Luisa Raimondo1, Tomas Knapen1,2, ĺcaro A.F. de Oliveira1, Xin Yu3,4, Serge O. Dumoulin1,5, Wietske van der Zwaag1, and Jeroen C.W Siero1,6
1Spinoza Centre for Neuroimaging, Amsterdam, Netherlands, 2VU University, Amsterdam, Netherlands, 3Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 4Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, SC, United States, 5Experimental and Applied Psychology, VU University, Amsterdam, Netherlands, 6Radiology, University Medical Centre Utrecht, Utrecht, Netherlands
Synopsis
We present
initial results of line-scanning fMRI in humans. The potential of this
technique lies in the combination of both high spatial and temporal resolution
while sacrificing spatial coverage outside the region of interest. We reached a
250 μm resolution along the line direction with a temporal resolution of 200
ms. Coil sensitivity profiles and the average tSNR per channel were used to optimize the line reconstructions. We obtained
similar BOLD sensitivity compared to standard 2D GE-EPI BOLD and high spatial
specificity for a visual task. Hence, we demonstrate the feasibility of
ultra-high spatiotemporal resolution in humans using line-scanning.
Introduction
Neurons with similar properties cluster
together into sub-millimeter columnar and laminar structures, moreover
neural activity occurs at millisecond resolution. Advances in fMRI approaches increase
either spatial or temporal resolution but never both. Instead, line-scanning
fMRI in rodents1 can achieve very high resolution across cortical
depth (50 μm) and time (50 ms), by sacrificing volume coverage and resolution along
the cortical surface. This high spatiotemporal resolution can also allow us to
isolate microvessel responses and to characterize the distribution of blood
flow and laminar fMRI profiles across cortical depth. Here, we present the
first human line-scanning implementation and results. First, we will focus on
the evaluation of the quality of the saturation pulses, followed by the
description of optimal coil combination for the reconstruction and finally we
will demonstrate the sensitivity of line-scanning fMRI in a comparison with
standard 2D GE-EPI BOLD fMRI using a visual task.Methods
We scanned five healthy volunteers at 7T
MRI (Philips) with a 32 channel receive head coil (Nova Medical). Line-scanning
data acquisition used a modified 2D gradient-echo
sequence: line resolution=250μm, TR/TE=200/13-22ms, 520 timepoints, flip
angle=16°, array size=720, line thickness=2.5mm, in-plane line width=4mm, fat
suppression using SPIR. Two saturation pulses (5 ms pulse duration) suppressed
the signal outside the line of interest. We computed the outer volume
suppression quality as the ratio signal along the line and outside the line.
The phase-encoding in the direction perpendicular to the line was turned off2.
The line was positioned along the right-left axis, crossing the visual cortex
(Figure 1). We acquired functional data using a block design in 6 runs. Visual
stimuli were 20 Hz flickering checkerboard,
presented for 10 s on/off. Reconstruction was performed offline (MatLab,
Gyrotools). We combined multi-channel coil data in four different ways: 1)
sum-of-squares (SoS), 2) tSNR-weighted SoS, 3) coil sensitivity weighted SoS
(csm), 4) tSNR and coil sensitivity weighted SoS (tSNR+csm). Resulting tSNR was
used to select the best coil combination. Functional data were analyzed using a
GLM approach and t-statistic values were computed to detect active voxels. We
also compared line-scanning data to a standard 2D GE-EPI BOLD acquisition with:
1x1 mm2 in-plane spatial resolution, TR/TE=200/22 ms, 600 dynamics,
flip angle=30°, FOV=176x176 mm2, SENSE factor 3, partial
Fourier=0.8. Here, we averaged the 2D BOLD timeseries data in the region of
interest coinciding with the line before computing t-statistics, resulting in
an activation profile along the line. Following manual coregistration and
averaging of line-scanning data every 4 voxels (in order to match the different
spatial resolutions of the two acquisitions), we also calculated the
correlation between the t-statistic values of line-scanning and 2D GE-EPI BOLD.Results
The achieved undesired signal
suppression outside the relevant cortical area was (97±0.4)% (Figure 1).
Figure 2 shows the tSNR for the four
different coil combinations. The weighted combination of both tSNR and csm
outperformed the other variants in terms of final tSNR. This coil combination
was selected for subsequent line-scanning reconstructions. Line-scanning data
averaged over 6 runs is shown in Figure 3a, for a representative subject. The
average line signal intensity profile through the occipital lobe is shown in
Figure 3b.
Figure 4a shows the t-statistic values
overlaid on the anatomical scan. Note the good spatial correspondence between
the positive BOLD t-statistic values and the grey matter ribbon, indicated with
white arrows. Figure 4b shows the time course of an example active voxel along
with the GLM.
Finally, in Figure 5 the comparison between the
line-scanning (left column) and standard 2D GE-EPI BOLD (right column) shows a
high similarity in BOLD sensitivity expressed by the t-statistic line profile.
Good correlation (R=0.75±0.17) was found between the t-statistic values for the
two acquisitions. Discussion & Conclusion
We report first line-scanning fMRI
results in humans with very high spatiotemporal resolution and show similar
BOLD sensitivity to standard 2D GE-EPI BOLD. Adequate outer volume suppression
can be achieved with saturation pulses. A coil combination including coil
sensitivity maps and tSNR/channel for the reconstruction furnishes a promising
temporal stability, since the resulting tSNR values are comparable to
sub-milllimeter 3D imaging and sufficient for BOLD signal detection3,
relatively to voxel size. Note that for the current line-scanning data
processing no temporal filtering was applied. Future experiments will examine
physiological noise contributions and its removal. Moreover, further
development will investigate improvements in signal suppression outside the
line of interest (2D spatial excitation, spin-echo beam excitation using
orthogonal 90-180° pulses) and increased SNR using surface coil arrays4.
Overall, the line-scanning fMRI
technique seems very promising due to its potential in detecting evoked BOLD
responses with sub-millimeter and sub-second resolution. Potential applications
for line-scanning are microvessel function measurements in clinical research on
cerebrovascular diseases but also fMRI at the mesoscopic scale such as cortical
lamina5.Acknowledgements
This study was supported by the Royal Netherlands Academy of Arts and Sciences Research Fund 2018 (KNAW BDO/3489) and the Visiting Professors Programme 2017 (KNAW WF/RB/3781) granted to the Spinoza Centre for Neuroimaging.References
-
Yu X et al.
Deciphering laminar-specific neural inputs with line-scanning fMRI. Nature
Methods. 2013; 10.1038/nmeth.2730.
- Siero J.C.W et al. BRAIN 2019, #10538.
- Van der Zwaag W et al. Temporal SNR Characteristics in Segmented 3D-EPI
at 7T. MRM. 2012; 10.1002/mrm.23007.
- Peridou N
et al. Pushing the limits of high-resolution functional MRI using a simple
high-density multi-element coil design. NMR in Biomedicine. 2012; 10.1002/nbm.2820.
- Petridou N
and Siero J.C.W. Laminar fMRI: What can time domain tell us? NeuroImage.
2019; 10.1016/j.neuroimage.2017.07.040.