Davide Cicolari1, Domenico Lizio2, Patrizia Pedrotti3, Monica Teresa Moioli2, Alessandro Lascialfari1, Manuel Mariani1, and Alberto Torresin2,4
1Department of Physics, Università degli Studi di Pavia, Pavia, Italy, 2Department of Medical Physics, ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy, 3Department of Cardiology, ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy, 4Department of Physics, Università degli Studi di Milano, Milan, Italy
Synopsis
Relaxation times measurement
standardization, a great issue in clinical inter-centre and inter-scanner
applications, is studied by comparing relaxation times maps of a MnCl2
phantom, scanned with two different MRI imagers, with reference values measured
with an NMR laboratory spectrometer.
For this study standard
sequences were used: NMR reference T1 and T2
values were obtained from IR (Inversion-Recovery) and CPMG
(Carr-Purcell-Meiboom-Gill) sequences respectively; MRI maps were generated
from clinical IR and SE (Spin-Echo) sequences.
The MRI and NMR results
agreement within the experimental error limits (5%) suggests that the
estimation of the relaxation times is independent from the spectrometer/scanner
utilized.
Introduction
Performing clinical diagnosis by means of MRI pure
relaxation times maps 1 is a procedure that requires a high level of
standardization between methods of images acquisition and post-processing 2,3
and, most of all, among different scanners 4: qualitative and
quantitative information on tissues physical properties should be obtained
through the appropriate image acquisition and analysis.
User-verified validation of MRI scanners
standardization in measuring relaxation times is necessary especially using MRI
scanner-independent methods.
This study has the aim to assess the standardization
in measuring relaxation times values for 1H nuclei from
post-processing of phantom images acquired by means of different vendor
clinical MRI scanners taking as a reference T1
and T2 values estimated
with an NMR laboratory spectrometer.Methods
For this study an MRI
phantom was realised: the phantom is composed by twelve vials (30
ml) made of borosilicate glass with Teflon caps filled with different
concentrations of MnCl2 aqueous solutions. 5
Reference T1
and T2 values at 1.5 T of the solutions were measured with an NMR spectrometer
(Techmag Apollo spectrometer, Bruker electromagnet) scanning twelve smaller
vials (2 ml) made of the same materials and filled with the same solutions of
the bigger ones (see Figure 1). The temperature dependence of relaxation time
values was measured as well: the vials were analysed inside a cryostat in which
the temperature was set by a flux of liquid nitrogen and a heating resistance;
the temperature was measured with a thermocouple nearby the vial with a
sensibility of ±0.2 K.
The NMR sequences used to obtain the reference values
were standard sequences: SR (Saturation Recovery) and IR (Inversion Recovery)
for T1 measurements; SE (Spin-Echo) and CPMG
(Carr-Purcell-Meiboom-Gill) for T2 measurements. 6
MRI measurements of
relaxation times were performed with two different scanners: a Siemens Magnetom
Aera (1.5 T) and a General Electric Signa (1.5 T) MRI scanners, using clinical
SE and IR sequences. The temperature of the
phantom was monitored by means of an optical fibres sensor TempSense (Opsense
Inc., Quebec, Canada, Figure 2) with a sensibility of ±0.2 K. Images were post-processed
by the vendor independent open source software Segment 7 (http://www.medviso.com) in order to obtain
the relaxation times maps: measurements of relaxation times were performed with
Segment itself selecting ROIs (Regions Of Interest) on the generated maps.Results
NMR results concerning the
dependence of relaxation times from MnCl2 concentration and from
temperature are shown in Figure 3. An experimental error of ±5% was applied to
all values (from previous studies on the systematic error of the NMR
experimental set up). SR results are not shown because statistically coincident
with the IR ones (within the experimental error of 5%, as shown in the
Bland-Altman analysis 10 in Figure 4). SE results were affected by a
high diffusion that induce a severe underestimation of T2
values: the suppression of the signal due to diffusion is related also to the
static field inhomogeneity, that in our experimental system is estimated to be
approximately of 90 ppm. 6
A conservative experimental
error of ±5% was associated to clinical MRI imagers results too.
