Dimitri Welting1, Carel C. van Leeuwen1, Alexander J.E. Raaijmakers1, and Dennis W.J. Klomp1
1Department of Radiology, University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
An automated 32 channel vector network analyzer
switch was developed in-house, to be used for high density coil arrays. Using
this device full S11 and S21 measurements of 32 channels can be acquired on the
bench in less than 2.5 minutes. Performance of the device was evaluated using a
32 channel 7T headcoil and a 8 channel 7T dipole body array.
Introduction
Due to
the need for high accelerations, high density coil array designs are now
commonplace. However measurement equipment is generally not designed for such
high number of ports. The relatively low number of ports on vector network
analyzers (VNA) makes bench measurements of the coil interaction slow and error
prone due to the need to keep switching the coil connectors on these ports. This study presents an automated 32 channel RF
switching device (Figure 1) which increases the total number of ports in a VNA from 2 ports to a
virtual 32 ports.Methods
Hardware design:
The design of
the 32 channel RF switch is centered around low insertion loss RF switches,
with a working frequency range encompassing 20 to 3000 MHz to include all
commonly used MRI RF frequencies. These RF switches are connected to 50 Ohm
coplanar waveguides and striplines to minimize losses. Special care was taken
to ensure coupling to neighboring channels was minimal, resulting in a coupling
between channels of less than -30 dB. By incorporating a 4-way RF switch in the
final stage of the circuit, the input impedance of ports that are not currently
measuring can be set to 50 Ohm or to a preamp decoupling impedance. The final
stage of the VNA switch is a wideband 50 Ohm matching network to ensure the
attached coil is always connected to a 50 Ohm system. Extended features such as
the ability to impose DC bias for detuning and malfunction checking are also
included to further facilitate the measurement procedure.
The ports of
the VNA are connected to the back of the VNA switch. The signal is then led
through 3 RF switches to the selected channel on the front of the device. With
these RF switches any port combination can be realized, giving it a total of 1024
(322) possible port configurations which can be used in the analysis
of the connected coil.
The controlling
hardware consists of an ARM cortex m0+ microcontroller. The microcontroller
gives the device USB capabilities, to enable the possibility for fully
automatic measurements.
The board
schematics and PCB files are designed in the KiCad EDA suite (kicad-pcb.org).
Software design:
Software was
written in Python 3 to interface with the VNA switch and display measurement
data acquired from the VNA. The written software interfaces with a TR1300 2-port
VNA (Copper Mountain Technologies, Indianapolis, USA), but can easily be rewritten
to encompass other VNA’s capable of interfacing with personal computers.
Validation:
To test the VNA
switch performances and accuracy a comparison was performed between traditional
manual measurements and the measurements acquired via the device. The available VNA is limited to S11
and S21 measurements; therefore these two types of measurements will need to be
validated.
S11
measurements were performed on a 32 channel 7T headcoil (Nova Medical, Wilmington, USA) (Figure 2A). S21 validation was performed on an in-house built 8 channel 7T dipole
body array1 (Figure 2B), resulting in 64 measurements. Cable
traps were added to avoid measurement discrepancies due to cable positioning
differences between the manual and automatic measurements.Results and Discussion
S11
measurements of the 32 channel headcoil are shown in figure 3. Results show good agreement
between manual and automatic measurement; on average there is 98.7% accuracy in
absolute power on the 7T proton resonance frequency. These differences can be
explained by difference in cable positioning, slight differences between 50 Ohm
terminators used in the manual measurement and the wide-band 50 Ohm matching
performed on the VNA switch.
S21
measurements of the 8 channel dipole body array are shown in figure 4. Results show
good agreement; on average there is 99.7% accuracy in absolute power on the 7T
proton resonance frequency. However, an unexpected resonance peak can be seen
on multiple channels during the automatic measurement (Figure 4D). This resonance peak can occur
even without any coils attached to the VNA switch, which indicate improper
decoupling of the channels on VNA switch PCB boards causing a resonant loop to
be formed. Due to the lower power of the measurements the distortions caused by
these resonant peaks do not have a big impact on the accuracy.
S21 coupling
matrices of the two measured setups are shown in figure 5. Figure 5 A and B show matrices generated
from the automatic measurements, while figure 5C shows the coupling matrix
generated from the manual measurement.
Large
measurements consisting of 1024 individual measurements can be performed in
approximately 2.5 minutes, using 501 data points per measurement. Reducing the
amount of points to the minimum for single frequency measurements can
reduce this time to approximately 30 seconds.Conclusion
A 32 channel
vector network analyzer switch was built to facilitate measuring high density
coil arrays on the bench. With the 32
channel switch and corresponding software full 32 channel S-parameter matrices
can be acquired in as few as 30 seconds (one frequency) or 2.5 minutes (501
frequency points). Verification and characterization of transmit and/or receive
coil arrays is now sped up considerably. However, minor spurious resonances
exist in the current device, which cause erroneous dips in low-magnitude
S-parameter curves.Acknowledgements
No acknowledgement found.References
1. Raaijmakers
et al. The fractionated dipole antenna: A new antenna for body imaging at 7
Tesla. Magnetic Resonance in Medicine; 2015