coil isolation
with no need to select Cmode [2,3]. It induces considerable coil losses and
decreases the SNR, as shown in previous works [2] and also validated here
(Figures 4 and 5). By combing the differential mode with the SDC, there is a
tradeoff between coil losses and Cmode values. For a 6-cm-diameter coil at 7T
placed 1.5cm apart from a tissue-equivalent phantom, the B1 efficiency can be
maintained with the coil-to-passive loop's distance >1.5mm. We note that this
distance varies with different loading cases. Generally, it can be smaller with
heavier loading but has to be larger in lighter loading cases. The dielectric
between the coil and passive loop provides another degree of freedom (its
relative permittivity) to manipulate the resonance frequency, and so the Xarm can
be reduced for small self-decoupled coils at lower fields.Acknowledgements
The authors thank Dr. Feng Wang of Vanderbilt University
for acquiring the in vivo images and Dr. Andrew Webb of Leiden University for
helpful discussions on using the differential mode to improve the isolation of
coaxial cable coils.
This work was partially supported by NIH Grant R01 EB 016695.
References
1. Yan,
X., Gore, J. C., & Grissom, W. A. (2018). Self-decoupled radiofrequency
coils for magnetic resonance imaging. Nature communications, 9(1), 3481.
2. Yeh,
J. N. T., Lin, J. F. L., Li, Y. T., & Lin, F. H. (2018). A Flexible and
Modular Receiver Coil Array for Magnetic Resonance Imaging. IEEE transactions
on medical imaging, 38(3), 824-833.
3. Ruytenberg, T., Webb,
A., & Zivkovic, I. (2019). Shielded‐coaxial‐cable coils as receive and transceive array
elements for 7T human MRI. Magnetic resonance in medicine.