Ali Caglar Özen1,2, Johannes Fischer1, Hao Song1, Yanis Taege1, Christian Schuch3, Rianne Schutter4, Cyril Moers4, Ronald JH Borra5, and Michael Bock1
1Dept. of Radiology, Medical Physics, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, 2German Consortium for Translational Cancer Research Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg, Germany, 3NUKEM Isotopes GmbH, Alzenau, Germany, 4Department of Surgery – Organ Donation and Transplantation, University Medical Center Groningen, Groningen, Netherlands, 5Medical Imaging Center, University Medical Center Groningen, Groningen, Netherlands
Synopsis
Direct 17O-MRI can be used to measure
renal metabolism in perfused kidneys in an organ transplantation setup. To
optimize SNR, a dedicated 17O Rx array was designed that fits into a
perfusion box used for functional metabolism tests of the donor kidneys. The
increased filling ratio resulted in higher SNR compared to the volume and surface
Tx/Rx coils. In combination with a 17O birdcage Tx coil for
homogeneous excitation the 4-element Rx array could also be used for parallel
imaging.
Introduction
Most organ
transplants are renal transplants [1–3], but even though biomarkers
exist to diagnose complications after kidney transplantation [4],
survival rates decrease drastically with the time after transplantation [5,6].
One reason for post-transplantation complications is the insufficient
functional characterization of the transplanted kidney. We have shown
previously that, in addition to perfusion and renal filtration, tissue
oxygenation of a resected kidney can be quantified in vitro before
implantation using dynamic 17O MRI [7].
A
homogeneous excitation field (B1) and high SNR are important for high-resolution
17O MRI of the kidney, which in whole body 1H MRI is
achieved by combining a local Rx array with a large volume Tx coil. In this
work, we designed and constructed a 17O RF coil array that fits on a
sterilized perfusion box used for functional metabolism tests of the donor
kidneys which was combined with a 17O birdcage head coil for
homogeneous RF excitation.Methods
Rx
Array
A schematic of the coil arrangement is shown in
Figure 1A. Three of the 4 Rx elements (C1-C3) were mounted on a 3D-printed
holder (Fig.1B) that fits under the perfusion box in which the kidney is stored
between resection and transplantation. C1-C3 were constructed using two turns
of a tin-plated copper wire (∅=1mm)
on a rectangular shaped former of 3x5cm2 size. The tuning capacity
was distributed on two elements, one of which was used as the input port for
tuning and matching (Fig.1C). A larger (∅=15cm)
Rx element (C4) was placed on top of the perfusion box using a single turn (∅=5mm)
of copper tube (Fig.1D). C2-4 were geometrically decoupled, and all elements
were further isolated from each other by preamplifier decoupling [8]. A
separate custom-made quadrature-driven 4-leg birdcage volume coil was used as the Tx only element. Measurement of the reflection and transmission coefficients, Sij, were performed on a vector network
analyzer (ZVB 4; Rohde&Schwarz, Munich Germany) for each channel pair with all
other channels terminated with 50Ω. Noise correlation matrix of the Rx
array was also measured using a FID sequence (TR/TE 200 ms/0.56 ms, BW 10 kHz,
Peak Tx RF voltage 0V) [9].
MRI
Measurements
A
cylindrical water phantom composed of three separate compartments was used to
test performance of the coil array at a clinical 3T MR system (Prisma; SIEMENS,
Erlangen, Germany). The inner cylinder (modeling the medulla) was filled with 17O2-enriched
saline solution (c17O2 = 0.072%) and has an inner diameter
of 3 cm and a length of 4 cm. In total, volume (V = 247mL) and axial
dimensions are similar to those of a human kidney [10].
For
image acquisition a radial 3D UTE sequence with golden-angle projection
acquisition pattern was used [11]. Kaiser-Bessel-regridding [12] of
k-space data and Hanning-filtering was subsequently applied.
