Xinyuan Miao1,2, Yuankui Wu1,2,3, Dapeng Liu1,2, Qin Qin1,2, Peter C.M van Zijl1,2, Jay J. Pillai4,5, and Jun Hua1,2
1Neurosection, Division of MRI Research, Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, MD, United States, 2F.M. Kirby Research Center for Functional Brain Imaging, Kennedy Krieger Institute, Baltimore, MD, United States, 3Department of Medical Imaging, Nanfang Hospital, Southern Medical University, Guangzhou, China, 4Johns Hopkins University School of Medicine, Division of Neuroradiology, Russell H. Morgan Department of Radiology and Radiological Science, Baltimore, MD, United States, 5Department of Neurosurgery, Johns Hopkins University School of Medicine, Baltimore, MD, United States
Synopsis
Metallic objects such as dental braces bring substantial susceptibility
artifacts in MR images acquired using echo-planar-imaging (EPI) sequences. Gradient-echo
EPI is currently the most commonly adopted sequence for blood oxygenation
level–dependent (BOLD) functional MRI (fMRI) in humans, which suffers from
susceptibility artifacts including signal dropout and geometric distortion in
the presence of metallic implants. Here, we demonstrate that T2-prepared
(T2prep) BOLD fMRI can significantly reduce susceptibility artifacts that are
commonly seen in GRE EPI in the presence of metallic orthodontic braces.
Introduction
MR images acquired using echo-planar-imaging (EPI) sequences are
sensitive to susceptibility artifacts such as signal dropout and geometric
distortion in regions affected by large susceptibility effects. Dental fillings
and orthodontic braces containing various metals can cause such artifacts
extending from the facial region into the brain in EPI1,2,3. This is particularly a problem for functional MRI (fMRI) studies
as gradient-echo EPI is currently the most commonly adopted sequence for blood
oxygenation level–dependent (BOLD) fMRI in humans. A whole-brain T2-prepared
BOLD fMRI approach4 was recently demonstrated, which uses a T2-preparation module to
induce the BOLD contrast, immediately followed by a 3D fast GRE readout with
short echo time (< 2 msec). T2-prepared BOLD showed little susceptibility artifacts
throughout the brain, and greater functional sensitivity than GRE EPI BOLD in
regions near air-filled cavities in healthy participants4, and around the lesions containing blood products in presurgical MRI
patients5,6. In this study, we apply T2-prepared (T2prep) BOLD fMRI4 in healthy human subjects wearing metallic orthodontic braces to
evaluate its ability to minimize susceptibility artifacts in the presence of
metallic objects7.Methods
Six
healthy participants (40±6yo, 3 females) were scanned. Removable dental braces
with bonding trays were used so that MRI images can be acquired with and
without braces in the same participants. Figure
1 illustrates the T2-prepared BOLD fMRI sequence, which consists of a
double refocusing T2-preparation module for generating the BOLD contrast,
followed by a 3D fast GRE readout. The following scans were acquired for each
subject on a 3T Philips MRI scanner: 1) MPRAGE (voxel=1x1x1mm3); 2) GRE-EPI-BOLD-fMRI (TR/TE=2000/30ms, flip angle (FA)=90°,
voxel=3.75×3.75×4 mm3, 40 slices, single-shot EPI, SENSE=3, fat
suppression); and 3) T2prep-BOLD-fMRI
(TR/TE=2000/50ms, FA=20°, voxel=3.75×3.75×4 mm3,
40 slices, single-shot 3D fast GRE, SENSE factor=2×1.5, centric phase encoding
profile, TRGRE/TEGRE=3.2/1.34ms). The order of fMRI scans was pseudo-randomized.
To assess BOLD fMRI signal changes in the entire brain, a breath-hold task8
was performed, which consists of 4 blocks of 40-second normal breathing,
4-second inhalation (5), and 16-second breath-holding, with an additional
20-second normal breathing period after the last block (total duration=4min20sec).
