Zhixing Wang1, Steven Allen1, Xue Feng1, John P. Mugler2, and Craig H. Meyer1
1Biomedical Engineering, University of Virginia, CHARLOTTESVILLE, VA, United States, 2Radiology & Medical Imaging, University of Virginia, CHARLOTTESVILLE, VA, United States
Synopsis
2D Cartesian
turbo spin-echo (TSE) is widely used in the clinical neuroimaging, yet the high
specific absorption rate (SAR) induced by a large number of refocusing RF
pulses limits its use in high magnetic field. Thus, this study describes a new TSE sequence with annular spiral ring
acquisitions, dubbed “SPRING TSE”, for fast T2-weigthed imaging and reducing SAR.
Preliminary results show that two sets of high spatial resolution images with
intermediate and strong T2-weighted
contrast characteristics can be obtained by the proposed method within a few seconds.
INTRODUCTION
Spiral trajectories have a number of
benefits for fast imaging [1], including efficient k-space sampling
and robustness to motion artifacts when compared to the conventional Cartesian
sampling. Spiral acquisitions have been incorporated into a 2D TSE signal
generation module via two strategies: an interleaved, rotated spiral-arm
segmentation and an annular ring segmentation. The first requires a double
encoding strategy [2] and a signal demodulation method to mitigate
the swirl-like artifacts due to T2-decay induced signal variation, resulting in
an increased scan time. The annular ring strategy [3, 4] splits long spiral trajectories into
several annular segments, with the benefit of reduced T2 decay artifacts by
converting the T2-dependent signal modulation into a k-space apodizing filter.
However, the latter ring strategy has not been fully explored in either single
excitation or multi-shot acquisitions. In this work, a new technique combining
shared-view acquisition [5] with multi-shot spiral ring TSE is
proposed to obtain high spatial resolution and two sets of T2-weigthed images
within a short scan time. METHODS
A
schematic of the pulse sequence is shown in Figure 1. Spectral-spatial [6]
RF water excitation pulses
were used to suppress lipids, and 180° sinc RF pulses were utilized for refocusing. The first
interleaved, black colored spiral arms were used for field map acquisitions as
in Block [3]
et al. After that, the center of k-space was sampled both at an early echo time
and a late echo time by the first and second spiral rings, respectively, while
the outer k-space regions were sampled by the same outer spiral rings. These
green common echoes were then shared by both the orange and blue echoes, which
allowed data for the short and long TE images to be acquired simultaneously.
Spiral
rings were designed by splitting the original constant density, multi-shot
spiral-out arms into segments of equal time duration, each of which was played
during one of the echoes in the TSE module. A model-based spiral trajectory
measurement [7] was
utilized to reduce the image blur and distortion induced by eddy current
effects and other hardware imperfections, including the anisotropic timing
delays on different physical axes. A simple center frequency offset correction
was used to correct for off-resonance effects. Sequence parameters are given in
Table 1.
All
experiments were performed on a 1.5T scanner (MAGNETOM Avanto, Siemens
Healthcare, Erlangen, Germany) with a 12-channel head coil array. For both
phantom and healthy volunteer studies, multiple slices with 4 mm thickness and
2 mm gap were imaged both by the proposed method and by conventional Cartesian
TSE as a reference.RESULTS AND DISCUSSION
Figure 2
demonstrates the performance of both the spiral ring trajectory correction and
the center frequency offset correction. Without correction, image distortion
and strong off-resonance artifacts are present in the images. With
trajectory correction, edge artifacts (red arrows) are reduced. By further
performing the center frequency correction using the field map data, blurring
artifacts (green arrows) are significantly reduced. Phantom and in-vivo results
are shown in Figures 3 and 4, respectively. In the phantom study, only late
echo images were acquired. In vivo results show that the image quality of
the late echo images from the SPRING TSE method are, in general, comparable to
that of Cartesian TSE, in terms of SNR and image contrast, with the addition of
the early echo images. With three image averages, the total acquisition time for SPRING TSE is 55.8 s, which is still faster than that from Cartesian TSE without any averaging. In
terms of SAR, even though in typical Cartesian TSE, the refocusing RF angle is
reduced to 120°, the relative SAR from the refocusing RF pulses is approximately $$$(180^2 \ast 8)\ / (120^2 \ast 24) \sim75\%$$$ of the SAR of Cartesian TSE. However, signal loss was seen
in some regions where there are strong susceptibility gradients, as shown in
Figure 4, and ghosting artifacts, likely induced by concomitant gradients, were
observed in sagittal and coronal views (not shown). Future work will focus on
improved B0 inhomogeneity correction [8] and concomitant gradient
correction. CONCLUSION
In this study, 2D
T2-weighted brain imaging using spiral ring TSE was implemented and tested. The
results show that within several seconds, the proposed method achieves
15-slice, dual-contrast T2-weighted
images at 0.71 x 0.71 x 4 mm3
spatial resolution per slice and with lower SAR.Acknowledgements
This research was partly supported by
Siemens Medical Solutions USA and by NIH R21
EB022309.References
[1]
Meyer CH, Hu B, Nishimura DG, Macovski A. Fast spiral coronary artery imaging. Magn
Reson Med. 1992;28:202-213.
[2] Li Z, Karis JP, Pipe JG. A
2D spiral turbo-spin-echo technique. Magn Reson Med. 2018;80:1989–1996.
[3] Block W, Pauly J, Nishimura
D. RARE spiral T2-weighted imaging. Magn Reson Med. 1997;37:582–590
[4]. Hennig
J, Menza M, Barghoorn A, Riemenschneider B, Kroboth S, Zaitsev M. Spiral RARE
with annular segmentation. Proceedings 27th Annual Meeting ISMRM, Montreal.
2019:4632.
[5] Johnson B, Fram EK, Drayer
BP, et al. Evaluation of shared-view acquisition using repeated
echoes (SHARE): a dual-echo fast spin-echo MR technique. Am J Neuoradiol
1994;15:667-673.
[6] Meyer CH, Pauly JM, Macovski
A, Nishimura DG. Simultaneous spatial and spectral selective excitation. Magn
Reson Med. 1990;15:287–304.
[7] Tan H, Meyer CH. Estimation of k-space trajectories in
spiral MRI. Magn Reson Med 2009;61:1396–1404.
[8] Chen W, Sica CT, Meyer CH.
Fast conjugate phase image reconstruction based on a Chebyshev approximation to
correct for B0 field inhomogeneity and concomitant gradients. Magn Reson Med.
2008. 60(5): 1103-1111.