Dong-Hoon Lee1, Do-Wan Lee2, Chul-Woong Woo3, Hwon Heo4, Jae-Im Kwon3, Yeon Ji Chae4, Su Jung Ham5, Jeong Kon Kim2, Kyung Won Kim2,5, and Dong-Cheol Woo3,4
1Faculty of Health Sciences and Brain & Mind Centre, The University of Sydney, Sydney, Australia, 2Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea, 3Convergence Medicine Research Center, Asan Institute for Life Sciences, Asan Medical Center, Seoul, Republic of Korea, 4Department of Convergence Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea, 5Asan Image Research, Asan Institute for Life Sciences, Asan Medical Center, Seoul, Republic of Korea
Synopsis
We evaluated the effects of a reference image and
keyhole factor selections for high-frequency substitution on a keyhole imaging
technique for application in glutamate CEST (GluCEST) imaging to reduce data
acquisition time. The calculated GluCEST signals and visually inspected results
from the reconstructed GluCEST maps indicated that a combination of unsaturated
image as a reference image and >50% of keyhole factors showed consistent
signals and image quality as opposed to the fully-sampled CEST data. Combining
the keyhole imaging technique with GluCEST imaging enables stable image
reconstruction and quantitative evaluation, and this approach is potentially implemented
in various CEST imaging applications.
INTRODUCTION
Chemical exchange saturation transfer (CEST)
imaging requires a long scanning time for collecting the whole voxel data at
multiple frequency offsets. Here, to provide an approach for solving this
issue, we investigated the effect of a reference image and keyhole factor selections
on CEST results with respect to the application of keyhole technique1,2
in pre-clinical glutamate CEST (GluCEST) data acquisition. We evaluated the
selection of appropriate parameters for implementing the keyhole imaging
technique in the GluCEST experiment conducted for the healthy control rat group
and demonstrated the feasibility of implementing keyhole imaging combined with
GluCEST by applying the selected optimal parameters to the status epilepticus (SE)
rat group.METHODS
CEST Imaging:
All MR data acquisitions were conducted using a Bruker 7-T MRI scanner. For CEST
data acquisition, a fat-suppressed RARE sequence with a continuous-wave RF
saturation pulse (power/time=3.6 μT/1 s) was applied to 21 frequency offsets (−3.67 to +3.67-ppm; 0.33-ppm
increments), and an unsaturated image (S0 image) was obtained. For B0- and
B1-field inhomogeneity correction, B0 map was
acquired using the double TEs (1.9 and 2.6-ms) method, and B1 map
was acquired using the double flip-angle (30° and 60°) method.3,4
Keyhole Imaging:
We controlled the following two keyhole imaging parameters: 1) keyhole factor
(Kf), which is the ratio of the completely sampled encoding steps to
the encoding steps of low frequencies in the central portion and 2) selection
of reference image, for which the range of Kf varied from 16.67% (Kf
= 6) to 75% (Kf = 1.3) of the complete k-space. The reference images
corresponding to −3 ppm and +3 ppm images and unsaturated image were
respectively selected. For comparing the results by keyhole imaging, the zero-padding
algorithm was also applied to the missing k-space lines in the images collected
at low frequencies.5
Animal Imaging:
Ten male Sprague–Dawley rats were used and randomized into two groups [healthy
control (CTRL); n = 6 and SE group (i.p. injection of kainic-acid); n = 4]. The
keyhole imaging technique was first applied to the brain data of the CTRL rats for
evaluating the effect of Kf and reference image selections.
Subsequently, the best-performing approach was demonstrated in the brain data
of SE rats.
