Giske Opheim1,2, Oula Puonti3,4, Jan Ole Pedersen5, Vincent O. Boer3, Ane Kloster1,2, Martin Prener1,2, Helle Juhl Simonsen6, Olaf B. Paulson1,2, Lars H. Pinborg1,2, and Melanie Ganz1,7
1Neurobiology Research Unit, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark, 2Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark, 3Danish Research Centre for Magnetic Resonance, Funktions- og Billeddiagnostisk Enhed, Copenhagen University Hospital Hvidovre, Copenhagen, Denmark, 4Dept. of Health Technology, Technical University of Denmark, Kongens Lyngby, Denmark, 5Philips Healthcare, Copenhagen, Denmark, 6Functional Imaging Unit, Department of Clinical Physiology, Nuclear Medicine and PET, Rigshospitalet Glostrup, Copenhagen, Denmark, 7Dept. of Computer Science, University of Copenhagen, Copenhagen, Denmark
Synopsis
Automated cortical segmentations benefit from higher SNR and spatial resolutions on 7T MR images, but are also challenged by B1 inhomogeneities, causing faulty surface inflations primarily in the temporal lobes. We investigated how FreeSurfer outputs were affected by applying eight different preprocessing schemes prior to reconstructions of submillimeter 7T MPRAGE images. The highest segmentation robustness across subjects was obtained by setting bias-field correction FWHM to 60mm and adding light regularization, and additionally performing intensity normalization.
Introduction
Computational analysis of structural MRI scans provides valuable
insights into anatomical changes in different brain diseases. Ultra-high field
(UHF) scanners provide unprecedented spatial resolution and SNR1
with potentially higher sensitivity to detect minute anatomical changes aiding
the discovery of novel biomarkers2-3. However, non-uniform intensity
distributions and inhomogeneous B1 transmit and receive fields at UHF MRI lead
to undesired variations in image intensity and hamper the automated
segmentation of structural images using software such as FreeSurfer (FS)4.
These issues could lead to selection bias where only the subjects in which the
automated analysis is successful can be included. On the FS wiki-page5,
bias-field correction with 18mm FWHM and sampling distance = 2 is recommended
prior to running recon-all on 7T data, together with tuned expert options adapted
to the temporal lobe intensities in your images. These settings somewhat improve
recon-all output, but still entail consistent failures in several regions in
our datasets. We therefore investigated how different preprocessings affect the robustness of FS segmentations
based on a submillimeter 3D MPRAGE sequence acquired using a the most widely
used commercial head-coil for 7T.Methods
3D MPRAGE images (0.7mm isotropic) were acquired on a 7T MR system
(Philips, Achieva, Best, The Netherlands), with 19x19cm dielectric pads on both
sides, with a quadrature 32/2 Rx/Tx coil (Nova Medical, Wilmington, MA). Six
subjects were included in the study, informed consent was obtained according to
the local ethical guidelines. The data were processed using FS’s recon-all
stream (conf2hires) with eight different preprocessing schemes: 1) No bias-field correction (BFcorr), 2)
BFcorr with low (30mm) FWHM and no regularization, 3) BFcorr with high (60mm) FWHM with liberal
regularization, and 4) BFcorr with high FWHM with tight regularization - all
performed with and without intensity normalization by a spatially adaptive non-local
means filter. An overview of the different preprocessing steps can be seen in
Figure 1. BFcorr was performed with a
downsampling factor of 3, and done in SPM12. Intensity normalization was done
in SPM12 with functions provided in the CAT12 toolbox. All 48 segmentation
outputs were visually quality controlled by the same observer. Variances in
average thicknesses across six typically problematic cortical regions and
across all six patients were calculated for all eight preprocessing
combinations.Results
Visual inspection of segmentations revealed improved quality after
applying scheme 2, 3 and 4, compared to scheme 1. Scheme 1, 2 and 3 further improved quality after
combining it with intensity normalization, whereas scheme 4 did not. Scheme 2
seemed to yield the lowest variances (see Figure 2), but visual quality
control revealed consistent failures in all subjects. Scheme 3 with intensity
normalization gave the highest robustness based on qualitative assessment of
segmentations (see Figure 3), and on the variances (Figure 2).Discussion
The results above
highlight that both visual inspection and quantitative validations are
important when assessing output from FreeSurfer. One should also note that even
though we significantly improve the segmentation performance after
preprocessing, manual edits might
still be necessary. Furthermore, our results show that it
is better to do bias-field correction and intensity normalization than not to, but also that surface quality control failures occur more often and consistently in the
least and most flexible schemes
(“fwhm60_tightreg”
and “fwhm30_noreg”). The reason for these failures in the former case is likely
that the bias correction is not flexible enough to correct for all the
intensity inhomogeneities, whereas in the latter case it might be too flexible
and start fading out contrast differences between tissues.Conclusion
Based on our findings, we
recommend performing bias-field correction with FWHM set to around 60mm along
with intensity normalization before automatic whole-brain segmentation on 7T
MPRAGE images from a classical 7T MR scan set-up.Acknowledgements
The project is supported by the Independent Research Fund Denmark. The 7T
scanner was donated by the Danish Agency for Science, Technology and Innovation
grant no. 0601-01370B, and The John and Birthe Meyer Foundation.References
- Trattnig, S., Springer, E., Bogner,
W., Hangel, G., Strasser, B., Dymerska, B., ... & Robinson, S. D.
(2018). Key clinical benefits of neuroimaging at 7 T. Neuroimage, 168, 477-489.
- Fujimoto, K., Polimeni, J. R., Van Der
Kouwe, A. J., Reuter, M., Kober, T., Benner, T., ... & Wald, L. L.
(2014). Quantitative comparison of cortical surface reconstructions from
MP2RAGE and multi-echo MPRAGE data at 3 and 7 T. Neuroimage, 90, 60-73.
- Zaretskaya, N., Fischl, B., Reuter,
M., Renvall, V., & Polimeni, J. R. (2018). Advantages of cortical
surface reconstruction using submillimeter 7 T MEMPRAGE. Neuroimage, 165, 11-26.
- Fischl, B. (2012). FreeSurfer. Neuroimage, 62(2), 774-781.
- https://surfer.nmr.mgh.harvard.edu/fswiki/HighFieldRecon