Emre Kopanoglu1, Cem M. Deniz2, and Richard G. Wise1,3
1CUBRIC, School of Psychology, Cardiff University, Cardiff, United Kingdom, 2Department of Radiology, New York University Langone Health, New York, NY, United States, 3Institute for Advanced Biomedical Technologies, University of Chieti-Pescara, Chieti, Italy
Synopsis
This study investigates the effect of within-scan patient
motion on local-SAR for simultaneous multi-slice (SMS) imaging at 7T. A virtual
body model was simulated at 104 different positions. 1-/2-/3-spokes pulses were
designed to excite a region of 60 slices covering the cerebellum and the brain,
using SMS-factors of 1 through 5. Local-SAR was observed to increase by up to 2.75-fold
due to patient motion. Pulses with higher SMS-factors were up to 50% less
sensitive against changes in local-SAR due to patient motion, compared to SMS:1
pulses. Pulses with higher SMS-factors yielded more consistent local-SAR
throughout the scan.
Purpose
To investigate the effect of within-scan patient motion on
the local specific absorption (SAR) rate for simultaneous multi-slice imaging
at 7T.Introduction
Simultaneous multi-slice (SMS) imaging has been
increasingly used in brain imaging 1. At ultrahigh field (UHF),
SMS is combined with parallel transmit (pTx) pulses to provide uniform excitations in
multiple slices/slabs while reducing the total scan time through the
acceleration provided by SMS 2,3. To control local-SAR, computational
models are commonly used in pTx pulse design for UHF imaging 4.
For single-slice pTx pulse optimization, it was shown that
within-scan patient motion can cause local-SAR to increase by up to 3.1-fold 5. As the utilization of
SMS imaging with pTx pulses is expected to grow in the future, in this study,
we investigate changes in local-SAR due to within-scan patient motion when SMS
pTx pulses are employed. Methods
The virtual body model Ella (IT’IS Foundation, Zurich, CH) 6 was simulated in
Sim4Life (Zurich MedTech,Zurich, CH) at 104 different relative positions inside
a generic 8-channel parallel transmit array (Figure 1). These positions covered
all six degrees-of-freedom of motion as well as off-axis motion on axial and
coronal planes. Translations of 1,2,5,10,15,20 mm along right or inferior, 1,2,5,10
mm along posterior (range limited due to geometrical constraints); and
rotations of ±1,±2,±5,10,15,20 degrees in pitch, 1,2,5,10,15,20 degrees in
roll, and ±1,±2,±5,±10,±15,±20 degrees in yaw were simulated 5. The body model
consisted of 73 different tissues and the shoulders were included in the
computation domain 7. The vectorized body
model was discretized at an isotropic resolution of 2 mm.
Using the simulated fields at the centred position, 1-spoke
(RF-shimming), 2-spokes and 3-spokes slice selective RF pulses were designed to
excite 60 axial slices (slice-thickness: 1.8 mm) that span the cerebellum and
the brain (Figure 1c). The cost function was defined as the sum of normalized
root-mean-squared error in the magnitude excitation profile (to enable phase
relaxation) and the RF power, the latter regularized by a Tikhonov parameter 8. The Tikhonov parameter
was determined using an l-curve analysis. Spoke locations that minimize the
cost function were individually selected for each designed pulse using the matching
pursuit-guided-conjugate gradient method 9. Pulses were designed
for interleaved simultaneous multi-slice (SMS) factors of 1 through 5 (e.g.,
first SMS:3 pulse excited slices 1-21-41).
A total of 411
pulses were designed (60,30,20,15,12 for SMS:1,2,3,4,5, respectively, for
1-/2-/3-spokes). The sensitivity of each pulses to patient motion was
investigated by comparing their local-SAR at the off-centre positions to their local-SAR
values at the centred positions. The ratio of the off-centre peak local-SAR
values to the values at the centre are reported. Regions with higher local-SAR at the off-centre position than the peak
local-SAR at the centred position are also compared for SMS:1 and SMS:5 cases
for the worst-case positions for 2-spokes pulses. Pulses were also compared in terms of how the peak local-SAR changes throughout
the scan.Results
SMS:1 pulses yielded a wide range of peak local-SAR values
across the slices. For all pulses (1-/2-/3-spokes), increasing the SMS-factor reduced
the variation of peak local-SAR across different slices/slice-groups and led to
more consistent instantaneous peak local-SAR throughout the scan (Figure 2).
