Belinda Ding1, Catarina Rua1, Johan D Carlin2, Marta M Correia2, Ajay D Halai2, Patrick Liebig3, Robin Heidemann3, Iulius Dragonu4, and Christopher T Rodgers1
1Wolfson Brain Imaging Centre, University of Cambridge, Cambridge, United Kingdom, 2MRC Cognition and Brain Science Unit, Cambridge, United Kingdom, 3Siemens Healthcare Limited, Erlangen, Germany, 4Siemens Healthcare Limited, Firmley, United Kingdom
Synopsis
Parallel transmit (pTx) has developed in recent years to
show promising reductions in signal dropouts and imaging artifacts from B1+
field inhomogeneities in 7T MRI. However, pTx methods have rarely been applied
for functional MRI. To our knowledge, no published task fMRI study has used pTx
with online pulse calculation. We therefore implemented pTx spokes excitation
in our vendor’s product EPI sequence, and tested it in volunteers using two
different task-based fMRI paradigms. Comparing with CP+ (or “TrueForm”) mode, using
pTx spokes pulses significantly improved both the tSNR and fCNR in task-based
fMRI acquisitions at 7T.
Introduction
At ultra-high fields(≥7T), the B1+ fields
generated by single channel coils are typically rather inhomogeneous in the
human head, resulting in signal dropouts in some brain regions. Parallel
transmit (pTx) techniques are able to improve image homogeneity and overall
quality at 7T1. Previous 7T pTx-fMRI
acquisitions have been largely limited to resting state data2,3. Here, we compare the
performance of a pTx excitation pulse against sinc excitation in the circularly
polarised (CP+, or “TrueForm”) RF shim for an fMRI study with two different task
fMRI paradigms.Methods
pTx sequence: The vendor’s standard gradient echo EPI
sequence (ep2d) was modified to use the pTx pulse design framework (Siemens).
The imaging slab (30 slices) was divided into 5 groups and spokes-2 pulses were
designed using the framework on the 3rd (i.e. centre) slice in each
group. Control scans used a circularly polarised (CP+) RF shim and sinc excitation.
Image acquisition: 6 subjects (3
females, age=28.3±4.5years) were scanned with an 8Tx32Rx head coil (Nova
Medical, USA) on a Magnetom Terra 7T system (Siemens, VE11U/K software).
Acquisition parameters for all EPI scans were: 30 axial slices across the
temporal lobe, FA=40°, TR/TE=2500/25ms, resolution=2.0×2.0×2.3mm3(slice direction), GRAPPA=2(24 reference lines), readout
bandwidth=1748Hz/Px, phase-encode=A>P. A scan with the same parameters as
above but with reversed phase gradient polarity(5 volumes) and a whole-brain structural
image(MP2RAGE, 0.75mm3 isotropic resolution) were also acquired.
Visual stimuli and task: fMRI datasets were acquired during two different visual paradigms
with pTx-EPI and CP+-EPI. Paradigm 1 consisted of a visual stimulus involving
images of faces (F), objects (O), scenes (S) and patterns (P) in 16s blocks (25
images/block). 16 image blocks (4 each) were interleaved with 4 blocks of rest
(16s), giving a total time of around 6 minutes per run. In each block, some
images were repeated and subjects were required to detect immediate repetitions
by pressing a response key (one-back task). Paradigm 2 consisted of a semantic
association (Se) and a pattern matching task (Pa). In the semantic association
task, three pictures were presented for each trial - a probe picture on the top
and two pictures at the bottom. The subject was required to select the picture
that was more related in meaning to the probe. The pattern-matching task had an
identical layout but the subject now had to choose the pattern identical to the
probe. Each run (about 9 minutes) consisted of 11 alternating blocks of each
task (16s, 4 trials) with rest blocks (10s) between them. In addition, one
extra resting state dataset was acquired for each sequence (100 volumes).
Data processing: Data were
analysed using FSL (v6.0.1). All EPI runs were distortion corrected with topup,
then motion corrected with MCFLIRT before further analysis. From the resting
state data, tSNR maps were obtained by taking the mean of the time series divided
by its standard deviation, before registering to the standard MNI152 brain.
Percentage tSNR change [δtSNR=(tSNRpTx-tSNRCP)/(tSNRpTx+tSNRCP)*200%] was calculated in 7 ROIs(Table 1)4.
