Qing Li1, Xiaoyue Zhou1, and Yi Sun1
1MR Collaborations, Siemens Healthcare Ltd., Shanghai, China
Synopsis
A dictionary-based
reconstruction method was applied to estimate T1 mapping from a series of
highly undersampled images acquired with an inversion recovery–prepared FISP
sequence and a radial readout. Motion robustness improved using interleaved slice
acquisition and data rejection in T1 estimation. The measurement time was 4
s/slice. Phantom and in vivo knee results show an up to 50% data rejection results
in less than 10% of quantification accuracy loss, compared to the T1 map
reconstructed with the full dataset.
Introduction
Patients
with Parkinson’s disease are with involuntary movements and require fast and
motion-robust imaging techniques. Fast T1 mapping with an inversion recovery (IR)–prepared
spoiled gradient echo (SPGR) sequence using radial acquisitions has been
reported by Wang et al. 1 k-space samples at different inversion
times (TIs) were grouped and reconstructed using a model-based sparse method.
However, the image is a mixed contrast of data from different TIs. Modified
Look-Locker inversion recovery (MOLLI) 2 is a frequently used fast
T1-mapping technique for myocardia that uses a balanced steady-state free-precession
(bSSFP) sequence with a Cartesian readout. The image weighting was mainly dependent on
the inversion time of the central k-space line. Thus, MOLLI and its variants were sensitive to
motion.3 Meanwhile, bSSFP-based sequences suffer from banding artifact especially at high field. In this study, a dictionary-based reconstruction method was
applied to an IR-prepared FISP sequence with radial readouts to achieve fast
and motion-robust T1 mapping.Methods
To
maximize the signal-to-noise ratio (SNR), the excitation flip angle (FA) was
optimized via Bloch simulation using different T1 and T2 relaxation times, and
the results are shown in Figure 1. The optimal FA increased with T1 and T2 value
increases. In this work, an FA of 15º and 20º was used for an in vivo knee
study and a phantom study, respectively.
Figure 2
shows the sequence diagram and reconstruction scheme. To improve motion
robustness, radial readouts were used and associated with interleaved slice
acquisitions after inversion preparation pulse. Radial spokes were rotated at a
small golden angle of 23.6º to maximize the data incoherence between adjacent
spokes from the same slice. With interleaved slice acquisition, the T1 sample redundancy
allowed for a retrospective motion correction via data rejection. Each spoke
was reconstructed to an image with high-undersampling aliasing using nonuniform
Fast Fourier Transform (NUFFT)4, shown at the bottom of Figure 2.
A dictionary-based
method was used to find the best match T1 value for each pixel, without
changing the image contrast by directly combining the k-space data. The dictionary
was generated using the IR signal model for all possible T1 values from the Bloch
simulation.
Experiments
were performed on a 3T MAGNETOM Prisma (Siemens Healthcare, Erlangen, Germany) using
a 20-channel head coil and a 15-channel knee coil. The imaging parameters were
an in-plane resolution = 1.0 × 1.0 mm2, slice
thickness = 5 mm, repetition time (TR) = 8 ms, and echo time (TE) = 3.7 ms. The
time interval between the inversion pulses was 8 s. Alternatively, to minimize
the waiting time for volume recovery, only two slices were acquired, leaving
other slices recovering. The acquisition time for each slice was 4 s.Results
Figure 3
shows a slice from the phantom results. The image rejection rate increased from
10% to 50% for the pattern shown in the left column. Compared with the T1 map
reconstructed with the full dataset, tubes with longer T1 maps were more robust
at data dropping.
Figure 4
shows the multi-slice knee results from the in vivo study. Images from slice numbers
1, 10, and 17 were selected and reconstructed using 30% and 50% data rejection.
Compared with the T1 map reconstructed from the full dataset, the
quantification error from the chosen region of interest (ROI; marked by a red
box) was 6.8% and 9.3% for 30% and 50% data rejection, respectively.Discussion
In this study, a fast and motion-robust T1 mapping method
was implemented with IR-prepared radial acquisitions and a dictionary-based
reconstruction. The proposed method was based on a FISP sequence, which should
not have banding artifacts at high field strengths as seen with the bSSFP sequence used in MOLLI.
Thus, the proposed method could achieve full FOV coverage, and could be applied to
abdominal areas. Moreover, the proposed method had a higher SNR than the IR
SPGR sequence, because the transverse magnetization was fully spoiled in the SPGR
sequence.
Dictionary-based
reconstruction also improved motion tolerance in the proposed method. Motion
influences for continuous TI samples on the same slice was reduced with an
interleaved slice acquisition strategy. Thus, motion artifacts would not have a
continuous impact on the TI samples from the same slice. The dictionary-based
T1 estimation used temporal sparsity of signal changes to allow data dropping. The
highly undersampled images from different TIs were not directly combined using
any sparse reconstruction method. Therefore, the image contrast from different
TIs was retained. However, streaking artifacts caused by radial undersampling were
observed in both the phantom and in vivo knee studies. One possible solution is
to apply a key-hole view-sharing strategy.5Conclusions
In this study,
we proposed a fast T1 mapping method with a high data rejection tolerance. This method directly
benefits the use of T1 mapping in patients with Parkinson’s disease.Acknowledgements
No acknowledgement found.References
1. Wang X, Roeloffs V, Klosowski J, et al. Model‐based
T1 mapping with sparsity constraints using single‐shot inversion‐recovery
radial FLASH. Magn
Reson Med, 2018, 79(2): 730-740.
2. Messroghli D,
Radjenovic A, Kozerke S, et al. Modified Look-Locker inversion recovery (MOLLI)
for high-resolution T1 mapping of the heart. Magn Reson Med, 2004, 52(1): 141-146.
3. von Knobelsdorff-Brenkenhoff, F.,
Prothmann, M., Dieringer, M.A. et al. Myocardial T1 and T2
mapping at 3 T: reference values, influencing factors and implications. J
Cardiovasc Magn Reson, 2013, 13:53.
4. Fessler JA, Sutton BP. Nonuniform fast Fourier transforms
using min-max interpolation. IEEE Trans. Signal Process. 2003;51:560-574.
5. Ehses, P., Seiberlich, N, Ma, D, et al. IR TrueFISP with a
golden‐ratio‐based radial readout: Fast quantification of T1, T2, and proton
density. Magn Reson Med, 2013, 69: 71-81.