Yuriko Suzuki1,2, Thomas Okell2, Joseph G. Woods2,3, and Michael Chappell1,2
1Institute of Biomedical Engineering, Department of Engineering Science, University of Oxford, Oxford, United Kingdom, 2Wellcome Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom, 3Department of Radiology, University of California, La Jolla, CA, United States
Synopsis
For
fMRI using ASL, multi-PLD acquisitions may have advantages, improving
reliability and specificity. However, the varying static-tissue signal in multi-PLD
ASL can confound motion estimation when conventional motion correction is
applied. In this study, we propose a novel framework using Gaussian processes
to address this problem, in which motionless ASL images are predicted, so that they
can be used as a reference for motion correction for each ASL volume. Simulation
and in-vivo studies show the new motion correction framework using Gaussian Processes
eliminates the influence of multi-PLD and provides a suitable reference for
each volume.
INTRODUCTION
Cerebral
blood flow (CBF) is an essential physiological parameter that shows a close
coupling with neuronal activity. Whereas the Blood Oxygen Level Dependent (BOLD)
technique is the traditional method for functional MRI (fMRI), several studies
have shown the advantages of Arterial Spin Labeling (ASL), such as better
spatial localization and CBF quantification1. For CBF quantification using ASL, while the
acquisition with single post-labeling delay (PLD) is currently the most common
approach, Mezue et.al. suggested that the use of multiple PLDs may improve the
reliability and specificity by accounting for arterial transit time (ATT)
variability1. However, when motion correction (MoCo) is required
(which is especially important for ASL-fMRI to detect statistically significant
activation from small regions of interest), choosing a suitable cost-function
for multi-PLD ASL is difficult: In ASL, background (static-tissue signal)
suppression (BGS) is performed to improve the SNR of CBF-map by suppressing
physiological subtraction artefacts. However, it introduces problems for
post-acquisition MoCo: The efficiency of BGS is often poorer for shorter PLD
due to less flexibility for optimization, which means the static-tissue signal
varies over different PLDs. Such varying tissue signal is not ideal for the use
with the conventional image similarity measure for motion estimation, which results
in poor motion estimates (Figure-1), and resulting artefacts following motion
correction.
In this
study, we propose a novel MoCo framework for multi-PLD ASL data, in which motionless
BGS-ASL images that present the identical static-tissue signal to each PLD is
predicted by employing Gaussian Processes (GP)2, so that they can be
used as a reference for MoCo for each ASL volume. METHODS
MoCo framework
Motion
estimation is performed using rigid-body registration between volumes in the
ASL time-series. To minimize artefactual overestimation as shown in Figure-1,
we aim to perform motion estimation on ‘ideal’ images of static-tissue having
removed other sources of variation besides motion. Although it is theoretically
possible to analytically predict the static-tissue signal after BGS if the underlying
function and variables are known (e.g. T1-recovery function with T1-value,
BGS-pulse timings, the efficiency of BGS and pre-saturation, as well as
magnetization transfer effects caused by pCASL labelling), it is not practical
to obtain all these parameters from the ASL time-series. Instead, we propose
the use of GP as a non-parametric approach to predict the ideal static-tissue
signal for each volume. Figure-2 explains the iteration process of the MoCo and
prediction of BGS-ASL data. The acquired multi-PLD ASL volumes are used as the
training dataset for GP. To minimize the influence of the subject’s motion in
the training dataset, (i) the mean-function µ was modelled by using the mean
value of the re-sliced ASL data for each PLD over entire epochs, and (ii) the
length-scale of the covariance function K (squared exponential in this study)
was set to 0.4 (larger than ΔPLD of 0.25s) to reflect that “smooth” variation
in static tissue with PLD introduced by BGS. This cycle is repeated several
times to obtain more accurate motion estimation.
In-vivo data acquisition and
motion-simulation
We
applied the framework to a previously acquired multi-PLD ASL-fMRI dataset1: single-shot
echo-planar imaging pseudo-continuous ASL (pCASL) images with 6 PLDs (250, 500,
750, 1000, 1250 and 1500ms (ΔPLD of 250ms) after pCASL labelling of 1400ms) repeated
10 times (10 epochs) without any intentional head motion. Additionally, a
BGS-ASL dataset was simulated by using the M0-image, CBF-map and T1-map and by
analytically applying the signal attenuation caused by BGS and perfusion for
each PLD. To help to observe the subtraction error, perfusion signal
attenuation was applied only to the left hemisphere, so that all pixel values
from the right hemisphere indicate the subtraction error. In 60 volumes of the
simulated dataset (i.e. 5 epochs), rotation was applied around z-direction
(yaw). RESULTS
Figure-3
shows the estimated motion parameters in the simulation study. When applying a
standard MoCo (Std-MoCo, using SPM12 in this study) to multi-PLD ASL, severe
overestimation caused by varying static-tissue signal was observed, although the
simulated motion (rotation around the z-axis) was correctly estimated. By
applying our framework using GP (GP-MoCo), the motion estimation was very close
to the ground truth. Figure-4 shows example images after control-label subtraction,
in which Std-MoCo resulted in obvious subtraction errors as indicated by
arrows. GP-MoCo avoided such subtraction errors, as well as successfully
correct motion. Finally, Figure-5 shows that GP-MoCo successfully eliminated
the wrong motion estimation of the in-vivo data shown in Figure-1. DISCUSSION AND CONCLUSION
This work
aims to address the limitation of the use of multi-PLD ASL for fMRI studies;
varying static-tissue signal over multiple PLDs hinders accurate motion
estimation. Our new MoCo framework using GP provides the reference images
presenting identical signal intensity to the BGS-ASL with multi-PLD, so that the
motion estimation is not influenced by different BGS-efficiency over multiple PLDs.
In this
study, several iterations were required to reach an accurate motion estimation
close to the ground truth. Potential future work includes further investigation
for the selection of the covariance function and hyper-parameters that could
improve the efficiency of results.Acknowledgements
This work was supported by Engineering and Physical Science Research Council (EP/P012361/1). The Wellcome Centre for Integrative Neuroimaging is supported by core funding from the Wellcome Trust (203139/Z/16/Z). TO is supported by the Royal Academy of
Engineering.References
1.
Mezue
M, Segerdahl AR, Okell TW, Chappell MA, Kelly ME, Tracey I. Optimization and
reliability of multiple postlabeling delay pseudo-continuous arterial spin
labeling during rest and stimulus-induced functional task activation. J Cereb
Blood Flow Metab. 2014; 34:1919-1927.
2.
Rasmussen,
CE and Williams, CKI. Gaussian processes for machine learning. MIT Press 2006.