For comparison, a custom built Tx/Rx head coil, and a loop coil were also used
to acquire 17O MRI data using the same protocol (TR/TE 6ms/0.52ms,
FA π/6, BW 390 Hz/px, radial spokes 80000). The SNR of a measurement was
quantified according to [13].Results
S-parameter
and noise correlation matrices are shown Figure 2. The range of coupling
between overlapping/non neighboring elements was 13.5-21.3 dB/ 13.5-18.3 dB, to
which preamplifier decoupling (75cm-long coaxial line and lumped element 3l/16 tank circuits between coil and
preamplifier interface) added another 13 dB. An average unloaded/loaded quality
factor of QU = 150.3 / QL = 80.2 was measured
resulting in a Q-ratio of 1.87. The active detuning efficiency was measured as
38+-3 dB for all elements. Noise correlation values ranged from 4% to 10%
(Fig.2).
The
reconstructed images of the kidney phantom are shown in Figure 3 as cross-sections
from different axes. The SNR values of the center/left/right section of the
phantom are 98/66/81 for single loop coil, 97/65/70 for the birdcage coil, and 481/425/404 for the 17O Rx
array, respectively, which is an up to 5-fold increase in SNR using the Rx array.Discussion
As expected, the SNR of the central cylinder was
higher for all coils due to the addition of H217O. The
performance of the Rx array was superior to single channel Tx/Rx loop and
birdcage coils due to the smaller coil element size, and the higher filling ratio.
Single loop coils, however, are not preferable for quantitative imaging tasks
due to an SNR difference of 19% between left and right cylinders, in the SNR
comparison. C4 is placed on top of the perfusion box, therefore has higher
coupling with C1-3. Since there is a distance of at least 8 cm between the
center of the kidney and the loop coil, fields of a smaller loop coil would not
be able to penetrate deep enough to reach the kidney. In a next step the
increased image quality of the coil array will be used to increase the spatial
and temporal resolution in dynamic 17O MRI measurements in kidneys to
identify e.g. induced ischemic lesions.Acknowledgements
Support
from NUKEM Isotopes Imaging GmbH is gratefully acknowledged.References
1. Branger, P. & Samuel, U. Annual Report 2017. 124 (Eurotransplant International Foundation, 2017).
2. Hart, A. et al. OPTN/SRTR 2016 Annual Data Report: Kidney. Am. J. Transplant. 18, 97.
3. Department of Health and Human Services, Health Resources and Services Administration, Healthcare Systems Bureau, Division of Transplantation. Annual Report of theU.S. Organ Procurement and Transplantation Network and the Scientific Registry of Transplant Recipients: UNOS (2016). Available at: https://unos.org/about/annual-report/2016-annual-report/. (Accessed: 30th October 2018)
4. Salvadori, M. & Tsalouchos, A. Biomarkers in renal transplantation: An updated review. World J. Transplant. 7, 161–178 (2017).
5. Pestana, J. M. Clinical outcomes of 11,436 kidney transplants performed in a single center - Hospital do Rim. J. Bras. Nefrol. 39, (2017).
6. Wang, J. H., Skeans, M. A. & Israni, A. K. Current Status of Kidney Transplant Outcomes: Dying to Survive. Adv. Chronic Kidney Dis. 23, 281–286 (2016).
7. Taege, Y., Fischer, J., Özen, A. C., et al. Protocol Optimization for Functional 17O-MRI of Donor Kidneys at 3T. In Proc. 27th ISMRM, Montreal, QC, Canada. p. 4223.
8. Roemer, P.B., Edelstein, W.A., Hayes, C.E., Souza, S.P., Mueller, O.M. The NMR phased array, Magnetic Resonance inMedicine 16, 192–225 (1990).
9. Kellman, P., & McVeigh, E. R. Image reconstruction in SNR units: a
general method for SNR measurement. Magnetic resonance in medicine,
54(6), 1439–1447 (2005).
10. Cheong, B. et al. Normal values for renal length and volume as measured by magneticresonance imaging. Clinical Journal of the American Society of Nephrology 2.1 (2007).
11. Chan, R. W., Ramsay, E. A., Cunningham, C. H. & Plewes, D. B. Temporal stability of adaptive 3D radial MRI using multidimensional golden means. Magn. Reson. Med.61, 354–363 (2009).
12. Jackson, J. I., Meyer, C. H., Nishimura, D. G. & Macovski, A. Selection of a convolution function for Fourier inversion using gridding (computerised tomography application).IEEE Trans. Med. Imaging 10, 473–478 (1991).
13. Gudbjartsson, H. H. and Patz, S. The Rician
Distribution of Noisy MRI Data. Magnetic Resonance in Medicine 34.6 (1995).