Sebastian Littin1, Feng Jia1, Huijun Yu1, and Maxim Zaitsev1
1Department of Radiology, Medical Physics, University of Freiburg, University Medical Center, Freiburg, Germany
Synopsis
Matrix gradient coils may be used for a non-linear phase preparation to suppress signals in unwanted regions. This allows for a reduction of the FOV smaller than the object without fold-over artifacts and can be seen as a viable alternative to selective excitation.
Introduction
MRI is rather slow
compared to other imaging modalities. This is due to the fact that
the data acquisition is performed consecutively. Each sample
represents a specific harmonic component within the full image.
Therefore, a reduction of the image size below the object size
results in fold-over artifacts.
We recently presented the design
and implementation of a matrix gradient coil [1,2] which enables for
a flexible generation of spatial encoding magnetic fields (SEMs). Nonetheless, due
to Maxwell's equations a SEM cannot be shaped completely freely, e.g. to only produce linear encoding fields within a fraction of a slice and no field otherwise. One approach in lthe literature is to use multi-dimensional excitation, deploying timely excitation patterns. Alternatively, non-linear SEMs may be used to
deliberately dephase MR signals outside the selected region [3]. Suppressed regions
may then be neglected during the image encoding which may be used to speed up the acquisition or enhance the spatial resolution while keeping the same acquisition time.Methods
Non-linear phase
preparation has been presented by Witschey et al. [3]. In that work a quadratic
SEM was used to shift signals out of
the acquisition window. The formalism of local k-space has been used to to
describe the induced signal shifts. The benefit of deploying
quadratic SEMs is that a combination with linearly varying SEMs (constant gradients) enables to induce simple shifts of the area, where no
signal scrambling is taking place throughout the FOV.
Contrary to second order fields, the individual components of the matrix gradient coil we recently
presented [1,2] are rather localized and do not represent spherical
volume harmonics. In addition the element fields vary along all three spatial dimensions. Due to this additional variation along the through-slice
direction, the local k-space formalism is not sufficient to describe
all effects leading to the signal loss. In addition intra-voxel dephasing is crucial for a full understanding of the observed phenomena.
A voxel based model
to describe the signal behaviour in presence of non-linear SEMs was recently
presented [4]. This model is based on the spatial extent of each voxel which is modeled as a boxcar-function. The corresponding Fourier transform is the sinc- function, which enables to described the intra-voxel signal behavior in presence of an non-linear SEM. The combination of signal shifts and intra-voxel dephasing effects enables a complete description of signal loss for arbitrary SEMs.
Phantom and in vivo
measurements were performed on a 3T system, (Siemens Healthineers,
Erlangen, Germany) equipped with 12 additional gradient power
amplifier channels (GPAs) (IECO, Helsinki, Finland). The additional
GPAs are controlled by custom interface hardware. Pulse sequences
were programmed in the open source pulse sequence framework Pulseq
[4,5]. Imaging experiments were performed with a gradient echo sequence with a prephasing time of 6ms and a readout time of 3.2ms. A slice thickness of 4mm, a flip angle of 25° and TE/TR of 10.6ms and 22ms were used.
A single chennel transmit/receive RF coil was used for phantom
experiments. In addition an 8 channel receive coil was used in vivo.
Images were acquired with a base resolution of 256.
The experienced volunteer gave his informed consent and the experiments were performed as self experiments.Results
Phantom experiments
demonstrating the signal signal suppression along one single direction are
depicted in Figure 1. Suppression of MR signals outside the central stripe can be appreciated. The FOV was reduced along with the measurement time to 1/4 of the original size. As expected, a plain reduction of the FOV results in a fold-over artifact, whereas non-linear phase preparation results in an artifact-free image.
The suppression pattern was rotated to
demonstrate the flexibility with the matrix coil. Different orientations are depicted in Figure 2.
Figure 3 shows resulting in vivo images of the reduced field-of-view with and without suppression along with the
original image for reference.Discussion
The flexibility of matrix gradient coils may be used to deliberately spoil signals in unwanted regions. This may be used to reduce the number of the necessary phase encoding steps, which can directly be translated to the measurement time saving during the image encoding. However, the signal-to-noise ratio (SNR) is substantially reduced, since the SNR scales with the square root of the number of acquisition steps. This may be observed in the acquired examples. Therefore reduced FOV imaging is mostly advantageous for SNR-rich time-consuming acquisitions, such as high-resolution 3D imaging.
One benefit of the method is that no additional time is needed during the acquisition since the phase preparation gradients are played out with the prephasing gradients. However, in the present experimental implementation the prephasing time had to be enlarged to allow for strong enough gradient amplitudes.
In theory non-linear phase preparation can be combined with parallel imaging techniques [6]. Yet, the effect of the additional phase on the signal has not been investigated yet in sufficient detail.
Acknowledgements
Deutsche Forschungsgemeinschaft (DFG) - Project number za422/5-1 and za422/6-1References
[1] F Jia, JMR
2017;281:217-228.
[2] S Littin, MRM 2018; 79:1181–1191.
[3] WR Witschey, MRM 2012; 67(6):1620-32.
[4] KJ Layton, MRM 2016;77:1544–1552.
[5] http://pulseq.github.io/
[6] M Zaitsev, MRM 2015; 73:1407–1419.