Increasing the SMS-factor reduced the sensitivity of
local-SAR to patient motion. Patient motion led to up to 2.75-fold increase in
peak local-SAR for 2-spokes SMS:1 pulses (Figure 3). Increasing the SMS-factor
to 3 and above reduced the sensitivity to motion approximately by half as the
maximum increase in peak local-SAR was around 1.4-fold for the higher
SMS-factors (Figure 3). A similar behaviour was observed for 3-spokes pulses as
well, with SMS:4 and SMS:5 reducing motion sensitivity by 50% compared to SMS:1 by reducing
the maximum peak local-SAR increase from 2.4-fold to 1.2-fold (Figure 4). The increase in
local-SAR for 1-spoke pulses reduced gradually from 2.6-fold to 2.2-fold as the
SMS-factor was increased (Figure 4).
Increasing the SMS-factor also reduced the size of the
region exposed to higher local-SAR than estimated at the centred position. Figure
5 compares the SMS:1 and SMS:5 cases for the 2-spokes pulse at the worst-case
position. Compared to the peak local-SAR at the centre, the SMS:1 case yielded
a maximum of 2.75-fold increase with a volume of 264 cm3 exposed to
higher SAR. For the SMS:5 case, the volume exposed to higher SAR reduced by 88%
to 32 cm3, and the maximum peak local-SAR increase was 39%.Discussion
We have investigated the effect of patient motion on local-SAR
for simultaneous multi-slice excitation at 7T. The results show that pulses
designed with higher SMS-factors are more robust against local SAR changes due
to patient motion when compared to pulses with lower SMS-factors. Moreover,
increasing the SMS-factor reduced the variation of peak local-SAR throughout
the scan when patient motion was not observed.
The local-SAR increased between 1.2-fold and 2.75-fold across all cases investigated. To ensure patient safety, the effect of
motion should be taken into consideration in pulse design, especially for
patient populations that may not stay still for extended durations 10,11.Acknowledgements
No acknowledgement found.References
1. Feinberg DA, Setsompop K. Ultra-fast
MRI of the human brain with simultaneous multi-slice imaging. J Magn Reson
2013;229(0):90-100.
2. Wu X, Schmitter S,
Auerbach EJ, Moeller S, Ugurbil K, Van de Moortele PF. Simultaneous multislice
multiband parallel radiofrequency excitation with independent slice-specific
transmit B1 homogenization. Magn Reson Med 2013;70(3):630-638.
3. Poser BA, Anderson
RJ, Guerin B, Setsompop K, Deng W, Mareyam A, Serano P, Wald LL, Stenger VA.
Simultaneous multislice excitation by parallel transmission. Magn Reson Med
2014;71(4):1416-1427.
4. Guerin B,
Setsompop K, Ye H, Poser BA, Stenger AV, Wald LL. Design of parallel
transmission pulses for simultaneous multislice with explicit control for peak
power and local specific absorption rate. Magn Reson Med 2015;73(5):1946-1953.
5. Kopanoglu E,
Plumley AJD, Erturk A, Deniz CM, Wise RG. Implications of within-scan patient
head motion on B1+ homogeneity and Specific Absorption Rate at 7T. Proc. ISMRM
2019; 2019; Montreal, Canada. p 4686. (Proc. ISMRM 2019).
6. Andreas C,
Wolfgang K, Eckhart GH, Katharina H, Marcel Z, Esra N, Wolfgang R, Rolf J,
Werner B, Ji C, Berthold K, Peter S, Hans-Peter H, Jianxiang S, Michael O, Dominik
S, Anthony K, Joshua WG, Niels K. The Virtual Family—development of
surface-based anatomical models of two adults and two children for dosimetric
simulations. Physics in Medicine & Biology 2010;55(2):N23.
7. Wolf S, Diehl D,
Gebhardt M, Mallow J, Speck O. SAR simulations for high-field MRI: how much
detail, effort, and accuracy is needed? Magn Reson Med 2013;69(4):1157-1168.
8. Grissom W, Yip CY,
Zhang Z, Stenger VA, Fessler JA, Noll DC. Spatial domain method for the design
of RF pulses in multicoil parallel excitation. Magn Reson Med
2006;56(3):620-629.
9. Kopanoglu E,
Constable RT. Radiofrequency pulse design using nonlinear gradient magnetic
fields. Magn Reson Med 2015;74(3):826-839.
10. Malviya S,
Voepel-Lewis T, Eldevik OP, Rockwell DT, Wong JH, Tait AR. Sedation and general
anaesthesia in children undergoing MRI and CT: adverse events and outcomes. Br
J Anaesth 2000;84(6):743-748.
11. Prasher V, Cumella
S, Natarajan K, Rolfe E, Shah S, Haque MS. Magnetic resonance imaging, Down's
syndrome and Alzheimer's disease: research and clinical implications. J
Intellect Disabil Res 2003;47(Pt 2):90-100.