Holger Eggers1, Christoph Katemann2, and Hendrik Kooijman2
1Philips Research, Hamburg, Germany, 2Philips Healthcare, Hamburg, Germany
Synopsis
Dual-repetition
gradient-echo sequences are currently not widely used for Dixon imaging, because
shorter scan times and better signal-to-noise ratios are usually achieved with
their dual-echo counterparts. In this work, the efficiency of dual-repetition gradient-echo sequences is optimized by introducing
partial echo sampling to extend the acquisition windows and by applying compressed
sensing to reduce scan times. On the example of knee imaging, this is shown to
enable high-resolution water-fat imaging with good image quality in
reasonable scan times.
Introduction
Dixon imaging with bipolar dual-echo gradient-echo
sequences is nowadays routinely included in a broad range of clinical examinations,
for example of the abdomen, pelvis, breast, and head and neck.1 The receive
bandwidth and echo spacing of these sequences typically increase with spatial
resolution, affecting the signal-to-noise ratio (SNR) and water-fat separation. By contrast,
dual-repetition gradient-echo sequences permit choosing the receive bandwidth
and echo shift independent of spatial resolution, which makes them particularly
suited for high-resolution imaging. In the present work, the acquisition duty
cycle of these sequences is enhanced by eliminating the dead times commonly
introduced to implement the echo shifting. Moreover, scan times acceptable for
clinical examinations are demonstrated with these sequences by employing compressed sensing.Methods
A standard
three-dimensional (3D) dual-repetition spoiled gradient-echo sequence is
illustrated in Fig. 1. Alternately, it samples full echoes at two echo times TE1
and TE2 by shifting the rephasing lobes of the readout gradient and
the acquisition windows (AQ). Dead times of ΔTE = TE2 - TE1
are inserted before or after the rephasing lobes to keep the
repetition time (TR) constant. Obviously, they decrease the acquisition duty
cycle, defined as the ratio of AQ duration and TR. To exploit these dead times,
the duration of the rephasing lobes is extended symmetrically by 2 ΔTE and the strength
is scaled accordingly. As shown in Figs. 1 and 2, this leads, for the first
echo at the shorter TE1, to an overlap with the preceding dephasing
lobe with opposite polarity and thus to a reduction in strength or duration of
the dephasing lobe. For the second echo at the longer TE2, an
overlap with the subsequent spoiling lobe with identical polarity occurs. It
requires, in general, a prolongation of the spoiling lobe. The AQ is stretched
by ΔTE, stopping later for the first echo and starting earlier for the second
echo. At the same time, it is shrunk for the second echo, stopping earlier, to
accommodate the longer spoiling lobe. In this way, the proposed sequence
samples only partial echoes, with different partial echo factors PE1
and PE2 at TE1 and TE2, respectively. A
further optimization, illustrated in Figs. 2 and 3, is obtained by adjusting the duration of
the rephasing lobe of the readout gradient and of the AQ for the first echo to
the minimum duration of the dephasing lobe of the readout gradient and the minimum duration of the
phase encoding and slab selection gradients. Usually, this results in smaller PE1 in the periphery of k-space.
Experiments were performed on a 3 T Ingenia scanner
(Philips Healthcare, Best, Netherlands) using a knee coil with eight receive
channels. Data were acquired on phantoms and in volunteers with the standard
and the proposed sequence. Complex single-echo images were individually generated
with a compressed sensing reconstruction and were then
jointly processed with a two-point Dixon method.2-6 Optionally, a
homodyne reconstruction enclosed both.7
Noise measurements were
carried out on phantoms with the radiofrequency excitation and any scan
acceleration disabled, selecting a resolution of
0.5 x 0.5 x 3.0 mm3 to limit scan times. Relative
SNR was predicted by $$rSNR_p=\sqrt{2\left(1-\cos(\theta)\right)/\left(\left(\frac{AQ_F}{AQ_{P_1}}\right)^2+\left(\frac{AQ_F}{AQ_{P_2}}\right)^2\right)},$$
where θ is the
dephasing angle of water and fat over ΔTE, and AQF and AQP
are the AQ durations in the standard and the proposed sequence, respectively. It was compared to the ratio of standard deviations in a region-of-interest in
the complex water image and in the complex single-echo images
obtained with the standard sequence.Results
Predicted and measured relative SNR of six noise measurements are compiled
in Tab. 1. They are in excellent agreement and are consistently higher with the proposed
sequence, with diminishing gain over the standard sequence towards lower
receive bandwidth. Representative examples of knee images are provided in Fig. 4. A field of view of
170 x 170 x 170 mm3 was covered with an
acquired resolution of 0.4 x 0.4 x 2.0 mm3
in 3:25 min in this case. Further parameters of the dual-repetition gradient-echo sequence included a flip angle of 12°, a
TR/TE1/TE2 of 10.0/4.5/5.5 ms, a minimum PE1/PE2
of 0.87/0.80, and a 4-fold scan acceleration.Discussion
In Dixon imaging, the minimum
receive bandwidth is not limited by the fat shift, since the fat shift can be compensated.8 With dual-repetition gradient-echo sequences, it
is not restricted by the robustness of the water-fat separation either. High
SNR efficiency can thus be achieved by choosing longer AQs and exploiting the
resulting longer TRs to increase acquisition duty cycle and steady-state
magnetization, respectively. The proposed partial echo sampling further contributes
to it. Finally, part of the SNR gain can advantageously be spent on higher scan
acceleration to keep scan times reasonable, as previously shown for parallel
imaging.9
Partial sampling of both
echoes allows keeping the fat shift and main field inhomogeneity-induced
distortion the same in both repetitions, facilitating an accurate water-fat separation. The
SNR gain it provides is limited by the required prolongation of the spoiling
gradient towards higher receive bandwidth and by the decreasing ratio of fixed
ΔTE to variable AQ durations towards lower receive bandwidth. Further
improvements in image quality are expected from applying an integrated reconstruction
and water-fat separation.10-11 Acknowledgements
No acknowledgement found.References
1. Eggers H, et al. J Magn
Reson Imaging 2014; 40:251-268. 2. Zhao C, et al. Proc ISMRM 2008; 1478. 3. Wu B,
et al. Proc ISMRM 2008; 1480. 4. King KF. Proc ISMRM 2008; 1488. 5. Liu B, et
al. Proc ISMRM 2008; 3154. 6. Eggers H, et al. Magn Reson Med 2011; 65:96-107. 7.
Reeder SB, et al. Magn Reson Med 2005; 54:586-593. 8. Reeder SB, et al. J Magn
Reson Imaging 2007; 25:644-652. 9. Weiger M, et al. Magn Reson Med 2005;
53:177-185. 10. Doneva M, et al. Magn Reson Med 2010; 64:1749-1759. 11. Ong F,
et al. Magn Reson Med 2018; 80:112-125.