Varut Vardhanabhuti1, Ho Ting Au2, Jie Ding1, Elaine Y Lee1, Peng Cao1, and Edward S Hui1
1Diagnostic Radiology, The University of Hong Kong, Hong Kong, Hong Kong, 2The University of Hong Kong, Hong Kong, Hong Kong
Synopsis
The purpose of this study is to assess the
repeatability of the magnetic resonance fingerprinting (MRF) sequence in a
Philips 3T scanner with reference to the ISMRM/NIST MRI system phantom. The
results showed a strong correlation between the T1 and T2
estimated from MRF versus the reference values (R2 =
0.999, and R2 = 0.981). A high level of
repeatability was achieved over the 15 scanning sessions, although the T1
estimated from MRF has a significantly higher degree of repeatability than T2.
Introduction
Magnetic
Resonance Fingerprinting (MRF) is a robust and efficient MR technique for the estimation
of MR parameters, such as T1, T2 and proton density
1,2. That these MR parameters are quantitative and largely do not depend
on MRI hardware and sequence configuration renders MRF an ideal technique for
longitudinal studies, such as treatment effect on cancer patients. It is
therefore imperative that the repeatability of MRF be evaluated. The
purpose of this study is to assess the repeatability of MRF based on the
inversion-recovery unbalanced steady-state (IR-FISP) sequence using the ISMRM/NIST
MRI system phantom3. Methods
All MRI experiments were performed using a 3.0T human
scanner (Achieve TX, Philips, Best, Netherland). The ISMRM/NIST MRI phantom
(Fig 1) was used for this repeatability study. It contains fiducial
spheres filled with well-characterized T1 and T2
relaxation times. The phantom was scanned using the IR-FISP MRF sequence over
15 sessions throughout 65 days. The IR-FISP MRF sequence
with spiral-in-spiral-out readout trajectory (acceleration factor = 58.4) was
developed in-house. Detailed imaging parameters were: TR = 8.5ms to 10.1ms,
flip angles = 5° to 60°, matrix size = 256 x 256, field of view = 30 x 30 cm2,
slice thickness = 5 mm, and 1000 dynamics per slice. An addition of a minimum 5
seconds delay was added before each acquisition to ensure the net magnetization
to be in its thermal equilibrium state. The resulting scan time per slice was
55s. MRF was performed 5 different times in each session. MRF dictionary was generated using the extended phase
graph algorithm using MATLAB (Mathworks Inc, Natick, MA). The dictionary
contained T1 values ranging from 300 to 6000 ms in 15 ms increment
and T2 values ranging from 10 to 6000 ms in 5 ms increment.
Region-of-interest analysis of each fiducial sphere
from all repetitions of each session was performed. In particular, the mean of
each ROI on T1 and T2
maps were measured, and the median and standard deviation of the means amongst
the 5 repetitions were recorded for all 15 sessions. Bland-Altman (B&A)
analyses were performed to assess the agreement between the T1 and T2
estimated from MRF sequence and those provided by ISMRM-NIST (Table 1). Results
There was a strong correlation between the T1
and T2 estimated from MRF versus the reference values (R2 =
0.999, and R2 = 0.981, respectively; Fig 2). In
Bland-Altman analysis, the T1 estimated from MRF has a bias of 74.8
ms, and the range of limit of agreement was between 8.37 and 141.2 ms (Fig 3a).
All measurements were within the limits of agreement. The T2 estimated
from MRF has a bias of -32.2 ms with range of limits of agreement between -94.3
and 29.9 ms (Figure 3b). There was a measurement that lied above the limit of
agreement (sphere T2-1) with a difference of 67.9 ms shorter than the reference
value. Conclusion
The MR parameters estimated from MRF display high
repeatability over the 15 sessions. The T1 estimated from MRF has a significantly
higher degree of repeatability than T2. Acknowledgements
NoneReferences
1. Ma D, Gulani V,
Seiberlich N, Liu K, Sunshine JL, Duerk JL, et al. Magnetic resonance
fingerprinting. Nature. 2013;495(7440):187-92.
2.
Jiang Y, Ma D, Seiberlich N, Gulani V, Griswold MA. MR fingerprinting using
fast imaging with steady-state precession (FISP) with spiral readout. Magnetic
resonance in medicine : official journal of the Society of Magnetic Resonance
in Medicine / Society of Magnetic Resonance in Medicine. 2015;74(6):1621-31.
3. Russek SE, Boss M, Jackson EF, Jennings DL,
Evelhoch JL, Gunter JL, Sorensen AG. Characterization of NIST/ISMRM MRI system
phantom. In Proceedings of the 20th Annual Meeting of ISMRM, Melbourne, Australia,
2012. Abstract 2456.