Ole Geldschläger1, Tingting Shao1, Jürgen Herrler2, Armin Nagel3,4, and Anke Henning1,5
1High-field Magnetic Resonance, Max-Planck-Institut for biolog. Cybernetics, Tübingen, Germany, 2Institute of Neuroradiology, University Hospital Erlangen, Erlangen, Germany, 3Institute of Radiology, University Hospital Erlangen, Erlangen, Germany, 4Institute of Medical Physics, Friedrich Alexander University Erlangen Nürnberg, Erlangen, Germany, 5Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, TX, United States
Synopsis
This study investigates universal parallel-transmission
(pTx) radio-frequency-pulses for 3-dimensional local-excitation
designed on different sized databases of B0/B1+-maps from
human heads at 7T. Thus, it prospectively abandons the need for time-consuming
subject specific B0/B1+-mapping and pTx-pulse calculation
during the scan session.
For the proposed calculation routine,
the design-database does not need to include more than five heads, to achieve a
pTx-pulse that excites the same 3-dimensional local-excitation target-pattern on
the tested 40 different heads. The resulting universal pulses created
magnetization-profiles with (in most cases) an only marginally worse Normalized-Root-Mean-Square-Error compared to the magnetization-profiles produced by pulses tailored to
individual heads.
Introduction
Parallel-transmission (pTx) radio-frequency (RF) systems
offer additional degrees of freedom for designing tailored RF-pulses (TPs).
However, pTx-pulse-design often includes measuring the B0 and
channel-wise B1+field distribution for each subject. In addition,
the pulse calculation is currently performed during the scan-session. This
procedure leads to lengthy scan times and prevents clinical and research-applications
of pTx.
Gras et al.1,2,3
introduced the concept of ‘universal’ pTx-pulses (UPs), which does not require
the knowledge of B0-variations and channel-wise B1+
distribution in individual subjects, and rather relies on a pre-calculated
pulse database. However, these studies only considered the case of homogenous whole-brain excitation pulses. In own previous studies4,5, we extend the UP
concept to 2- and 3-dimensional local excitation, based on a relative small
database of B0/B1+-maps. In this work, we investigate UPs
for 3-dimensional (3D) local-excitation-pulses optimized on different sized
(one to eight included heads) databases of B0/B1+-maps.
Furthermore, the calculated UPs will be tested on 32 B0/B1+-maps
that were not included in the design-database.Methods
For 40 different heads of healthy volunteers the B0-
and the B1+-maps from each transmit channel of an 8-channel RF-coil
were measured using a 7T whole body MRI scanner (Siemens Terra).
Eight different sized databases of B0/B1+-maps
were created, based on which the UPs were calculated. The first tested database
contained the B0/B1+-map from ‘head one’ (DB1), the
second database contained two B0/B1+-maps; from ‘head one’
and from ‘head two’ (DB2), etc. The eighth database contained the eight B0/B1+-maps
from ‘head one’ to ‘head eight’ (DB8). In advance, all maps were sorted randomly.
The goal was to design single UPs that would consistently
excite the same 3D local pattern with a desired flip angle (FA) on all heads
that are in the database, respectively. The UP designed based on DB1 is
referred to UP-DB1; the UP designed based on DB2 is referred to UP-DB2; etc.
Two 3D-local-excitation test-target-patterns
where chosen: A cylinder reaching from the top to the bottom of each head with
a target FA of 90° (FA90) (see left column in Figure
3) and a cuboid in the center of each head with a target
FA of 10° (FA10) (see left column in Figure
4).
After calculation, the UPs were
applied on all of the 40 heads, respectively. The performances of the UPs were compared
with the performances of the TPs by calculating the
Normalized-Root-Mean-Square-Error (NRMSE) between the resulting magnetization
profiles and the target-patterns.
To calculate the UPs, the method from Geldschläger et al.5 was used. The idea
of this approach is basically to extend the spatial domain method6 to include more than one subject for which the pulse will be optimized.
To calculate the TPs, the spatial information from only the
corresponding subject will be used as the design-database.
