Pallavi Bohidar1, Hongwei Sun2, Jonathan C. Sharp2, and Gordon E. Sarty1
1Division of Biomedical Engineering, University of Saskatchewan, Saskatoon, SK, Canada, 2Department of Oncology, University of Alberta, Edmonton, AB, Canada
Synopsis
Transmit Array Spatial Encoding (TRASE) is a novel MRI technique that achieves spatial encoding by introducing phase gradients in the transmit RF magnetic field (B1). In this study, Bloch simulations were performed to investigate and study the effects of B1 field perturbations arising from inductive coupling among RF coils for 2D TRASE imaging. Simulations show that a flip angle contribution of ~95% or higher from the primary (driven) transmit coil is required for 2D TRASE MRI. This result is of crucial importance for designers of practical TRASE transmit array systems.
Introduction
Transmit Array Spatial Encoding (TRASE) is a spatial encoding
technique in Magnetic Resonance Imaging (MRI) that uses phase
gradients of the transmit radio-frequency (RF) B1 field1. TRASE is
useful for spaceflight MRI systems since it no
longer requires the use of complex B0 gradient coils, and can provide
resolution at par with standard MRI2.
TRASE relies on B1 fields of different phase gradients for k-space traversal. A TRASE pulse sequence uses RF pulses that are produced by switching between the transmit coils. However, field interactions among the RF coils can cause un-driven coils to produce unwanted B1 fields that impair the spatial encoding. Therefore, TRASE is sensitive to B1 field perturbations arising from
inductive coupling between B1 transmit coils. Any B1 field isolation (coil
decoupling) technique will require a clear understanding of the effects of the B1 field interactions that have never been studied for TRASE imaging.
The purpose of this study was to investigate the effects of B1 field
coupling using Bloch3 simulations and determine the acceptable level of field interactions for 2D TRASE imaging. Methods
Biot-Savart calculations (using the BSmag4 toolbox in Matlab) were
used to generate B1 field maps to study the simplest 2D TRASE transmit coil setup
consisting of 3 coils (1 uniform saddle coil and 2 phase gradient twisted solenoid5 coils). Various weighting factors, representing
flip angle contributions from the secondary (un-driven) coils for each
transmit pulse, were used to add the B1 field components of primary and
secondary coil geometries to simulate coil coupling.
The coupled B1
field maps were then used in a Bloch simulation program to perform
2D TRASE MRI simulations. The reconstructed 2D TRASE images from our simulations of varying levels of coil coupling were compared to that of an ideal case (i.e. no coupling between the RF transmit coils) by simple visual inspections and also by computing metrics such as PSNR (peak-signal-to-noise-ratio) and SSIM (structural-similarity-index)6.Results
Figure 1 shows the contour maps for both coupled and un-coupled B
1 fields of all the three transmit coil geometries considered for 2D TRASE.
Some key observations about the effects of interacting B
1 fields on 2D
TRASE imaging are as noted below:
-
Reduced phase gradient strength - impairs spatial resolution.
-
A spatial shift in the phase gradient - causes a shifted image.
-
Increased non-linearity of the phase gradients - causes object shape
distortion in the image.
-
Increased non-orthogonality of the phase gradients - impairs spatial
encoding direction.
- A weak phase gradient induced in the uniform coil (that is supposed
to have no phase gradient) - impairs spatial encoding.
-
Increased |B1| inhomogeneity - causes large flip angle variation over
the FOV.
All the above-mentioned factors contribute to image distortions that
increase with an increase in coupling (see Figures 2 and 3).
A visual inspection of the above mentioned results in Figure 2 showed that a flip angle contribution of ~95% or higher is needed from the primary coil for image formation in 2D TRASE using the transmit coil setup described above.
Conclusion
Our results show that a flip angle contribution of ~95% or higher
from the primary transmit coil is required for 2D TRASE MRI, using a 3-coil setup. This
is a significant step in determining the acceptable levels of coil
interactions that is conducive to 2D TRASE imaging. This permissible limit of B1 field interactions will be applicable to any 2D TRASE transmit coil setup of 3 coils.
Currently, geometric decoupling is used between the
transmit coils for 1D TRASE imaging7. If the B1 field isolation
achieved from geometric decoupling alone, for any 2D TRASE transmit RF system,
is not as good as the required permissible limit (as determined from this
study), then alternate or additional means, such as the use of pin-diode switches, need to be employed to
increase the isolation up to the required level. Therefore, the results of this study are important for designers of practical TRASE transmit array systems.Acknowledgements
We acknowledge the funding provided by the Canadian Space Agency (CSA) FAST - Zero-G MRI Project grant to GES (Principal Investigator)
titled “Suborbital Development and Testing of a Space Flight Astronaut Magnetic Resonance Imager”.
References
- J. C. Sharp and S. B. King, “MRI using radiofrequency magnetic field phase gradients,” Magnetic resonance in
medicine, vol. 63, pp. 151-161, 2010.
- P. Bohidar, H. Sun, G. E. Sarty, J. C. Sharp, “TRASE 1D sequence performance in imperfect B1 fields,” Journal of
Magnetic Resonance, vol. 305, pp. 77-88, 2019.
- F. Bloch, “Nuclear induction,” Physical review, vol. 70, p. 460, 1946.
- L. Quéval, “BSmag toolbox user manual,” Technical Report, Department of Electrical engineering, University of
Applied Sciences Düsseldorf, Germany, April 2015. Available: http://www. lqueval.com
- H Sun, S Yong, J C. Sharp, “The twisted solenoid RF phase gradient transmit coil for TRASE imaging,” Journal of
Magnetic Resonance, vol. 299, pp. 135– 150, 2019.
- Z. Wang, A. C. Bovik, H. R. Sheikh, E. P. Simoncelli, “Image quality assessment: from error visibility to structural similarity,” IEEE transactions on image processing, vol. 13.4, pp. 600-612, 2004.
- H. Sun, A. AlZubaidi, A. R. Purchase, J. C. Sharp, “A geometrically decoupled, twisted solenoid single‐axis gradient
coil set for TRASE,” Magnetic resonance in medicine, pp. 1-15, 2019.