Data Processing: GluCEST signal was calculated in the region
of interest (ROI), which was carefully drawn in the hippocampus region using the
following equation: GluCEST(%)=100 × [(S-ω - S+ω)/S-ω],
where S±ω are signals at ±3.0 ppm.3 GluCEST contrast was
corrected by relative B1 values that were calculated based on the B1
map.3 For estimating the keyhole-GluCEST fidelity, artifact power
(AP) between the reconstructed images of keyhole-GluCEST and a fully sampled
image was calculated within whole brain regions using the following equation:
AP = Ʃ||Ifully-sampled(x,y)| - |Ikeyhole(x,y)||2 / Ʃ|Ifully-sampled(x,y)|2,
where Ifully-sampled is
the fully sampled image without keyhole imaging, and Ikeyhole is the image reconstructed using the keyhole
imaging.6,7RESULTS AND DISCUSSION
GluCEST results of various Kf values
and selected reference images are similar to those of the fully sampled data and
without statistical significance (Fig.1; all p ≥ 0.655). However, while using zero-padding technique, GluCEST
values decrease as Kf changes. Furthermore, GluCEST value at Kf
= 1.3 is similar across the results obtained from different selected reference
images and that from the fully sampled data. Compared with the reconstructed maps
of fully-sampled data (Fig.2), the images appear blurry, regardless of reference
image selection, when Kf is >4, and the boundaries of anatomical
structures are not clearly visible. Such blurring in the reconstructed images
are more dominant in case of zero-padding than in other reference images. The
results with Kf = 2 and Kf = 1.3 demonstrate relatively
clear anatomical structures in the brain. In all cases, AP value declines as Kf
decreases (Fig.3). In the results obtained via zero-padding, AP values are
relatively higher than those obtained from other reference images. There is no
significant difference in the Kf range, when images at −3 ppm or +3
ppm are selected as reference images. Notably, in the results obtained using the
unsaturated image as reference, AP values for all Kf changes were lower
than those obtained using other reference images. Additionally, we compared
CEST data acquisition time between fully-sampled data and keyhole-GluCEST data.
According to change in Kf, the scan time can be reduced by approximately
80% when using Kf = 6. However, considering all the results of this
study, the scan time with reasonable image quality and quantified glutamate
signal is reduced by approximately 48% when a minimum Kf value of
2 was applied. Keyhole-GluCEST values in the SE group corresponding to Kf
changes show a similar level, which is not significantly different from that associated
with the fully-sampled data (Fig.4; all p
≥ 0.992). Moreover, GluCEST values of the SE group corresponding to the
fully-sampled data and two different keyhole factors are significantly
different from those of the CTRL group (all p
≤ 0.033).CONCLUSION
We demonstrated the application of keyhole imaging
technique in GluCEST at 7-T. Our results indicated that substituting high-frequency
information in an unsaturated image and updating it with CEST data comprising at
least 50% of data collected at low frequencies facilitated a stable
reconstruction of consistent CEST signal and image quality.Acknowledgements
This work was supported by grants from the Basic
Science Research Program through the National Research Foundation of Korea
[NRF-2018R1C1B6004521 and NRF-2018R1A2B2007694], funded by the Korea Government
(MSIT).References
1. Hu X. On the
"keyhole" technique. J Magn Reson Imaging. 1994;4(2):231.
2. Varma G,
Lenkinski RE, Vinogradov E. Keyhole chemical exchange saturation transfer. Magn
Reson Med 2012;68(4):1228-33.
3. Cai K, Haris M,
Singh A, Kogan F, Greenberg JH, Hariharan H, Detre JA, Reddy R. Magnetic
resonance imaging of glutamate. Nat Med. 2012;18(2):302.
4. Singh A, Cai K,
Haris M, Hariharan H, Reddy R. On B1 inhomogeneity correction of in vivo human
brain glutamate chemical exchange saturation transfer contrast at 7T. Magn
Reson Med. 2013;69(3):818-24.
5. Du YP, Parker
DL, Davis WL, Cao G. Reduction of partial-volume artifacts with zero-filled
interpolation in three-dimensional MR angiography. J Magn Reson Imaging. 1994;4(5):733-41.
6. Park J, Zhang
Q, Jellus V, Simonetti O, Li D. Artifact and noise suppression in GRAPPA
imaging using improved k-space coil calibration and variable density sampling.
Magn Reson Med. 2005;53(1):186-93.
7. Xiao Z, Hoge
WS, Mulkern RV, Zhao L, Hu G, Kyriakos WE. Comparison of parallel MRI
reconstruction methods for accelerated 3D fast spin-echo imaging. Magn Reson
Med. 2008;60(3):650-60.