Neha Koonjoo1,2,3, Sheng Shen1,4, Charlotte R Sappo5, and Matthew S Rosen1,2,3
1Department of Radiology, A.A Martinos Center for Biomedical Imaging/MGH, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Department of Physics, Harvard University, Cambridge, MA, United States, 4State Key laboratory for Power Transmission Equipment and System Security and New Technology, Chongqing University, Chongqing, China, 5Institute of Imaging Science, Vanderbilt University, Nashville, TN, United States
Synopsis
High performance RF coil design is challenging at
ultra-low field as rules of thumb learned from high-field may not apply. Previously,
we built a 2-channel quadrature head coil with orthogonal B1 fields
using an air-core transformer to null the mutual coupling between the coil,
however this was unsuccessful in boosting the SNR beyond that of our
single-channel spiral volume coil for brain imaging at 276 kHz. Here, a new
optimized quadrature coil with a capacitive decoupling element and a new outer
layer coil was designed. Images acquired show an expected √2-factor or more enhancement in combined maximum SNR.
Introduction
In the ultra-low
field regime, the noise floor in RF coils is dominated by Johnson noise. High-performance
RF coil design at 6.5mT presents unique engineering challenges which sometimes
are not well advised by work done at high-field. Our previous coils focused on a one-channel spiral head coil1 with maximum coverage of the volume of
interest and a 2-channel quadrature head coil2 which had the goal of
increasing the SNR by √2 3. So far, our 1-channel coil has enabled
3D proton-density weighted brain images of 2.5 × 3.5 × 8.5 mm3
spatial resolution, acquired in 6 minutes (NA=30)1,4. As for the
quadrature coil, our previous results showed more 30% more noise in one of the
channels, with no net increase in SNR compared to the 1-channel coil. Here we describe
an optimized transmit/receive quadrature head coil (Figure 1). Our goal is to
improve the SNR, thus allowing less averaging to be performed and shortening scan
time. To evaluate the performance of the quadrature coil, both phantom and in
vivo brain images were acquired, and the results were compared with the
one-channel coil. Materials and Methods
Coil description: As
in the previous work2, the same quadrature helmet made up of two
nesting layers was used. The inner was not modified and consisted of 30 spiral turns
with a B1 field in the axial (x)-direction. The outer layer coil was rewound into a biplanar configuration with 29
orthogonal loops and a resultant B1 field in the anterior-posterior
(y)-direction (Figure 1). Both B1
fields are orthogonal to the B0 z-axis.
In this optimized quadrature coil, the two layers were first
assembled together. Incomplete geometric decoupling resulted in some inductive
coupling between the channels. Here, instead of the air-core transformer as
used previously a capacitive decoupling element (Cd) was added in
the circuit (Figure 1 A and B). Hence
both the tuning/matching and decoupling were performed all at once. To null
this mutual coupling, the capacitive decoupling method was preferred because
the transformer was very position sensitive thus unstable and resulting in
constant tuning/matching. The outer layer and inner layer were tuned at 276.2 kHz
and matched to at least -35 dB (S11/S22 parameter), and both coils were
geometrically decoupled by –23 dB (S12 parameter).
Imaging: A water-filled
structured phantom3 was placed inside the coil. The quadrature head coil was placed in our ULF MRI scanner, with the inner coil was
connected to the active T/R switch and the outer coil directly to
a second receiver channel. A 3D b-SSFP sequence was used with 50%
undersampling ratio, matrix size of 75×64×15, voxel size of 2.5×3.5×8.5 mm3,
the number of averages of 40 and the scan time was 8.5 minutes. The
performance of the quadrature coil was compared to that of the one-channel
spiral coil. All the imaging parameters, the receiver gain and shim set values
were kept constant. This same experiment was also performed in vivo on
the brain. All experiments were duplicated. Noise correlation for the
two-channel array are calculated using noise vectors acquired by the loaded
coil with the transmit RF voltage set to zero (Figure 5).
Data Analysis: Raw
data was reconstructed with the sum of squares (SOS) method from both channels and
compared to the previous single-channel spiral head coil. The maximum SNR and
mean SNR were calculated by taking either the maximum signal or the mean signal
over the ROI of the phantom/brain and dividing it with the standard deviation
(std) of noise in a specified region-of-interest (ROI).
Results
Figure 2 and Figure 4A show the images obtained from the
phantom experiment with the corresponding SNR values. The images acquired with
the quadrature coil have more contrast than the single channel coil. We observe
that the center part of all the slices is brighter for the inner layer coil probably
due to a higher Q coil, which can be lowered by adding a resistor. (Note: the
slices acquired with quadrature coil are not an identical match to that of the
single channel coil). The SNR evaluated across the slices for the quadrature
coil shows a significant increase the SNR. The maximum SNR of the quadrature
coil was of 144.3 versus 93.5 for the one-channel coil. Data was also acquired
with NA=20 with the quadrature coil and the maximum SNR was evaluated to 99.7,
showing that with half the scan time the same SNR can be achieved as with the
single channel coil. For the in vivo brain experiment, similar results were
obtained in Figure 3 and its corresponding SNR analysis in Figure 4B. Images
obtained with the quadrature coil were brighter and the maximum SNR enhancement
was of 1.42 when compared to the one channel coil.
Discussion and Conclusion
An increase in the maximum SNR of 40% was observed in vivo with
the quadrature coil as compared to the one-channel coil. Noise correlation
values of around 5% was obtained, showing minimal coupling between channels.
The phase images of each channel were out of phase with each other, showing
that the 2 layers were geometrically decoupled. Conclusion: We
have demonstrated an optimized quadrature head coil for use at ULF that will
allow us to accelerate overall imaging. Acknowledgements
No acknowledgement found.References
1. ‘A single channel spiral volume
coil for in vivo imaging of the whole human brain at 6.5 mT’, C. D. LaPierre, M.
Sarracanie, D. E. J. Waddington and M. S. Rosen, ISMRM abstract Proc. Intl.
Soc. Mag. Reson. Med. 23 (2015) 5902
2. ‘Quadrature Head Coil for Brain Imaging at 6.5 mT’, N.
Koonjoo, B. Primavera, J.P Stockmann, T. Witzel, L.L Wald and M. S. Rosen, ISMRM
abstract Proc. Intl. Soc. Mag. Reson. Med. 25 (2017) 2664
3. ‘Quadrature detection in the laboratory frame’, D. I.
Hoult, C.-N. Chen, V. J. Sank, Magnetic Resonance in Medicine 1 339–353 (1984)
4. ‘Low-Cost High-Performance MRI’, M. Sarracanie, C. D.
LaPierre, N. Salameh, D. E. J. Waddington, T. Witzel and M. S. Rosen, Scientific
Reports 5 15177 (2015)