Yang Xia1, Farid Badar1, and Simon Miller1
1Physics, Oakland University, Rochester, MI, United States
Synopsis
Imperfect compensation for the eddy currents can cause artifacts in UTE
images. Several post-acquisition processing methods can reduce artifacts in the
images that are acquired in the radial k-space trajectories.
Introduction
Very short T2 relaxation times are common in biological tissues and porous
media, which make them invisible by conventional spin-echo or gradient-echo
based MRI. Reduction of the echo time (TE) in MRI can be achieved in Ultra-short
echo (UTE) sequences. However, shortening the TE generally makes the signals
susceptible to the residual eddy currents and other imperfections in the
imaging system, which manifests itself as artifacts in the UTE images. In this study, we aimed to reduce the
artifacts seen in the UTE images using various post-acquisition image processing
techniques (regridding algorithm, oversampling, deapodization)1-3.Methods
A
Bruker AVANCE II micro-imager (Billerica, MA) with a 7T/9cm vertical-bore
magnet was used to image several glass phantoms that contained doped water (0.1
and 1% CuSO4). Two imaging sequences were used, a 2D UTE sequence
and a multi-slice-multi-echo (MSME) sequence (in the single-slice single-echo
mode). The MSME sequence had a TR of 2000 ms, TE of ~ 8 ms, and a matrix size of
256x256 with 25 mm field of view. The UTE sequence had the same imaging setting
except for a TR of 100 ms and a minimum TE of 0.27 ms. The gradient trajectory
was pre-measured on a glass sphere of water. A number of Matlab (Natick, MA)
routines were developed to reconstruct the images from the same UTE raw data. The
images were examined using ImageJ (NIH, MD) where similar ROIs were
selected for comparative profiles.Results
Fig
1 shows the MSME and UTE images from (a) a single tube and (b) a multi-tube
phantom. The single tube (a) was small in size when comparing with the size of
the rf coil (a 25 mm resonator); however, the edge of the tube had an enhanced
intensity than the center of the tube. Since the outside diameter of the
multi-tube phantom (b) was similar to the size of the rf resonator, some
inhomogeneity in the peripheral of the image was expected. However, the UTE
image of the multi-tubes had much severe non-uniformity in the center part of
the tubes.
Fig 2 shows the effects of
the various post-acquisition
reconstruction methods: (a) MSME image (acquired using Cartesian sampling in k
space) and (b-e) UTE images from the same data set (acquired using radial
sampling in k space). It is clear that the image produced by re-gridding algorithm
(c), which re-grids the radial data to a Cartesian grid before a 2D FFT, had
better uniformity than the image by the commercial reconstruction (b). Several
additional approaches were incorporated into the re-gridding algorithm,
including (d) oversampling, which interpolates on to a 512 x 512, twice-fine Cartesian
grid from the radial samples, and (e) de-apodization,
which applies a filter in the image domain to the oversampling reconstruction. Oversampling
is used to eliminate wraparound artifacts that are present in larger images.
The use of de-apodization is motivated by typical re-gridding methods. The applications
of oversampling and de-apodization (d, e) returned the images that appeared similar
to the commercial algorithm (b). Since the amount of the improvement depends
upon the individual data sets (i.e., the actual trajectory and data acquired),
a min artifact in the UTE images could be achieved by iterations in image analyses.
Discussion
Reduction
of TE in MRI increases artifacts in the UTE images. In this configuration of
the Bruker MRI hardware, the electric current for B0 compensation was
added to the shim coils analogically, which is described by 3 exponential
functions (each with an amplitude term and a time-constant). This analog B0
compensation is never perfect and can deviate from the actual situation in
individual signal acquisition, due to different patterns of eddy currents in
different k-space trajectories. These imperfections in the central
k-space data can lead to artifacts in the UTE images. The same phantoms have also
been imaged using another Bruker micro-imager with a different configuration of
the B0 compensation method (Bruker AVANCE IIIHD). The same post-acquisition processing methods
yielded improved but different appearances of the images.Conclusion
Different
post-acquisition processing
methods can point out the sources of the artifacts in UTE images, which are imaging
hardware dependent. Iteration in post-processing can reduce the UTE artifacts.Acknowledgements
The authors are grateful to the
National Institutes of Health (NIH) for a R01 grant (AR 69047). The
authors thank Dr. Dieter Gross and Dr. Thomas Oerther from Bruker (Germany) for
their assistance with the UTE sequence and understanding of hardware setup. References
1. J. O'Sullivan, IEEE Trans.
Med. Imaging 1985, 4(4):200–207.
2. J. Jackson, C. Meyer, D. Nishimura, A. Macovski, IEEE Trans.
Med. Imaging 1991 10(3): 473–478. 3. J.A. Fessler, J Magn Reson. 2007
Oct;188(2): 191-5.