For task data, each time-series was smoothed with a Gaussian
kernel (FWHM=4mm). The BOLD response for each run was modelled by convolving
the corresponding explanatory variables(4 for paradigm 1 and 2 for paradigm 2)
with a hemodynamic response function. General Linear Model analyses that
included motion parameters as regressors were performed. For paradigm 1, the
following contrasts were computed: S>F, S>O, with a threshold of p<10-4
(uncorrected); while for paradigm 2, the Se>Pa contrast was computed with
threshold of clusters determined by z>3.1 and a corrected significance
cluster threshold of p<0.05. Functional contrast-to-noise ratio
(fCNR=amplitude/standard deviation of noise) and number of active voxels
(nVoxels) were calculated for various ROIs and their percentage changes (δfCNR
and δnVoxels) are defined similarly to δtSNR. ROIs for paradigm 1 were based on
previous literature data5,6 (Table 1) while the same 7
ROIs used for tSNR calculations were used for paradigm 2.Results
tSNR: Figure 1 shows the tSNR map from a transverse slice in each volunteer. CP+ acquisition displayed
signal loss in the lower temporal lobe region (green circles) which was
recovered by using pTx excitation. Figure 2 shows the δtSNR across the 7 ROIs
for all volunteers. On average across all subjects, the pTx-mode outperformed
the CP+ mode in terms of tSNR in all 7 ROIs, with statistically significant
increases (Student’s paired t-test) in: lat. occ. inf., IT, MT and ST gyrus.
The average increase across all ROIs was 10.4%.
Functional data: With both
paradigms, CP+-EPI and pTx-EPI scans showed significant activation in the
temporal and fusiform regions as expected5–7. Figure 3 shows the
subject-average δfCNR and δnVoxels for paradigm 1 and 2. Larger cluster sizes
and increased fCNR were observed with the pTx-mode compared to CP+ in the
regions of interest.Discussion & Conclusions
The results from this experiment show the potential value of spokes-2 pTx excitation for task-based fMRI. Compared to the standard sinc CP+
excitation approach (CP+-EPI), pTx-EPI was able to increase tSNR across the
whole brain and reduce signal dropouts in the temporal lobes. In addition, with
simple targeted task-based fMRI, we showed that the pTx pulse design enabled an
average of 9.5% increase in fCNR and 26.1% increase in cluster size in brain
regions related to place-object perception and visual association.Acknowledgements
BD is supported by
Gates Cambridge Trust. CR is funded by the NIHR Cambridge Biomedical Research
Centre and the Isaac Newton Trust. CTR is funded by a Sir Henry Dale Fellowship
from the Wellcome Trust and the Royal Society [098436/Z/12/B]. This study was funded by the NIHR Cambridge Biomedical
Research Centre and MRC Clinical Research Infrastructure Award for 7T.References
1. Padormo, F.,
Beqiri, A., Hajnal, J. V. & Malik, S. J. Parallel transmission for
ultrahigh-field imaging. NMR Biomed. 29, 1145–1161 (2016).
2. Wu, X. et
al. Human Connectome Project-style resting-state functional MRI at 7 Tesla
using radiofrequency parallel transmission. Neuroimage 184,
396–408 (2019).
3. Gras, V.,
Poser, B. A., Wu, X., Tomi-Tricot, R. & Boulant, N. Optimizing BOLD
sensitivity in the 7T Human Connectome Project resting-state fMRI protocol
using plug-and-play parallel transmission. Neuroimage (2019).
doi:10.1016/J.NEUROIMAGE.2019.03.040
4. Rua, C. et
al. Improving fMRI in signal drop-out regions at 7 T by using tailored
radio-frequency pulses: application to the ventral occipito-temporal cortex. Magn.
Reson. Mater. Physics, Biol. Med. 31, 257–267 (2018).
5. Watson, D. M.,
Hartley, T. & Andrews, T. J. Patterns of response to visual scenes are
linked to the low-level properties of the image. Neuroimage 99,
402–410 (2014).
6. Hodgetts, C.
J., Shine, J. P., Lawrence, A. D., Downing, P. E. & Graham, K. S.
Evidencing a place for the hippocampus within the core scene processing
network. Hum. Brain Mapp. 37, 3779–3794 (2016).
7. Jung, J.,
Williams, S. R., Sanaei Nezhad, F. & Lambon Ralph, M. A. GABA
concentrations in the anterior temporal lobe predict human semantic processing.
Sci. Rep. 7, 15748 (2017).