Ziyi Pan1, Huilou Liang2,3, Kaibao Sun2, Danny J.J. Wang4, Rong Xue2,3,5, and Hua Guo1
1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China, 2State Key Laboratory of Brain and Cognitive Science, Beijing MRI Center for Brain Research, Institute of Biophysics, Chinese Academy of Sciences, Beijing, China, 3University of Chinese Academy of Sciences, Beijing, China, 4Laboratory of FMRI Technology (LOFT), Stevens Neuroimaging and Informatics Institute, University of Southern California, Los Angeles, CA, United States, 5Beijing Institute for Brain Disorders, Beijing, China
Synopsis
Multiline
bSSFP is an acceleration technique that has the potential to approach the speed of EPI while overcoming inherent EPI artifacts in fMRI. However, the highly segmented EPI readout in multiline bSSFP can cause severe Nyquist
artifacts. Additionally, extra
magnitude and phase differences among the echoes also exist within
the echo train, making it more complicated. In this
study, a joint-GRAPPA based 2D phase correction method is proposed, which can
robustly correct artifacts in multi-echo bSSFP with different ETLs while
maintaining the image SNR. To evaluate its performance, this approach is also compared
to other parallel imaging based correction methods.
Introduction
To
accelerate acquisition of balanced steady-state free precession (bSSFP), multiline
bSSFP, which employs a highly segmented EPI readout with a interleaved phase-encoding order, has recently been
proposed1. It can approach the speed of echo-planar imaging (EPI) and show an increased BOLD sensitivity
in fMRI. In addition, it is immune to artifacts
like signal dropouts and image distortions2 commonly seen in gradient-echo EPI images,
especially at ultrahigh magnetic fields like 7T. However,
in highly segmented EPI, alternating the readout polarity causes misalignment
of the k-space lines due to eddy current, field inhomogeneity and other hardware imperfections3,
producing severe ghost artifacts in multi-echo bSSFP with a sequential phase-encoding order, as shown in Fig. 1c.
Unlike
conventional EPI, multi-echo bSSFP with odd echo train length (ETL) has
identical readout polarities at the segment edges in k-space. For example, when
ETL = 3, the 3rd
echo of the first echo train and the 1st echo of the second echo train are both
positive readout, as seen in Fig. 1b. Thus this leads to N/ETL shift aliasing
artifacts. Furthermore, multi-echo bSSFP introduces extra magnitude difference
and phase errors within the echoes for each excitation. When data are combined
together for final image output, the signal difference makes linear phase
correction (LPC) methods and 2D phase correction methods (such as PEC-SENSE3) unsuitable.
By
estimating 2D phase difference maps and incorporating them into
auto-calibration signals (ACS), we propose a joint-GRAPPA4 based ghost
correction method to correct for the aliasing artifacts in sequential multiline or multi-echo bSSFP. It
can also preserve image signal-to-noise ratio (SNR), especially when ETL is
large. Methods
Data acquisition
Experiments were conducted on a 7T MRI scanner (Siemens,
Erlangen, Germany) with a volume transmit/32-channel receive head coil. Both multi-echo bSSFP and multi-echo integrated SSFP2 (iSSFP) sequence (a recently proposed method that can overcome susceptibility artifacts of bSSFP) were performed on a healthy subject using
three different ETLs (ETL 1/3/5) with the following acquisition parameters: TR =3.94/7.0/10.2ms , TE = 1.97/3.5/5.1 ms, slice-acquisition
time=864/574/542 ms, field of view = 200 x 200 x 3 mm3, resolution = 1 x
1 x 3 mm3,
flip angle = 35°.
Reconstruction
Basic GRAPPA is used
for reconstruction as shown in Fig. 2a. 1) The multi-echo k-space
data are separated into different groups based on their echo order within a TR,
and the missing data are filled with zeros. 2) Using the central part of the single echo k-space data (ETL=1) as the ACS lines, GRAPPA generates a set of
ghost-free images $$$I_{i}(x,y),\ i=1,2,\cdot\cdot\cdot,ETL$$$
recovered
from different echoes, respectively. 3) A final image can be obtained by simply
averaging the recovered set of images. However, this basic GRAPPA based method still suffers from image SNR decrease attributed to noise amplification during
parallel imaging reconstruction3, especially for a high ETL factor.
