Dario Bosch1, Jonas Bause1, Philipp Ehses2, Moritz Zaiss1,3, and Klaus Scheffler1,4
1High Field Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2German Center for Neurodegenerative Diseases, Bonn, Germany, 3Department of Neuroradiology, University Hospital Erlangen, Erlangen, Germany, 4Institute for Biomedical Magnetic Resonance, University Hospital Tuebingen, Tuebingen, Germany
Synopsis
Robust and fast measurements of the transmit field strength is of great importance particularly in parallel transmit applications at ultra-high field. In this work, a pre-saturation TurboFLASH B1+ mapping sequence was optimized for 3-dimensional single-shot acquisition with spiral-centric reordering. Improved SNR and reduced artifacts were achieved by using a variable flip-angle readout. The ability of the proposed sampling scheme to perform fast whole-brain B1+ mapping with low SAR was demonstrated in phantom and in-vivo experiments at 9.4 T. The obtained B1+ maps were comparable to those obtained with the conventional 2D-SatTFL approach but required significantly less scan time.
Introduction
Fast and accurate mapping of transmit-B1 (B1+)
is of great importance for Ultra-High Field (UHF) MRI. The required
time and deposited energy for creating B1+ maps
becomes even more relevant when B1+ maps need
to be recorded for several transmit channels.
The
saturated TurboFLASH method1
(SatTFL) allows for accurate B1+ mapping with a
high dynamic range and low SAR2.
It has, however, only been described with a 2D readout (2D-SatTFL),
which is unsuitable for whole-brain applications due to its poor
temporal efficiency. This work investigated the feasibility of
single-shot SatTFL with a 3D spiral-centric reordering (3D-SatTFL).Methods
In SatTFL two small flip-angle (FA, α) spoiled GRE images are
acquired. The initial longitudinal magnetization (M
Z) in
one of the images is prepared with a high FA (β) presaturation pulse
which reduces the detectable signal by a factor of cos(β). The ratio
between the saturated (sig
sat) and unsaturated (sig
ref)
image intensity permits the calculation of the effective
presaturation FA β:
$$β=acos(\frac{sig_{sat}}{sig_{ref}})$$
However, this technique is limited by signal decay as well as the
number and FA of the excitations in the image acquisition module.
These effects drive both signals (sig
sat, sig
ref)
towards a common steady state without any dependency on β.
Due to the low spatial frequency of B
1+
inhomogeneities and the fact that the largest signal difference
exists at the beginning of the readout modules, a 3D-SatTFL sequence
with spiral-centric readout order was implemented (Fig.1).
All measurements were performed on a 9.4T whole-body MRI scanner
(Siemens Healthineers, Germany), 8 independent transmit channels (pTx
step 2.3), using a custom built head coil (8 transceiver, 8
receive-only elements)
3.
Simulations:
Bloch Simulations to determine the maximum readout duration and
optimal readout FA were performed on a Shepp-Logan-Phantom (T1=2.0s;
matrix=64x64x64, spiral-centric single-shot readout). In order to
reduce signal blurring compared to a readout with constant FA
(3D-SatTFL
const), a variable FA readout module
(3D-SatTFL
var) was investigated.
Phantom
scans:
Phantom scans were performed in a head-shaped agar phantom (T1=2.6s)
using both 3D-SatTFL variants. The resulting FA maps were compared
against the conventional 2D-SatTFL. The B
1+
linearity was evaluated by performing measurement with varying the
saturation pulse voltage using the 3D-SatTFL
const readout.
In
vivo scans:
Whole-brain maps were obtained from two healthy volunteers. B
1+
maps were recorded using 2D-SatTFL, 3D-SatTFL
const, and
3D-SatTFL
var. As in the phantom scans, the B
1+
linearity of the 3D-SatTFL
const was verified. In addition,
individual channel maps (magnitude and phase) were recorded using B
1+
interferometry
4.
Sequence
parameters:-
3D-SatTFL: TR=2.72ms; 3.2 mm isotropic, matrix=64x64x64;
readout-FA=constant (4°) or variable (4°-0°); GRAPPA5 2x2;
elliptical scanning; k-space lines/volume=795; TA=22s/volume.