The comparisons between MRI
results and NMR results (these last ones taken as reference) are illustrated in
Figure 5. To obtain the reference values for the MRI results, which were
acquired at different temperatures with respect to the NMR measurements, a
linear regression of NMR data vs. temperature was used: with the coefficients
obtained from the linear regressions we could determine the relaxation times
reference values of each vials at any temperature. Again, an error of ±5% is
associated to the NMR reference values calculated in this way.
Compatibility tests between
MRI values and NMR reference values were performed by means of the two-sample
t-test, assuming the experimental errors equal to the standard deviations, thus
obtaining for all the comparisons significance level (p-values) greater
than 0.1.Discussion
In our study we assessed T1 and T2
values obtained from NMR measurements and set them as reference values for
inter-scanner T1 and T2 evaluation. A good agreement was found for both relaxation times
of each vials in the limits of experimental errors (±5% for both T1 and T2) using different and independent MR scanners (an NMR
spectrometer and two MRI scanners).
An error of ±5% was applied to MRI
results because Segment provides the mean value and the error (standard deviation) related to the
pixel intensities within a selected ROI (the pixel intensity in a relaxation
time map corresponds to the relaxation time value): the error calculated in
this way does not provide an accurate estimation of the error related to the
several processes that occur in the mapping process (image acquisition and
generation, pixel per pixel fitting procedure). Analysis methods as those
proposed by Kellman et al. in 2013 11 could lead to a better
estimation of experimental errors. Conclusion
Results suggest
that standard sequences are scanner-independent methods to establish pure
relaxation times values within the experimental error limits (5%): the
sequence-dependence and scanner-dependence, well known in literature especially
for fast mapping sequences 3,4,12, could be overcome by an harmonization
of acquisition parameters in order to obtain results as close to standard
sequences ones as possible.Acknowledgements
No acknowledgement found.References
1. Cheng H-LM, Stikov N, Ghugre NR, Wright GA.
Practical medical applications of quantitative MR relaxometry. J Magn Reson
Imaging 2012;36:805-824.
2. Roujol S, Weingärtner S, Foppa M, et al. Accuracy, precision, and
reproducibility of four T1 mapping sequences: a head-to-head comparison of
MOLLI, ShMOLLI, SASHA, and SAPPHIRE. Radiology 2014;272(3):683-689.
3. Baeßler B, Schaarschmidt F, Stehning C, Schnackenburg B, Maintz D, Bunck
AC. A systematic evaluation of three different cardiac T2-mapping sequences at
1.5 and 3T in healthy volunteers. Europ J Radiol 2015;84:2161-2170.
4. Raman FS, Kawel-Boehm N, Gai N, et al. Modified
look-locker inversion recovery T1 mapping indices: assessment of accuracy and
reproducibility between magnetic resonance scanners. J Cardiovasc Magn Reason
2013;15:64.
5. Thangavel K, Saritaş EU. Aqueous paramagnetic
solutions for MRI phantoms at 3 T: A detailed study on
Relaxivities. Turk J Elec Eng & Comp Sci 2017;25:2108-2121.
6. Brown RW, Cheng YN, Haacke EM, Thompson MR,
Venkatesan R. Magnetic resonance imaging: physical principles and sequence
design. 2nd ed. New Jersey: Blackwell; 2014.
7. Heiberg E,
Sjögren J, Ugander M, Carlsson M, Engblom H, Arheden H. Design and validation
of Segment – freely available software for cardiovascular image analysis. BMC
Med Imaging 2010;10:1.
8. Solomon I.
Relaxation process in a system of two spins. Phys Rev 1955;99(2):559-566.
9. Bloembergen N,
Morgan LO. Proton relaxation times in paramagnetic solutions. Effects of
electron spin relaxation. J Chem Phys 1961;34(3):842-850.
10. Altman DG, Bland JM. Measurement in medicine: the
analysis of method comparison studies. Statistician 1983;32:307-317.
11. Kellman P,
Arai AE, Xue H. T1 and extracellular volume mapping in the heart: estimation of
error maps and the influence of noise on precision. J Cardiovasc Magn Reson 2013;15:56.
12. Kim PK, Hong
YJ, Im DJ, et al. Myocardial T1 and T2 mapping: techniques and clinical
applications. Korean J Radiol 2017;18(1):113-131.