In
Fig. 2b, we demonstrate a joint-GRAPPA based ghost correction method to further
preserve image SNR. For joint-GRAPPA, a set of ACS data are first calculated to
match the phase of the multiple echoes. As seen in Fig. 3, 2D phase difference
maps between the GRAPPA recovered images and
the single echo images
are first calculated as $$$\triangle\varphi_{i}(x,y)=Arg\left[\
I_{i}(x,y)\ I^{*}_{ETL=1}(x,y)\ \right]$$$.
They are then incorporated into the single echo ACS data to generate a set of phase-matched
ACS data for the separated k-space groups of different echoes. Notice that
phase difference added to the single echo reference should be the same across channels. Thus, joint-GRAPPA can recover the
missing data jointly using different echo data together. As illustrated in Fig.
2b, the k space sampling is staggered, and joint-GRAPPA fits kernels across the
3 different echoes.
Similarly, sensitivity encoding (SENSE) based
correction methods are also implemented for comparison, applying single echo data to obtain coil sensitivity maps. SENSE can be easily accomplished, but the
joint-SENSE reconstruction, PEC-SENSE3, cannot be directly implemented due to the
magnitude difference of the images $$$I_{echo=i}(x,y) $$$. Instead, MUSSELS5, a method for the
recovery of multi-shot diffusion weighted images, is adopted. The multi-shot DWI
images with motion-induced phase difference are replaced by the multi-echo
bSSFP images with EPI-induced phase difference here. The low-rank matrix is
obtained from the separate groups with different echoes jointly. Results and Discussion
Fig.
4 demonstrates
the ghost corrected multi-echo bSSFP images, with ELT = 3 and 5.
Residual aliasing artifacts are observed on the SENSE-based reconstruction, but
are well corrected using GRAPPA and joint-GRAPPA.
Fig.
5 demonstrates the ghost corrected multi-echo iSSFP images with ETL=5. It is
observed in Fig. 5a that EPI acquired k-space data introduce both magnitude
differences and phase difference, even with the same readout polarity (e.g. echo1,
3, 5), as mentioned in the introduction. Fig. 5 b and c show that the proposed
joint-GRAPPA method can preserve the image SNR when ETL is large, both for the one-echo
and averaged images. GRAPPA, however, obtains noisy one-echo images. Conclusion
A
robust ghost correction method incorporating 2D phase difference maps with
joint-GRAPPA is proposed for multil-echo bSSFP and iSSFP at 7T.
This approach is reliable for different ETLs and demonstrates better SNR
performance over other parallel imaging based ghost correction methods. fMRI
results can be found in another abstract from our team.Acknowledgements
This
work was supported by the National Nature Science Foundation of China
grants (81871350, 31730039).References
1. Ehses P, Scheffler K. Multiline balanced SSFP for rapid
functional imaging at ultrahigh field. Magn Reson Med 2018;79(2):994-1000.
2. Sun KB, Xue R, Zhang P, et al. Integrated SSFP for
functional brain mapping at 7 T with reduced susceptibility artifact. J Magn
Reson 2017;276:22-30.
3. Xie VB, Lyu MY, Liu YL, Feng YQ, Wu EX. Robust EPI
Nyquist ghost removal by incorporating phase error correction with sensitivity
encoding (PEC-SENSE). Magn Reson Med 2018;79(2):943-951.
4. Bilgic B, Kim TH, Liao C, et al. Improving parallel
imaging by jointly reconstructing multi-contrast data. Magn Reson Med
2018;80(2):619-632.
5. Mani M, Jacob M, Kelley D, Magnotta V. Multi-shot
sensitivity-encoded diffusion data recovery using structured low-rank matrix
completion (MUSSELS). Magn Reson Med 2017;78(2):494-507.