- 2
D-SatTFL: TR=4.7ms; FOV=205x205mm; 3.2x3.2x3.5 mm³, matrix=64x64,
readout-FA=4°; k-space lines/slice=64; TA=18s/slice.
-
Common parameters: non-selective saturation-FA=90°; recovery time
7.5s; acquisition in circular polarized mode (45° phase
shift/channel).
Results
Simulations:
The simulations of the signal and estimated FA for the single-shot
3D-SatTFL illustrate that the majority of presaturation decays over
the course of the readout scan (Fig.2).
This effect can be partially reduced by using a variable readout FA.
A numeric estimation of the point spread function yielded a blurring
of approximately 30% when using the 3D-SatTFLvar.
Phantom Scans:
The 3D-SatTFL data shows a B1+ pattern similar
to the reference 2D-SatTFL (Fig.3).
However, there are differences in areas with very high and very low
preparation FA. Compared to the 3D-SatTFLconst data, the
3D-TFLvar scan results in a lower deviation from the 2D
map, as well as reduced ringing. Linearity for B1+
is only given for 30°<preparation FA<150°.
In
vivo Scans:
The in vivo scans confirm the similarity between 2D and 3D maps as
well as the linear relation between pulse voltage and measured
presaturation FA (Fig.4a).
The individual channel maps, as they would be required for pTX
pulse-calculation, are depicted in Fig.4b.Discussion
While the variable readout FA reduced artifacts, it also introduced
image blurring and resulted in a lower effective resolution of the
B1+ maps. It may therefore be further improved
by optimizing the sequence with attention to the readout point spread
function. We also observed slight differences between transmit field
maps obtained with the 2D and 3D-SatTFL which likely sourced from
poor slice profiles or inaccuracies in the alignment of the different
data sets. Further research will therefore include comparisons with
other 3D B1+ mapping techniques such as AFI6
and DREAM7
as well as experiments with a multi-shot 3D-SatTFL in order to
investigate the impact of readout blurring further.
One
also has
to keep in mind that a presaturation FA in range of 30°-150° is
required in order to obtain reliable results. However, this is even
possible in case of strong B1+
inhomogeneities at 9.4T if the pre-saturation pulse voltage is chosen
carefully.Conclusion
This work shows that spiral-centric reordered 3D-SatTFL could be used for efficient B1+ mapping at 9.4 T despite the decay of presaturation magnetization during the readout train. The implementation of a variable FA readout allowed for a minimization of imaging artifacts at the cost of slightly increased image blurring. Nevertheless, the presented approach has the capability to perform fast and robust B1+ mapping with low SAR which is essential for pTx pulse calculation, RF coil testing and retrospective B1+ correction at UHF.Acknowledgements
References
-
Chung, Sohae, et al. "Rapid B1+ mapping using a preconditioning
RF pulse with TurboFLASH readout." Magnetic Resonance
in Medicine 64.2 (2010): 439-446.
-
Pohmann, Rolf, and Scheffler, Klaus. "A theoretical and
experimental comparison of different techniques for B1 mapping at
very high fields." NMR in Biomedicine 26.3 (2013):
265-275.
-
Avdievich, Nikolai I., et al. "Combination of surface and
‘vertical’ loop elements improves receive performance of a human
head transceiver array at 9.4 T." NMR in Biomedicine 31.2
(2018): e3878.
-
Brunner, David O., and Pruessmann, Klaas P. "B1+ interferometry
for the calibration of RF transmitter arrays." Magnetic
Resonance in Medicine 61.6 (2009): 1480-1488.
-
Griswold, Mark A., et al. "Generalized autocalibrating partially
parallel acquisitions (GRAPPA)." Magnetic Resonance in
Medicine 47.6 (2002): 1202-1210.
-
Yarnykh, Vasily L. "Actual flip‐angle imaging in the pulsed
steady state: a method for rapid three‐dimensional mapping of the
transmitted radiofrequency field." Magnetic Resonance in
Medicine 57.1 (2007): 192-200.
-
Nehrke, Kay, and Börnert, Peter. "DREAM—a
novel approach for robust, ultrafast, multislice B1 mapping."
Magnetic Resonance in Medicine 68.5 (2012): 1517-1526.