Guoxiang Liu1,2, Adnan Shah1,2, and Takashi Ueguchi1,2
1Brain Function Analysis and Imaging Lab, CiNet, NICT, Osaka, Japan, 2Graduate School of Frontier Biosciences, Osaka University, Osaka, Japan
Synopsis
A
novel method for simultaneous functional and anatomical imaging is presented
for high-resolution fMRI at 7T. An RF inversion pulse is used before two EPI acquisition
trains where
functional scans are acquired alternating between shorter and longer inversion times
(TIs). Post-processing division of the shorter TI functional scan
with the longer TI functional scan allows improved localization of brain activity
in high-resolution fMRI at 7T. The proposed method allows simultaneous imaging
of brain function and anatomy as the grey matter tissues can easily be
segmented out from the resulting functional images, ensuring their perfect alignment
required for detecting functional activity.
Introduction:
Discriminating
BOLD responses at the level of cortical layers require isotropic submillimeter
resolution fMRI, which is possible at ultra-high-field (UHF, 7T or higher)
scanners. However, functional imaging at 7T suffers from large image distortions
resulting in a strong misregistration between function and anatomy much
exacerbated for submillimeter-resolution imaging. In ISMRM 2018, we proposed an
imaging and reconstruction method referred to as Block-Interleaved Segmented
EPI (BISEPI) for reconstructing fMRI volumes using discontinuous segmented k-space
data [1]. BISEPI method allows us to perform isotropic submillimeter resolution
BOLD fMRI while simultaneously preserving the hemodynamic response. In this work, we report a novel imaging technique based on BISEPI that
allows us to investigate sub-millimeter-level brain activity without the need
of an additional anatomical scan. An RF inversion pulse is used before two EPI acquisition
trains where
functional scans are simultaneously acquired alternating between shorter and longer
inversion times (TIs). Post-processing division of the shorter TI
scan with the longer TI scan allows improved localization of brain activity in
high-resolution fMRI. Moreover, the proposed method improves the
specificity of detected brain activity by suppressing contribution from
arteries and veins. The proposed method avoids functional to structural
transformation and spatial smoothing, giving us an opportunity to investigate brain
activity dynamics at the columnar or even layer level spatial resolution. Our
results showed that the proposed method can be considered as a choice to
perform voxel-wise high-resolution layer fMRI studies at 7T, as the functional
and anatomical information can simultaneously be obtained from the functional
scan. Materials & Methods:
In the proposed sequence, an inversion pulse is applied before two EPI
acquisition trains as shown in Figure 1. A human brain was scanned using this sequence
on a Siemens MAGNETOM 7T scanner with a 32-channel phased array head coil (Nova
Medical, MA, USA) to obtain two EPI series data (48 volumes each) at a spatial
resolution of 0.6×0.6×2.0 mm3. Acquisition parameters for
functional/structural scans were as follows: TR = 3000 ms, TE = 23 ms, flip
angle (FA) = 90°, # segments = 6, # shots in block = 16, PF = 6/8. A bilateral
finger tapping task was performed using BISEPI design [1] with ON and OFF for 12
s each. Six slices were scanned alternating between shorter and longer groups
of inversion times (TIs). TI1: 850ms to 1200ms, and TI2:
2350ms to 2700ms. Division of EPI1 (shorter TIs) by EPI2
(longer TIs) in each TR was calculated to build a new time series
called DVI=EPI1/EPI2. Three series (EPI1, EPI2,
DVI) were analyzed in BrainVoyager QX (Version 2.8.4.2645 Brain Innovation,
Maastricht, The Netherlands). The following pre-statistics processing was applied;
interscan slice time correction, high-pass filtering using a general linear
model (GLM) Fourier bases set of 2 sine/cosines as well as temporal Gaussian
smoothing with a full width at half maximum (FWHM) kernel of 2 data points. No
spatial smoothing was performed. GLM analysis was applied with stimulation
ON/OFF as a binary regression variable.Results and Discussion:
Figure 2 shows the same slice from EPI1,
EPI2, and DVI in our experiment, and the grey mater tissues segmented
from DVI using 2D k-means algorithm [2]. Figure 3 compares the brain activity
detection results of each series for two different slices and the different TIs.
EPI2 shows similar normal BOLD response with some veins suppressed
by the inversion pulse. BOLD-like response detected in EPI1 series is
more likely coming from arteries, because uninverted blood should enter the microvasculature
later than 1.4s after the inversion pulse applied [3] , while the TI1 in this
experiment is shorter than 1.2s. Negative BOLD-like response in DVI series is observed
to be located in grey matter area, showing the possibility that the effect of arteries
was canceled by the division. Another possible component of the negative BOLD-like
response in DVI series is VASO [4], even though we could not detect VASO
response directly from EPI1 series. Although signal inhomogeneity in both EPI1 and EPI2
is serious, but the division processing of EPI1 and EPI2
removed most of the signal inhomogeneity caused by varying tissue T2*, B1 and
even instability of the scanner, resulting in a pure T1-weighted image, which can
be segmented into grey/white matter tissues for next step analysis.Acknowledgements
This
study was supported in part by Japan Society for the
Promotion of Science (JSPS) Grants-in-Aid for Scientific Research “KAKENHI” (Grant
Numbers JP26282223 and JP26350471 and JP19K08244).References
[1]
Liu G, Shah A, Ueguchi T. Block-Interleaved Segmented EPI for voxel-wise
high-resolution fMRI studies at 7T. Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
5450. [2] MacQueen JB. Some Methods for Classification and Analysis of
Multivariate Observations. In: Cam, L. M. L., Neyman, J. (Eds.), Proceedings of
5-th Berkeley Symposium on Mathematical Statistics and Probability. Vol. 1.
University of California, Berkeley, Berkeley, USA; 1967; 1:281–297.
[3] Chen Y, Wang DJJ, Detre JA. Comparison of arterial transit times estimated
using arterial spin labeling. Magnetic Resonance Materials in Physics, Biology
and Medicine, 2012; 25:135–144. [4] Lu H and van Zijl PC. A review of the
development of Vascular-Space-Occupancy (VASO) fMRI. Neuroimage. 2012 Aug
15;62(2):736-42.