3D-stack-spiral-in k-space-trajectories
and a pulse-duration of 2.21ms (8.02ms) for the cylinder (cuboid), were chosen.
T1- and T2-relaxation times were
not considered during this 7T simulation-study.Results
As depicted in Figure 1 and Figure 2, the NRMEs of the UPs
decrease with increasing number of heads in the design-database for both
target-patterns. For database sizes ≥5 the performances of
the UPs are almost identical on each head. For the sake of clarity, the results
from the UP-DB7 and UP-DB8 are not shown. They differ negligibly from the
results of UP-DB5 and UP-DB6. Except of some heads for the cuboid-target
pattern, the TPs outperform every UP. The different scales of the y-axis in Figure 1 and Figure 2 is worth noting.
In Figure 3 and Figure 4, the
magnetization-profiles after application of the TPs and UP-DB5 for both
targets, respectively, are highly similar. The same is valid for the results in
Figure 5, where UP-DB5 creates
similar profiles on eight example heads, which were not in the design-database
included.Discussion
Except of on head 38, the UPs with a
database size ≥5 perform almost
identical on each head, respectively. Possibly, head 38 exhibits anatomical
abnormalities or measurement errors occur during the B0/B1+-map
acquisition.
Due to the lower target FA10 for the
cuboid target-pattern compared to the FA90 for the cylinder target-pattern, the
excitation (non-excitation) of a voxel that is not excited (excited) in the
target has not that strong influence on the NRMSE. That results in lower NRMSEs
for the cuboid target-pattern as for the cylinder target-pattern. However, for
both targets, the performances of UPs with database size ≥5 are excellent, as they perform marginal worse compared to the TPs. That is
valid for heads that were in the design-database included, as well as on heads
that were not included.Conclusion
With the proposed UP-design routine, not
more than five database heads are necessary to create reliably pulses that achieve
the same 3D-local-excitation pattern with a desired FA across a large cohort of
subject heads. As this is especially true for heads that were not the database
inherent, the time-consuming per subject B0/B1+-map-measurements
can potentially be eliminated from the pTx-scan-sessions in future. This
significantly reduces the scan time, abandons the need of subject specific
pulse optimization while the subject is waiting in the scanner and might open
up new possibilities in the field of pTx-pulse-design methods.Acknowledgements
Funding by the European Union (ERC Starting Grant, SYNAPLAST
MR, Grant Number: 679927) is gratefully acknowledged.References
1. Gras V, Vignaud A,
Amadon A, Bihan DL, Boulant N. Universal pulses: A new concept for
calibration-free parallel transmission. Magnetic Resonance in Medicine. 2016
Feb; 77: 635-643.
2. Gras V, Mauconduit F,
Vignaud A, Amadon A, Bihan DL, Stöcker T, et al. Design of universal
parallel-transmit refocusing kT -point pulses and application to 3D T2
-weighted imaging at 7T. Magnetic Resonance in Medicine. 2017 Nov; 80: 53-65.
3. Gras V, Pracht ED,
Mauconduit F, Bihan DL, Stöcker T, Boulant N. Robust nonadiabatic T2
preparation using universal parallel-transmit kT -point pulses for 3D FLAIR
imaging at 7 T. Magnetic Resonance in Medicine. 2019 Jan; 81: 3202-3208.
4. Geldschläger O, Shao
T, Henning A. Universal Parallel Transmit Pulse Design for Local Excitation.
In Proceedings of Joint Annual Meeting ISMRM-ESMRMB 2018, Paris, France.;
2018.
5. Geldschläger O, Shao
T, Henning A. Universal Parallel Transmit Pulse Design for 3-Dimensional
Local-Excitation – A 9.4T Simulation Study. In Proceedings of Annual Meeting
27th ISMRM, Montreal, Canada; 2019.
6. Grissom W, Yip Cy, Zhang Z, Stenger VA, Fessler JA, Noll
DC. Spatial domain method for the
design of RF pulses in multicoil parallel excitation. Magnetic Resonance in
Medicine. 2006; 56: 620-629.