Xueying Zhao1, Ying-Hua Chu2, Qianfeng Wang1, and He Wang1,3
1Institute of Science and Technology for Brain-inspired Intelligence, Fudan University, Shanghai, China, 2MR Collaboration, Siemens Healthcare Ltd., Shanghai, China, 3Human Phenome Institute, Fudan University, Shanghai, China
Synopsis
Precise quantification of the contrast agent uptake in
DCE-MRI is still challenging. Here, we redesigned the reconstruction scheme of MR
fingerprinting by introducing a sliding window into the process of dictionary
matching, which allows dynamic quantification of T1 and T2* in DCE-MRI with high
temporal resolution. The performance of this proposed dynamic MRF method is simulated
and tested on MATLAB. The time-varying trajectories of T1 and T2* have been
successfully captured with a temporal resolution of 4.5 seconds. This method
may open the door of utilizing MR fingerprinting to quantify time-varying
variable with adjustable temporal resolution.
Introduction
Magnetic resonance fingerprinting (MRF)1 is a novel quantitative imaging approach that allows simultaneous measurements of multiple tissue properties during a single
scan. It has been applied to DCE-MRI studies for dynamic T1 and T2* mapping
with a two-minute temporal resolution2.
Although quantitative T1 and T2* can be addressed at the same time, the temporal
resolution is still not enough for human DCE-MRI.
The existing idea of DCE-MRF2
is doing MRF sequence repeatedly to achieve dynamic tracking and the temporal
resolution is restricted by the acquisition time of a single MRF sequence. Here,
we redesigned the reconstruction scheme by introducing a sliding window into
the dictionary matching process of MRF, in which the time-varying T1 and T2*
can be calculated dynamically within a single MRF acquisition. And by doing
this, high temporal resolution can be achieved without much changes in MR
sequence, because the temporal resolution will merely depend on the step length
of the sliding window.Theory
Inspired by the application of the sliding window in MR data
acquisition3, in which the K-space
data can be shared between time-consecutive images, we found that the data
frames used for the dictionary matching of MRF can also be shared.
As a proof-of-principle implementation, we adopted an
EPI-based MRF sequence with varying FAs and TEs4.
The data frames are composed of a series of GE-EPI images. The design for dynamic
reconstruction is illustrated in Figure 1(c). The data frames used for dictionary matching will be
selected by a sliding window. This window will start at the beginning of our
acquired data frames, then gradually slides with a step length.
For each window position, T1 and T2* values at the center of
the window will be reconstructed through dictionary matching, a matching between
the data frames selected by this window and a calculated dictionary. Then the
obtained T1 and T2* values will be used to calculate the initial magnetization
as well as the corresponding dictionary of the next window.Methods
Our MRF sequence has 760 frames in total, covering 41.59
seconds. The FAs and TEs patterns are shown in Figure 1(a-b). Minimal TR was chosen for any
given TE. The width and step length of the sliding window are 160 and 80
frames, separately.
The MR signal of the DCE experiment was simulated as
follows. First, the time-varying T1 and T2* curves were calculated based on the
empirical concentration of contrast agents during DCE-MRI in human brain5. The blue lines in Figure 2 (c-d) shows the
simulated T1 and T2* curves. Then, we simulated the measured MR signal under the
specific FA and TE patterns above using the Bloch equation. At last, the time-varying T1 and T2* were reconstructed by our proposed reconstruction
scheme without any prior knowledge including the characteristics of the bolus,
the initial T1 or T2* value, etc.Results
The simulated MR signal under the time-varying T1 and T2* is
exhibited as the blue line in Figure 2(a). The red circles represent the
predicted magnetization of each window calculated during the dynamic
reconstruction process, which shows a high agreement of the blue line. The
goodness of dictionary matching is illustrated in Figure 2(b), where the dashed
blue line represents the acquired MR signal, and the best matching curve from dictionary
matching are superposed. The final reconstructed T1 and T2* curves are shown in Figure 2(c-d). The trend of the time-varying relaxation time has been
successfully captured especially for T2*. Discussion
The results in Figure 2 showed that it is possible to track
time-varying variables through this sliding window aided dictionary matching
scheme. The temporal resolution is only determined by the step length of the
sliding window. In the present demonstration, the temporal resolution is 4.5
seconds approximately, based on our choice of 80-frame step length. There is no limit to the choice of step length, which means in principle, the temporal
resolutions of our propose method can reach as high as we want.
Inconsistence in dictionary matching was found around frame #200~300 in Figure 2(b), which
may be addressed by further adjusting the width and step length of the sliding
window. This choice of window parameters should be optimized together with the
MRF sequence design, because the matching process will be influenced by the
coherence between them. Also, a sliding window with different weights over frames could be used for further studies. Besides, the transmit magnetic field B1 can also be included in the dictionary calculation and be retrieved along with T1
and T2*. Last but not least, experimental data will be added in our future
results, to fully evaluate the performance of this dynamic reconstruction scheme.Conclusion
Our proposed dynamic reconstruction scheme of MRF has shown
promising simulation results of capturing the time-varying T1 and T2* curve
without any prior knowledge. This may open the door of dynamic MRF, which can
be used to quantify the contrast agent uptake with high temporal resolution. Acknowledgements
This work was supported by Shanghai Municipal Science and Technology Major Project (No.2017SHZDZX01), Shanghai Municipal Science and Technology Major Project (No.2018SHZDZX01) and ZJLab, Shanghai Natural Science Foundation (No. 17ZR1401600) and the National Natural Science Foundation of China (No. 81971583).References
1. Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance
fingerprinting. Nature. 2013;495(7440):187-92.
2. Gu Y, Wang CY, Anderson CE, et al. Fast magnetic
resonance fingerprinting for dynamic contrast-enhanced studies in mice. Magn
Reson Med. 2018;80(6):2681-2690.
3. Cao X, Liao C, Wang Z, et al. Robust sliding-window
reconstruction for accelerating the acquisition of MR fingerprinting. Magn Reson
Med. 2016;78(4):1579-1588.
4. Rieger B, Zimmer F, Zapp J, et al. Magnetic resonance
fingerprinting using echo-planar imaging: Joint quantification of T1 and T2∗
relaxation times. Magn Reson Med. 2017;78(5):1724-1733.
5. Lind E, Knutsson L, Kämpe R, et al. Assessment of MRI
contrast agent concentration by quantitative susceptibility mapping (QSM):
application to estimation of cerebral blood volume during steady state. MAGMA.
2017;30(6):555–566.