Yan Tong1, Peter Jezzard1, Caitlin O'Brien1, and William T Clarke1
1Wellcome Centre for Integrative Neuroimaging, FMRIB Division, NDCN, University of Oxford, Oxford, United Kingdom
Synopsis
T2‐relaxation under‐spin‐tagging (TRUST)
is a robust spin tagging based method to quantify oxygen extraction fraction
(OEF), but it lacks spatial specificity. Recently O’Brien et al. proposed a
method involving multiple saturation pulses to achieve spatial specificity. Parallel
transmission (pTx) provides additional degrees of freedom for spatial
localisation. A pTx RF pulse design strategy based on a shells trajectory is applied
to perform regional OEF measurement at 7 T. Initial in-vivo results
acquired from one healthy subject showed that spatial localisation of OEF could
be achieved.
Introduction
Oxygen extraction fraction
(OEF) is an important parameter for the understanding of normal and abnormal physiology
in the brain. Lu and Ge1 introduced an MR-based method named T2‐relaxation-under‐spin‐tagging (TRUST)
to measure OEF by labelling spins in the venous blood and performing pair-wise subtraction
in the sagittal sinus. OEF is then inferred based on the known relationship
between blood T2 and blood oxygenation levels1. Although
proven to be a robust non-invasive technique to measure cerebral OEF, TRUST has
one key limitation: it does not provide spatial specificity. O’Brien et al. applied
a train of water suppression enhanced through T1 effects (WET) saturation
pulses to measure regional OEF using selective localised-TRUST (SL-TRUST) at 3 T2. Parallel transmission
(pTx) allows different RF waveforms to be played out in different transmit
channels, thus providing more degrees of freedom for spatial selectivity. In
this work, we present a single pTx pulse design strategy to measure regional
OEF in the brain at 7 T. Methods
B1+ and
B0 maps were acquired from a single subject (25 years, male) on a
Siemens (Erlangen, Germany) Magnetom 7 T scanner equipped with an 8 channel pTx
system, and a Nova Medical Inc. (Wilmington, MA, USA) 8Tx/32Rx head coil. B1+ maps were acquired using a 2D “STE first” phase-cycled DREAM sequence3,4. B0 maps were
acquired using a dual-echo gradient-echo sequence. A brain mask was generated from the raw images in the B0 mapping sequence using BET5.
The labelling strategy consists
of the following steps. First, a spatially selective pTx pulse excites a 3D
shape that excludes the region-of-interest (ROI), which is then saturated by a spoiler
gradient. Second, a slab-selective inversion pulse is played out to invert the
remaining longitudinal magnetisation in the ROI. In the control case, the same
spatially selective pulse is used, but without a subsequent inversion pulse,
therefore creating a signal difference solely from the venous blood in the ROI
(see Fig. 1 and Fig. 2). The T2 encoding is moved to the front of
the sequence, unlike the conventional TRUST where it is at the end.
A shells trajectory
proposed by Davids et al. was used for the pulse design6 (see Fig. 3). No gradient
trajectory optimisation was performed in this study. A cylindrical ROI with a height
of 40 mm was chose. The pulse design employed the variable-exchange algorithm,
which was based on a method proposed by Dupas et al.7. The regularisation constant λ values were
chosen heuristically to produce acceptable results in FA with a feasible RF
voltage. The
optimisation algorithm mentioned above was implemented in MATLAB (The
MathWorks, Natick, MA, USA)8. The designed RF pulse
was then incorporated into the TRUST sequence to evaluate its performance. T2
encoding, pre-saturation and double inversion pulses were turned off to isolate
the effects proposed RF pulse. The inversion time was set to lowest possible
value of 73ms to minimise relaxation effects. Five slices in the were acquired
using an EPI readout. Results
Simulation
results were shown in Fig 4. A root mean square error (RMSE) of 6.71% was achieved.
Larger errors occurred in the ROI compared to the rest of the slices.
