Tsinghua Zheng1, Xiaoyu Yang1, Blaise Whitesell1, Martin Domondon1, Paul Taylor1, Samuel Musilli1, and Labros Petropoulos1
1Quality Electrodynamics, LLC, Mayfield Village, OH, United States
Synopsis
We propose
a novel skipped-rung birdcage coil as the local transmitter for head imaging at
7 Tesla. This concept allows sizable openings on the coil anterior. The
openings improve both patient comfort and workflow. A 7T head coil using the skipped-rung
birdcage transmitter and a 32-channel receiver array was constructed and evaluated.
The preliminary test results show a good transmitting efficiency and a good
image quality compared with a commercial 7T head coil.
Introduction
Birdcage
coil has been used as a volume transmitter for its uniform B1
profile and good transmit efficiency at 3T and below(1). The
birdcage coil transmitting efficiency is affected by the coil loss, the sample
loss and the radiation loss. At ultra-high frequency (UHF), such as 7T and above, its transmitting efficiency reduces significantly
due to the excessive radiation loss. This is because the radiation power Pave
is proportional to the square of the frequency f(2).
PAve=4πf2μεE0(r)H0(r)
The amplitude
of the RF current through each rung of a birdcage may be written as Irung=
I0/N, where I0 is a scaling factor from the targeted B1
field strength. The total radiation power P may be the sum of all rungs that
can be written as P~(I0/N)2*N=I02/N.
Thus, one approach for reducing radiation loss of the 7T birdcage coil is to
increase the number of rungs. In many previous 7T head coil designs, the number
of rungs for the birdcage transmitter is equal or greater than16(3)(4)(5).
In addition, an RF shield must be placed
closely outside the birdcage for further reduction of radiation loss.
The obvious
drawback of more rungs and the RF shield is that it is becoming impossible to
have openings large enough for eyes on the anterior of the coil. The eye
openings are crucial for reducing claustrophobic reactions and facilitating
visual stimulation for 7T head applications. Therefore, the head imaging
experiences at 7T in terms of patient comfort and workflow are suffering.
We address
this issue by proposing a novel birdcage coil. The novelty is to strategically
remove a small number of rungs in the birdcage coil and the shield near eyes, and
create openings on the mechanical housing. The B1 field will have local
disturbance at where the rungs are removed. However, the overall B1 uniformity
across the whole imaging volume may not be affected significantly. We call this
concept a skipped-rung birdcage.Method
B1
uniformity evaluation was performed using EM simulation software Remcom XFtdt. A
birdcage coil with a partial shield model was created with the following parameters:
diameter=30cm, length=22cm, partial shield diameter=33cm, frequency=297 Mhz. Two
configurations were evaluated: (1) conventional birdcage coil with evenly
distributed 24 rungs, and (2) the same dimension birdcage coil except two rungs
adjacent to the top center rung are removed. The ring currents for both
configurations are the same. Fig.1 shows the simulation results of B1
profiles for both configurations using a 20cm radius no-load phantom. The
result shows the B1 uniformity within the phantom for both cases are
almost identical. We conclude that the removal of 2 rungs on the 24-rung
birdcage has minimal impact on uniformity.
Then a 7
Tesla T/R head coil with a skipped-rung birdcage as the transmitter and a 32-channel
receive array was constructed. Fig. 2 shows the section of the birdcage coil where
the rungs) were removed to create openings. A partial shield(6)
was implemented for further reduction of radiation loss.
The
receiver array was constructed on the same mechanical housing under the transmit
coil. It consists of 8 rows of 4 overlapped receive-only coils. 4 rows are in the
anterior and 4 are in the posterior. Each coil is connected to a low impedance
preamplifier to reduce coupling with its neighboring coils.
Fig. 3
shows the anterior and posterior housing of the 7T T/R 32 channel head coil.
Besides the two eye openings at the top, there is another opening at the
superior end of the coil that allows the EEG probes to go through. In addition,
the internal profile of the housing has mechanical cutouts to accommodate
headphones. Comparing to a commercial 7T head coil (Nova Medical), the
split-top design makes the workflow much easier. Even with larger inner
diameter, the new coil overall dimension is smaller (37cmx48cmx38cm) than the comparison
7T head coil (68cmx 48.5cmx 37.5cm). The coil is also 40% lighter (8kg vs.14kg).Results
The 7T T/R head coil with skipped rung birdcage was tested on a Siemens MAGNETOM 7T
scanner. Axial GRE images show uniform B1 profile on a no-load
spherical phantom, as shown in Fig. 4. The transmit adjustment voltage was 20%
higher (250V vs. 210V) than the comparison coil. We attribute the higher transmit voltage to a longer S-I dimension of the birdcage, which extends the
coverage to the upper c spine.
Volunteer
studies were conducted using the 7T T/R skipped-rung 32 channel head coil and
the comparison 7T 32 head coil. Results show that the two coils have comparable
image qualities with commonly used head imaging protocols, such as GRE, MP2RAGE
and EPI. Fig. 5 shows an example of an MP2RAGE result for comparison.Conclusions
This novel
coil concept allows the eye openings on the anterior of the coil. The openings
make the coil more patient friendly for head imaging at 7T. Preliminary
volunteer evaluation has demonstrated a good transmitting efficiency and a comparable
image quality as a commercial 7T 32 channel head coil. To further optimize the
skipped-rung birdcage concept, our future work may explore various configurations
of the rungs for the balance among the coil opening, the transmitting
efficiency, and the B1 uniformity.Acknowledgements
The authors are grateful to Dr. Mark Lowe, Dr.
Ken Sakaie and Dr. Sanghoon Kim at Cleveland Clinic Foundation (CCF) and Dr.
Ravi Seethamrazu at Brigham Women’s Hospital (BWH) for their Siemens 7T MR
system supports.References
(1) C.E. Hayes et al, An efficient, highly homogeneous
radiofrequency coil for whole-body NMR imaging at 1.5 T, J. of Mag. Reson. Vol
63, Issue 3, 622-628 1985
(2) M. Muhibbullah et al, Frequency dependent power and energy
flux density equations of the electromagnetic wave, Results in Physics
Vol. 7,
Pages 435-439, 2017
(3) P.J. Ledden, et al., 32 Channel Receive-Only SENSE Array
for Brain Imaging at 7T, Proc. Intl. Soc. Mag. Reson. Med. 15 2007.
(4) G.C. Wiggins, et al., A 32 Channel Receive-only Head Coil and
Detunable Transmit Birdcage Coil for 7 Tesla Brain Imaging, Proc. Intl. Soc.
Mag. Reson. Med. 14, 415 2006
(5) M. Finnerty et al, A 7-Tesla Transmit with 32-Channel
Receive-Only Array Head Coil for fMRI, Proc. Intl. Soc. Mag. Reson. Med. 2016-2133
2016
(6) M. Finnerty et al, A 7-Tesla High Density Transmit with
28-Channel Receive-Only Array Knee Coil, Proc. Intl. Soc. Mag. Reson. Med. 2010-642
2010