Anthony N Price1,2, Lucilio Cordero-Grande1,2, Shaihan J Malik1,2, and Joseph V Hajnal1,2
1Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 2Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom
Synopsis
In this work we demonstrate the application of
Fast Interrupted Stead-State (FISS) imaging with 2D Cartesian sampling and
multiband (MB) acceleration to regain scan efficiency. MB-FISS cardiac cine
imaging and a free-running sequence for vessel imaging are shown to work
reliably with good fat suppression.
For cine cardiac imaging FISS with readouts equal to number phase-encode lines
in each cardiac segment had minimal scan time increase, while FISS (n=1) lead to breath-hold
durations increase ~2.5 fold, but benefit from improved fat suppression and no additional
ghosting artefacts compared to bSSFP.
Introduction
The recently proposed fast interrupted steady-state (FISS)
sequence was first demonstrated with 2D radial sampling for cine cardiac
imaging and, combined with ASL, to visualise blood flow1,2. More
recently 3D cardio-respiratory resolved imaging of the heart has been
demonstrated with both radial3 and Cartesian sampling4.
The FISS sequence has been shown to offer similar signal and
contrast to conventional balanced steady-state free precession (bSSFP) for on-resonance
signal, but can also yield a large stop-band in the frequency response which
can be exploited for effective fat suppression, additional benefits include
reduced flow artefacts and off-resonance signal instabilities. However, the
method inevitably increases scan time, by ~2.5x for the case of one readout per
FISS module (n=1, see Fig.1). Although multiple readouts per module have been
used effectively with radial sampling, this can lead to ghosting artefacts in
Cartesian imaging due to signal fluctuation within the FISS module, in addition
to narrowing the usable stop-band for fat suppression.
Here we demonstrate FISS imaging with 2D Cartesian sampling
and use multiband (MB) acceleration to regain scan efficiency in multi-slice
imaging, and apply to a widely-used retrospective gated cine sequence in the
heart, and also to vessel imaging in the leg using a free-running multi-dynamic
sequence. Methods
The FISS technique was implemented along with blipped-CAIPI multiband
as previously described for cine bSSFP5 on a 3T Philips Achieva
system. The FISS sequence employed RF spoiling (117°) between each module, and variable
gradient spoiler duration - in most cases this was kept to the minimum required
for the α/2 slice
selection to avoid extending the shot duration. MB factors up to 4 were tested
with FISS and conventional bSSFP for reference.
For cardiac cine imaging two FISS operating modes were implemented
with a 2D Cartesian retrospective gated sequence: 1) FISS with n=TFE factor
(phase-encode lines per segment), determined by the R-R interval / number of cardiac
phases / TRb - (typically: 14-17), and 2) FISS with n=1 leading to a
reduced TFE factor = R-R/cardiac phases/TRf (typically: 6-7). Standard
parameters for short axis stack analysis were used: flip angle= 40-45°, TRb=2.8-3.6ms, TE=TR/2,
acquired resolution=2x1.6x8mm, 30 cardiac phases (64% minimum acquired before
temporal interpolation, partial-Fourier=80%), in vivo data was acquired from a
healthy adult male volunteer using a 32-channel cardiac coil and
vector-cardiogram gating.
A free-running bSSFP and FISS (n=1) sequence was also tested
on two healthy female volunteers using a Transmit-Receive knee coil (16-channel
receive) and 8-channel extremity coil. Imaging parameters: flip angle=40°, resolution=1.4x1.4x4mm, TRb=3.4,
TE=TR/2, 10 dynamics, MB=1-4.
Multiband reconstruction was performed with custom scripts
(Matlab) and MRecon (GyroTools, Zurich).Results and Discussion
2D Cartesian FISS cine imaging is presented in Fig.2 for both
modes of operating (n=TFE and n=1), alongside a conventional bSSFP scan. For
the TRb=3.4ms case, FISS (n=TFE) has performed well with good fat
suppression and relatively minor ghosting artefacts, whereas for minimum TRb=2.8ms
the ghosts are more prominent at the periphery. In both cases the small
extension of the TFE shot due to extra α/2 tip down and restore
interruptions have only reduced the acquired phase percentage by 5-6% whilst
maintaining the same breath-hold duration. For FISS (n=1) the TFE factor drops
from 14 to 6 and 17 to 7, resulting in a scan time increase from 12 to 27, and
23 seconds, respectively, for each of the TRs shown. However, there is a clear
benefit of improved fat suppression and no issues from ghosting or eddy current
artefacts.
In Fig.3 MB3 accelerated bSSFP and FISS cine data are shown.
Here, due to operating close to the SAR limits, adding multiband has led to a small
increase in TR=3.6ms and with MB FISS (n=1) the minimum TRb=3.3ms leading
to a dark-band encroaching the myocardium in the most apical slice.
Incidentally, it appears the bright signal enhancement in the adjacent blood is
less hyper-intense and more stable in FISS compared to bSSFP. The residual
multiband leakage artefacts in MB-FISS (n=1) are lower than in MB-bSSFP,
probably due to fat saturation.
In Fig.4 we demonstrate multiband accelerated FISS imaging in the leg and
compare to bSSFP acquisitions. Due to the
near 4x increased in the SAR limits for extremity imaging, MB acquisitions can
be acquired with negligible scan time or TR penalty, in addition to the ability
to shim the volume of interest without the presence of the air-tissue
interfaces which make cardiac imaging at 3T challenging - even the on-resonance
band-pass for FISS n=1 is sufficient over the whole volume.
Conclusion
In this work we demonstrate the application of multiband
acceleration to FISS imaging with Cartesian sampling. For cine cardiac imaging
we test two modes of operating FISS: n=1 and n=TFE (determined by the imaging
parameters and heart rate). Results showing good fat suppression can be
achieved with FISS n=TFE even with Cartesian sampling. However, the ghost
artefacts are both parameter and B0 shim dependant. MB-FISS (n=1)
has been shown to work reliably due to the excellent fat suppression and
immunity to ghosting artefacts that can affect n>1, whilst the application
to cardiac imaging at 3T is challenging due to banding artefacts, at lower
field and in other anatomic regions, like the extremities, MB-FISS can be used
without significant penalty. Acknowledgements
This work is supported by the Wellcome EPSRC Centre for
Medical Engineering at Kings College London (WT 203148/Z/16/Z) and by the
National Institute for Health Research (NIHR) Biomedical Research Centre based
at Guy’s and St Thomas’ NHS Foundation Trust and King’s College London. The
views expressed are those of the authors and not necessarily those of the NHS,
the NIHR or the Department of Health.References
1. Koktzoglou.
MRM (2018) 79:2077-2086.
2. Edelman et al. JCMR (2018) 20:12.
3.
Bastiaansen et al. MRM (2020)
83:45-55.
4. Küstner et al. MRM (2019) 82:1617-1630.
5. Price et al. MRM (2019) DOI: 10.1002/mrm.28086