fMRI data was analyzed using SPM12 and Matlab. Preprocessing steps include
motion correction, slice timing correction (for 2D multi-slice EPI BOLD scans
only, not needed for 3D T2prep BOLD scans), co-registration between fMRI and
anatomical images, segmentation, and normalization. A general linear model (GLM)
was employed to detect functional activation (adjusted P<0.05, cluster
size≥3). fMRI results from the EPI
and T2prep fMRI methods were compared using an region-of-interest (ROI) based analysis. Two ROIs were manually delineated in each subject: one with strong
susceptibility artifacts (dropout), and one covering bilateral motor cortex
with minimal susceptibility artifacts in EPI. To calculate the relative signal
change (∆S/S) between breath-hold and normal breathing, the BOLD time course
from each scan was first averaged across the 4 blocks, and the peak-to-trough
difference described in the literature with similar breath-hold paradigms was
taken as the signal change 9. Temporal signal-to-noise ratio (tSNR)
was calculated as the signal divided by standard deviation along the time
course in each voxel. Contrast-to-noise ratio (CNR) was defined as the product
of tSNR and ∆S/S. Results
Figure 2 demonstrates the representative
image quality of the fMRI scans from one subject wearing metallic braces. While
little visible artifact in the brain can be observed on the anatomical (MPRAGE)
images, large signal voids can be seen on the GRE EPI BOLD fMRI images, which
include mainly the orbitofrontal and ventromedial prefrontal cortex (referred
to as the “dropout region” in subsequent text). Similar to MPRAGE images,
little visible artifact can be observed on T2prep BOLD fMRI images in the
entire brain. Typical activation maps during breath-hold are shown in Figure 3. While activation was detected
in most regions with minimal susceptibility artifacts using both methods, less
activations were detected in the dropout region with GRE EPI BOLD than with
T2-prepared BOLD. The group-averaged quantitative results from all participants
are summarized in Table 1. As
expected, tSNR, ΔS/S, and CNR in the dropout region were all significantly
higher in T2prep BOLD than in EPI images. On the other hand, in the motor
cortex where the susceptibility artifacts are expected to be minimal, GRE EPI
BOLD showed a trend (not significant) of higher tSNR, ΔS/S and CNR compared to
T2prep BOLD. In the same subjects when
not wearing braces, tSNR, ΔS/S and CNR in the dropout region were comparable
(P>0.1) in T2prep scans, but were significantly improved (P<0.001) in EPI
scans compared to corresponding results with braces. The results in the motor
cortex were comparable (P>0.1) with and without braces.
Discussion & Conclusion
We demonstrate that T2prep-BOLD-fMRI
can acquire BOLD images in healthy human subjects wearing metallic dental
braces with preserved CNR in the entire brain, whereas conventional EPI
approaches showed significantly reduced CNR in regions with strong
susceptibility effects. T2prep-BOLD-fMRI is expected to provide an alternative approach
in studies suffering from large susceptibility
artifacts, for instance in the presence of metallic implants in the brain, or for
adolescents wearing braces.Acknowledgements
NINDS (1R01NS108452),
NIBIB (R21EB 023538 and P41 EB015909), NICHD (U54 HD079123). References
1.
Starčuková J, Starčuk Z, Hubálková
H, Linetskiy I. Magnetic susceptibility and electrical conductivity of metallic
dental materials and their impact on MR imaging artifacts. Dent Mater.
2008;24(6):715-723. doi:10.1016/J.DENTAL.2007.07.002
2.
New PF, Rosen BR, Brady TJ, et al.
Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical
and dental materials and devices in nuclear magnetic resonance imaging. Radiology.
1983;147(1):139-148. doi:10.1148/radiology.147.1.6828719
3.
Hinshaw DB, Holshouser BA, Engstrom
HI, Tjan AH, Christiansen EL, Catelli WF. Dental material artifacts on MR
images. Radiology. 1988;166(3):777-779.
doi:10.1148/radiology.166.3.3340777
4.
Hua J, Qin Q, Van Zijl PCM, Pekar
JJ, Jones CK. Whole-brain three-dimensional T2-weighted BOLD functional
magnetic resonance imaging at 7 Tesla. Magn Reson Med.
2014;72(6):1530-1540. doi:10.1002/mrm.25055
5.
Hua J, Miao X, Agarwal S, et al.
Language Mapping Using T2-Prepared BOLD Functional MRI in the Presence of Large
Susceptibility Artifacts—Initial Results in Patients With Brain Tumor and
Epilepsy. Tomography. 2017;3(2):105-113. doi:10.18383/j.tom.2017.00006
6.
Agarwal S, Hua J, Sair HI, et al.
Repeatability of language fMRI lateralization and localization metrics in brain
tumor patients. Hum Brain Mapp. 2018;39(12):4733-4742.
doi:10.1002/hbm.24318
7.
Miao X, Wu Y, Liu D, et al. Whole-brain
functional and diffusion tensor mri in human participants with metallic
orthodontic braces. Radiology. (in press)
8.
Zaca D, Hua J, Pillai JJ.
Cerebrovascular reactivity mapping for brain tumor presurgical planning. World
J Clin Oncol. 2011;2(7):289. doi:10.5306/wjco.v2.i7.289
9. Urback AL, MacIntosh BJ, Goldstein BI.
Cerebrovascular reactivity measured by functional magnetic resonance imaging
during breath-hold challenge: A systematic review. Neurosci Biobehav Rev.
2017;79(April):27-47. doi:10.1016/j.neubiorev.2017.05.003