Rasim Boyacioglu1 and Mark Griswold1
1Radiology, Case Western Reserve University, Cleveland, OH, United States
Synopsis
Magnetic Resonance Fingerprinting
(MRF) maps multiple tissue properties and system parameters simultaneously. The
accuracy of MRF maps depends on the simulation of all possible system
properties into the signal evolutions via the Bloch equations. We have observed
frequency drifts during MRF scans, similar to those seen in fMRI scans, which
might cause various artifacts if not accounted for. Here, it is shown that 2D
MRF frequency drifts can be compensated with a simple dictionary update. For
correction of 3D MRF frequency drifts, a novel reconstruction framework is
introduced. Results show significant improvements on quantitative maps for 2D
and 3D MRF.
Introduction
Magnetic Resonance Fingerprinting
(MRF) is a novel imaging technique capable of simultaneous mapping of multiple
tissue and system properties1. MRF data is typically acquired with
fast sampling of highly undersampled k-space with a temporally variable FA/TR.
The reconstructed MRF signal evolutions are assumed to be encoded by the
underlying tissue properties which are then obtained with dictionary matching.
The dictionary is simulated considering the possible tissue and system property
ranges (field inhomogeneities, slice profile, etc.). Because of this, the accuracy
of MRF maps depends on the exact knowledge of system properties throughout the
acquisition. Recent experiments revealed ~4-5 Hz/min scanner frequency drifts
during continuous 2D MRF acquisitions due to the eddy currents from fast
switching gradients. If not accounted for, any discrepancy between the expected
and the real on-resonance frequency might cause problems in the dictionary
matching stage. Here we incorporate experimental measurements of frequency drifts
into dictionary generation and suggest a novel MRF reconstruction framework for
3D. Improved image quality is shown when these drifts are taken into account.Methods
Recently, MRF with quadratic RF
phase2 (MRFqRF) was introduced to map off-resonance and T2*, in
addition to T1 and T2. Since MRFqRF is particularly sensitive to frequency
drifts with artifacts as blurring in T1 maps and overestimation of Γ, it is a
good candidate to test our hypothesis.
2D MRFqRF brain data3:
256x256 matrix size, 1.2x1.2 mm2 resolution, 5 mm slice thickness,
3516 time points (tp). 3D MRFqRF brain data4: 24 partitions
(Np), same acquisition parameters as in 2D (Figure 1) but with 3D excitation
and phase encoding in z-direction with no acceleration. Also for reference, 3D
FISP MRF data (matching MRFqRF FOV) were also acquired. Informed consent with
IRB approval was obtained prior to the data acquisition.
For frequency drift estimation,
we acquired 10 consecutive 2D MRFqRF scans and calculated the mean
off-resonance change per TR over an ROI. Then, for frequency drift compensated MRFqRF
dictionary, every entry was updated every TR with an increment of the resulting
mean frequency drift of 0.0013 Hz/TR within the Bloch simulations. Dictionary
and the data were compressed with randomized SVD5 (rSVD) prior to
the reconstruction. The dictionary was also undersampled in the tissue property
dimension. Quadratic interpolation6 recovers the high resolution in
the tissue property dimension. The same 2D MRFqRF data were reconstructed with
the generic and frequency drift compensated MRF dictionary.
During generic 3D MRF reconstruction
each k-space partition is compressed to Nk rank with the compression
matrix calculated from rSVD of the standard dictionary (Figure 2a). Then, Nk
fully sampled k-space data go through non-uniform FFT and the resulting Nk
images get matched to the compressed dictionary. However, in the case of
frequency drift, each partition will have a frequency offset of an integer
multiple of tp*0.0013 Hz compared to the very first time point in
the acquisition. In this case, the compression matrix, even though it’s
calculated from the frequency drift compensated 2D dictionary, would only apply
to the first partition and not the others.
To overcome this problem, we
suggest a new framework for 3D MRF reconstruction (Figure2b) where we calculate
one SVD that can compress all partitions than might have slightly different
signal evolutions due to the drift. The same 3D MRFqRF data were reconstructed
with the generic and frequency drift corrected dictionary for comparison.Results
2D results are shown in Figure 3.
Frequency drift correction mainly improves the underestimation of T1 over the
whole brain but most dramatically in ventricles. Γ values are overall decreased and T2*
contrast gets sharper.
Improvement in 3D T1 maps is
noteworthy (Figure 4). Compared to FISP, there’s significant blurring in T1
maps with standard reconstruction which is practically corrected with frequency
drift compensation. 3D Γ maps might be improved the most (Figure 5) since with
frequency drift it does not have to explain all the frequencies that
existed in the voxel throughout the acquisition. 3D T2* maps have similar
contrast to 2D T2* maps.Discussion and Conclusions
Here, upon the observation of
scanner frequency drifts during dynamic MRFqRF scans and its possible effects,
two potential retrospective correction schemes are suggested. For 2D MRF, the
dictionary can be updated with manual increments of measured frequency drifts
to compensate for unexpected differences between the acquired data and the
dictionary. For 3D MRF, a new reconstruction framework is introduced. The
results show significant improvements especially for T1 and T2* maps.
The frequency drift is specific
to our system and sequence and could be different on other scanners. Here this
value was manually hardcoded into the dictionary but one can also consider
making a frequency drift dimension in the dictionary (assuming its time
evolution, i.e. linear) and obtain the frequency drift with dictionary matching.
3D reconstruction framework
introduced here opens the door to different and optimized sequence parameters
for each partition. Thus, MRF optimization problem may have more degrees of
freedom and not be limited to number of time points per partition.
In conclusion, it is necessary to
compensate for the frequency drift in MRF, especially for bSSFP based MRF
sequences, to obtain accurate quantitative maps and a simple retrospective
framework achieving that was described.Acknowledgements
This work is supported by Siemens
Healthcare.References
1. Ma D, Gulani V, Seiberlich N,
et al. Magnetic resonance fingerprinting. Nature 2013;495: 187–192.
2. Wang C, Coppo S, Mehta B, et al. Magnetic resonance
fingerprinting with quadratic RF phase for measurement of T2* simultaneously
with δf
,T1, and T2. Magn Reson Med 2019 81(3):1849-1862.
3. Wang C, Boyacioglu R, McGivney D, et al. Magnetic
Resonance Fingerprinting with Pure Quadratic RF Phase. ISMRM proceedings 2019, p4552.
4. Boyacioglu
R, Wang C, Ma D et al. 3D Magnetic Resonance Fingerprinting with Quadratic RF
Phase. ISMRM proceedings 2019, p806.
5. Yang M, Ma D, Jiang Y, et al. Low rank approximation
methods for MR fingerprinting with large scale dictionaries. Magn Reson Med
2018 79(4):2392-2400.
6. McGivney
D, Boyacioglu R, Jiang Y, et al. Towards Continuous Tissue Property Resolution
in MR Fingerprinting using a Quadratic Inner Product Model. ISMRM proceedings
2019, p4526.