Maria Engel1, Lars Kasper1, Franz Patzig1, Bertram Wilm1,2, Benjamin Dietrich1, Laetitia Vionnet1, and Klaas Paul Pruessmann1
1Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland, 2Skope Magnetic Resonance Technologies, Zurich, Switzerland
Synopsis
In this work, we
show high-resolution stacks of spirals on a tilted hexagonal grid. The new
scheme provides flexibility in balancing readout and scan time, thereby
allowing for high-quality images in a temporal resolution regime suitable for
fMRI. 0.6 mm whole-brain coverage is achieved in below 5.
Introduction
Rapid MR scanning
of 3D volumes is aimed at in applications like fMRI with high spatial and temporal
resolution (1). 3D Fourier
encoding lends
itself to this task: it provides high SNR due to data acquisition from the
entire volume, while being amenable to undersampling by optimal parallel
imaging acceleration in all three dimensions.
The most widespread and obvious encoding strategies
used for this purpose include stacks of
EPIs and spirals(2). Each
k-space plane is covered by one shot of such a 2D trajectory.
However, just like in the 2D case, when aiming for high spatial resolutions, it is not praticable to cover the entire k-space
plane within a single shot. For such a long readout train, local field inhomogeneity would introduce severe
distortions and blurring in the image. Additionally, the broadening of the PSF
due to T2* decay
renders the resolution gain more and more inefficient(3).
This challenge is typically met by covering each
k-space plane by multiple interleaved shots.
However, the overhead
for other sequence modules such as RF excitation and spoiling has to be
repeated for every shot and makes multiplication in the number of shots unattractive
since it entails an increase in the
total scan time. This effect is highly undesirable for time-resolved
applications such as fMRI.
In this work, we
propose to balance readout length and scan-time more flexibly than possible so
far, relying on a tilted hexagonal grid (t-Hex) for fast k-space sampling (4,5).
We show 0.6mm resolved in-vivo whole-brain images using t-Hex spirals.Methods
T-Hex: The tilting of the hexagonal grid (6,7) underlying
a stack of spirals is shown in Figure 1. Each point with coordinates [f,g] in
the oblique coordinate system corresponds to one possibility of tilting the
grid. Its distance to the origin determines the TAQ per shot.
Hardware:
- Philips 7T Achieva system
- 32 channel head array (Nova Medical)
- Field camera
consisting of an array of 16 1H-based NMR field
probes and a dedicated MR acquisition system (8)
Sequence: Spiral k-space trajectories were designed for 0.6x0.6x2 mm
3
resolution and whole-brain coverage (FOV= 24x24x12cm
3), leading to
86 shots, each of which being 38ms long (Figure 2). A GRE sequence with two
different T
2*-contrasts was implemented:
- TE = 15 ms, TR = 56 ms, Total scan time = 4.8s
- TE = 25 ms, TR = 66 ms, Total scan time = 5.7s
Image reconstruction was performed by an iterative cg-SENSE reconstruction (9) extended to 3D, including multi-frequency-interpolation
(10,11) for static off-resonance correction and based on the pre-monitored
trajectories (12). Off-resonance and coil sensitivity
maps were computed from a 3D multi-echo, spin-warp pre-scan (6 echoes, TE 2-7 ms,
2x2x2mm
3 resolution).
Results
In Figure 3, scan time is plotted against TAQ per shot.
Ideally one would aim for moving towards the lower left corner. Spirals stacked
such that forming a hexagonal grid in the cross-section outperform in all
resolution regimes conventional stacks (Cartesian grid). T-Hex provides additional
flexibility in balancing total scan time and TAQ per shot. Figure 4
and 5 show in-vivo results.Discussion
The T-Hex spiral for rapid 3D imaging introduced in this work is tailored
to sampling less k-space volume than one k-space plane per shot. Just like its
kindred t-Hex trajectories (7,4,5) which sample more k-space volume than one k-space plane per shot, it
comprises a uniform undersampling, smooth T2* weighting and high average
k-space speed. However unlike them, it does not need time-consuming blips and
is attractive for the contiguous high-resolution domain. Its main benefit is to
gain flexibility, when balancing readout and scan duration. This is especially
true for a regime, where few shots per plane are suitable. Beyond that, an
integer number of shots per plane is increasingly flexible on its own
To achieve appealing image quality, the reconstruction was based on the
expanded signal model described above. Only dephasing in the ear canals and
lower brain regions compromises those partially. This effect is mitigated for
the shorter TE of 15ms.
The proposed t-Hex trajectory is equally applicable to different
sampling schemes such as EPI or spiral-in. It may furthermore prove useful for
multiband imaging.Acknowledgements
No acknowledgement found.References
1. Feinberg DA, Moeller S, Smith SM, et al. Multiplexed Echo
Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging
Valdes-Sosa PA, editor. PLoS ONE 2010;5:e15710 doi:
10.1371/journal.pone.0015710.
2. Poser BA, Koopmans PJ, Witzel T, Wald LL, Barth M. Three dimensional
echo-planar imaging at 7 Tesla. NeuroImage 2010;51:261–266 doi:
10.1016/j.neuroimage.2010.01.108.
3. Engel M, Kasper L, Barmet C, et al. Single-shot spiral imaging at 7
T. Magn. Reson. Med. 2018;80:1836–1846 doi: 10.1002/mrm.27176.
4. Engel M, Kasper L,
Barmet C, Schmid T, Pruessmann KP. A rapid 3D spiral readout with uniform sampling
density and smooth T2* weighting. In: Proceedings of the ISMRM 2018.
5. Engel M, Kasper L, Wilm B, Dietrich BE, Vionnet L, Pruessmann KP.
T-Hex: Spiral sampling on a tilted hexagonal grid. In: Proceedings of the ISMRM 2019. ; 2019.
6. Mersereau RM. The Processing of Hexagonally Sampled Two-Dimensional
Signals. Proc. IEEE 1979;67:930–949.
7. Engel M, Kasper L, Pruessmann KP. Rapid 3D imaging with multiplanar
spirals. Proc. ISMRM 2017.
8. Dietrich BE, Brunner DO, Wilm BJ, et al. A field camera for MR
sequence monitoring and system analysis. Magn. Reson. Med. 2016;75:1831–1840
doi: 10.1002/mrm.25770.
9. Pruessmann KP, Weiger M, Börnert P, Boesiger P. Advances in
sensitivity encoding with arbitrary k-space trajectories. Magn. Reson. Med.
2001;46:638–651 doi: 10.1002/mrm.1241.
10. Man L-C, Pauly JM, Macovski A. Multifrequency interpolation for fast
off-resonance correction. Magn. Reson. Med. 1997;37:785–792 doi:
10.1002/mrm.1910370523.
11. Barmet C, Tsao J, Pruessmann KP. Sensitivity encoding and B0
inhomogeneity – A simultaneous reconstruction approach. In: Proceedings of the
ISMRM. ; 2005. p. 682.
12. Barmet C, Wilm BJ, Pavan M, et al. Concurrent higher-order field
monitoring for routine head MRI: an integrated heteronuclear setup. In:
Proceedings of the 18th Annual Meeting of ISMRM, Stockholm, Sweden. ; 2010. p.
216