Raw EPI images were masked and shown in Fig 5. A circular pattern could be observed
in each slice, but the boundary of such a pattern varied across slices. Moreover,
saturation of signal outside of the ROI was not achieved with great accuracy. The sum of pixel values in
the ROI accounted for an average of 27.1% of the sum of pixel values in the
whole masked brain slice. Discussions
Higher intrinsic SNR at 7T
is beneficial to the TRUST method, enabling OEF measurements of smaller regions. However, spatial selection
based on multiple WET saturation pulses might not be suitable for 7T, as the
saturation performance may be impaired due to B1+
inhomogeneity. In this abstract, we introduced a single tailored pTx RF pulse
design to achieve spatial localisation in the presence of B1+
inhomogeneity.
Although
the magnetisation was not controlled beyond the slab of interest, OEF measurements
should not be affected since the spatially selective pulse was applied in both
the label and control conditions, meaning any errors should subtract.
Simulation results showed that spatial localisation could be achieved with 6.71% RMSE error. However, in-vivo
results showed that although the ROI was clearly identified, saturation of
signal outside of the ROI was not achieved with great accuracy. Imperfect timing
synchronisation between the RFPA and gradient system or inaccurate playout of
the prescribed complex gradient waveforms might explain the pulse performance. Calibrating
these imperfections has been shown to improve the accuracy of pTx pulses9.Conclusion
A pTx RF pulse design is
presented for regional OEF measurement at 7 T. Initial in-vivo results showed
the feasibility of the method, but RF/gradient system imperfections need to be
corrected for complete saturation of tissue outside the selected volume.Acknowledgements
The authors acknowledge the China Oxford Scholarship Fund, Royal
Academy of Engineering, Dunhill Medical Trust, University College Oxford and
the NIHR Oxford Biomedical Research Centre. The Wellcome Centre for Integrative
Neuroimaging is supported by the Wellcome Trust (203139/Z/16/Z). The support of
the UK Medical Research Council UK7T Network is acknowledged (MR/N008537/1). The support of support from the Engineering and Physical
Sciences Research Council (EPSRC) and Medical Research
Council (MRC) Centre for Doctoral Training in Biomedical
Imaging (grant EP/L016052/1) is also acknowledged.References
1. Lu, H.
and Y. Ge, Quantitative evaluation of
oxygenation in venous vessels using T2-Relaxation-Under-Spin-Tagging MRI.
Magn Reson Med, 2008. 60(2): p.
357-63.
2. O'Brien,
C., et al., Volume-localized measurement
of oxygen extraction fraction in the brain using MRI. Magn Reson Med, 2019.
82(4): p. 1412-1423.
3. Nehrke,
K. and P. Bornert, DREAM--a novel
approach for robust, ultrafast, multislice B(1) mapping. Magn Reson Med,
2012. 68(5): p. 1517-26.
4. Tse,
D.H., et al., Encoding methods for B1(+)
mapping in parallel transmit systems at ultra high field. J Magn Reson,
2014. 245: p. 125-32.
5. Smith,
S.M., Fast robust automated brain
extraction. Hum Brain Mapp, 2002. 17(3):
p. 143-55.
6. Davids,
M., et al., Fast three-dimensional inner
volume excitations using parallel transmission and optimized k-space
trajectories. Magn Reson Med, 2016. 76(4):
p. 1170-82.
7. Dupas,
L., et al., Two-spoke placement
optimization under explicit specific absorption rate and power constraints in
parallel transmission at ultra-high field. J Magn Reson, 2015. 255: p. 59-67.
8. Constrained Nonlinear Optimization Algorithms. [cited 2017; Available from: https://uk.mathworks.com/help/optim/ug/constrained-nonlinear-optimization-algorithms.html#brnox01.] 9. Cavusoglu,
M., et al., Correction of parallel
transmission using concurrent RF and gradient field monitoring. MAGMA, 2017.
30(5): p. 473-488.