Naoya Takagawa1, Masaya Ishikawa2, and Yasuhiko Terada1
1Institute of Applied Physics, University of Tsukuba, Tsukuba, Japan, 2Department of Forest Science, Graduate School of Agricultural and Life Sciences, The University of Tokyo, Tokyo, Japan
Synopsis
We
have developed a large-diameter, solenoid probe for a wide bore NMR
superconducting magnet that can image samples at different temperatures. The
controllable temperature range was -20 to 20 °C. To demonstrate the
performance, plant buds were imaged at different temperatures, and the frozen
process was visualized.
INTRODUCTION
NMR microimaging at variable temperatures has many applications,
such as plant imaging and cryobiopsies. A conventional variable-temperature
probe uses a saddle RF coil that has a lower SNR than a solenoid coil, or uses
a solenoid RF coil but with a limited sample area (<10 mm). In this study,
we developed a variable-temperature microimaging system with a 40 mm-diameter
solenoid coil for large sample observations. This system used a planar gradient
rather than a saddle one in order to provide a large space for samples. The
temperature was controlled by air and the whole system was simple, which
reduced the running and maintenance costs. We demonstrated microimaging of plant
samples during the freezing process, and showed the performance of the system. METHOD
The variable temperature RF probe was
designed for a 4.74 T/89 mm horizontal bore superconducting
magnet (Oxford Instruments plc, abingdon, UK). It consisted of a planar
gradient coil set, solenoid RF coil, and acrylic sample holder (the inner space
= 38 mm, length = 65 mm) (Figure 1). The solenoid RF coil (inner diameter = 40 mm
, length = 40 mm, 4 turns) was wound using a 1.5 mm polyurethane-coated Cu wire,
and divided into 4 segments by using 3 chip capacitors (4.8 pF) to achieve 50 Ω impedance matching at 202 MHz.
The planar gradient coil (Fig. 2) were
designed and fabricated as follows. The coil planes had holes (7 mm in
diameter) through which air pipes for cooling passed. The x and y gradients were
designed using GA-DUCAS 1 (homogenous area = 40 mm diameter of spherical volume (DSV),
coil plane diameter = 78 mm). The z gradient designed using an artificial bee colony algorithm 2 (homogenous
area = 40 mm DSV, coil plane diameter = 58 mm). The gradient coil was wound on FRP
plates using polyurethane coated Cu wire (wire
diameter = 0.4 mm). The efficiencies were 0.60 (x), 0.51 (y), 0.72 (z) G/cm/A, respectively.
The base metal of the probe was made of
acrylic instead of metal to avoid the eddy current field caused by gradient
switching.
Figure 3(a) shows a diagram of the cooling
system. Dried air from a dryer was stored in a tank, and cooled by a gas
chiller. Then the temperature of the flowing air was kept constant using a PID
controller and gas heater.
All experiments were performed on a
home-built digital MRI console (DTRX6, MRTechnology, Tsukuba, Japan). Flower buds, Aesculus
turbinate and Rhododendron, were imaged at different
temperatures using a 3D spin echo sequence (TR / TE = 800ms / 9ms, FOV = 51.2
mm × 25.6 mm × 25.6 mm, image matrix = 512 × 256 × 64, voxel size = 100 um ×
100 um × 400 um). RESULTS AND DISCUSSION
Figure
3(b) shows a temperature change in the empty sample holder. The temperature was
controllable ranging from -20 ℃ to 20 ℃. The temperature fluctuation after
reaching the steady state was within 0.8 °C at the target temperature of -10 ℃.
Figure
4 shows MR images of the flower bud acquired at 1, -3, -7, and -11 ℃. The
signals of florets and bud scales decreased as the temperature decreased. This indicates
that these tissues were frozen gradually within this temperature range. The
xylem was not frozen at -11 ℃ and was clearly visible in the image. This agrees
with the fact that xylem tissues are supercooled and hardly frozen in many plants 3.
Figure
5 shows MR images of the flower bud, Rhododendron, acquired at 0, -2, -4, -6,
and -8 ℃. The bud scales and the other tissues were frozen below -4℃, while the
anthers were partly frozen. The ovules were hardly frozen, which is the first
observation for Rhododendron.CONCLUSION
We
developed a variable temperature MR microimaging system using the
large-diameter solenoid coil for the vertical wide bore superconducting magnet.
We imaged plant flower buds at the low temperatures and demonstrated the system
performance.Acknowledgements
No acknowledgement found.References
1. K. Matsuzawa et al., A new method for
optimizing performances of gradient coils based on singular value decomposition
and genetic algorithm, 2017, 25th Annual Meeting & Exhibition
(ISMRM),Honolulu, USA.(4336)
2. Y. Terada et al., A Straightforward
Direct Optimization Method for Designing Biplanar Gradient Coils Using
Artificial Bee 3olony Algorithm, 2015, 23th Annual Meeting & Exhibition
(ISMRM), Toronto, Canada.(1834)
3. M. Ishikawa et al., Freezing
behaviours in wintering Cornus florida flower bud tissues revisited using MRI,
plant, Cell and Environment. 2016;2663-2675.