Artem Mikheev1, Louisa Bokacheva1, Heesoo Yang1, Jeremy Sobel1, Carlos Fernandez-Granda2, Hersh Chandarana1, and Henry Rusinek1
1Radiology, NYU School of Medicine, New York City, NY, United States, 2Courant Institute of Mathematical Science, New York University, New York City, NY, United States
Synopsis
We have implemented a method (BiCal) for
correction of image nonuniformity. Using objective criteria we have compared BiCal to widely
used N4 algorithm in several challenging clinical MRI applications. There
was a significant advantage of BiCal over N4 for 7T brain MRI and for accelerated radial GRASP of the abdomen. The performance of BiCal and N4 were comparable in breast imaging.
Introduction
Image intensity nonuniformity is a
common MRI artifact manifested by variation of signal intensity across the
image even within the same tissue. Nonuniformity may be caused by RF field
inhomogeneity, inhomogeneous reception coil sensitivity, and patient’s
influence on the magnetic and electric field. This artifact is detrimental to
tissue segmentation, inter-modality coregistration, parametric mapping, and
radiomics analysis, and therefore retrospective correction is often required.
Retrospective approaches entail
estimating a "smooth" multiplicative bias field B(r) to
correct the image (Figure 1). The most commonly used method is the N3 algorithm1
and its N4 refinement. 2 N3/N4 employ an iterative deconvolution that
estimates B(r) so as to maximize the
high frequency content of the corrected image. In our experience, however, the method
does not perform well in challenging imaging situations such as abdominal,
ultra-high field or accelerated MRI, and further improvements are necessary. We
have implemented a method we call BiCal where, after preprocessing to exclude sharp
edges and background region, the bias field is fitted directly. We compared the
performance of BiCal to N4 in three challenging applications: abdominal MRI
acquired at 3T using accelerated radial GRASP, 3T breast imaging using a
dedicated breast coil, and 7T head imaging. Methods
Patient datasets: This was an IRB-approved, retrospective study with waived informed
consent. Patient datasets (Figure 2) were randomly selected from image database
used in previous clinical research studies 3,4. (A). Abdominal images were
time-frames extracted from dynamic studies acquired using GRASP 3D stack-of-star
radial FLASH sequence, TR/TE= .27/1.88 msec, FA=12o, voxel size =1.5
x 1.5 x 3 mm. Breast images were acquired a 3T Siemens Trio using a seven-element
surface breast coil (Sentinelle, Invivo, Gainesville, FL, USA), TR/TE=4.74/1.79
ms, 448 x 358 x 116 matrix. MPRAGE head images (TR/TE/TI=2.6/2600/1100 msec,
FA=6°, 346 × 323 x 248
matrix) were acquired on a 7T Siemens magnet and a volume-transmit 24-element
receive coil array (Nova Medical, Boston, Mass).
Correction methods: N4 (a
variant of N3) expresses the bias field B(r) as a convolution of the intensity histogram by a
Gaussian. The computation consists of iterating between deconvolving the
intensity histogram by a Gaussian, remapping the intensities, and then
spatially smoothing this result by a B-spline model. The executable N4 code 2
is called explicitly by our software. BiCal method is implemented
in FireVoxel image analysis software from NYU Center for Biomedical Imaging. In BiCal B(r) is
estimated as a set of smooth bias functions such as direct cosines or 3D Lagarange
polynomials. After preprocessing to remove sharp edges and background (air)
signal, the partial derivatives of the bias field are fitted directly to the
partial derivatives of the intensity.
Outcome Measure: To quantify nonuniformity
before and after correction we used the coefficient of joint variation5
(CJV). CJV = (σ1+σ2 ) / |μ1-μ2|, where μ and σ are the
mean and stdev of the intensity of two sets of regions of interest (seeds)
placed across the image. CJV quantifies both the intensity variability in each set
and controls for the loss of tissue contrast.
Seeds: Four experienced readers independently placed two groups of seeds.
In the brain dataset, seeds were placed in white matter and gray matter; in the
breast, in fat and fibroglandular tissue (FGT); and in the abdomen, in
subcutaneous fat and muscle. Seeds were added incrementally until the change in
CJV after the addition of new batch of seeds did not exceed 2.5%. This resulted
in the following number of seeds (avg±st dev): in the brain, 194±14 in both
white and gray matter; in the breast, 181±71 in fat and 113±49 in FGT; and in
the abdomen, 253±102 in fat and 254±104 in muscle.
Training and testing: N4 and BiCal were optimized on the randomly selected training datasets.
The size of the training set is determined by successive addition of cases and
optimizing parameters until the change in average CJV becomes negligible. The
optimal parameters (Figure 3) are then used to process an independent set of test
cases and compare algorithm performance. Results
The process of parameter optimization (FIgure 3) successfully
converged in 3-4 iterations. The
original CJV and its change after correction (Figures 4 and 5) varied significantly across
the three applications. The abdominal (A) dataset showed the highest non-uniformity, the
breast (B) images were the most uniform, and 7T brain MRIs (C) were in
intermediate position. There was a significant improvement for each method and modality
as compared with originally acquired image: p<0.001 for (B) and (C),
p<0.05 for (A). For modalities A and C, but not for the breast there
was a significant advantage of BiCal over N4.
Discussion
Based on CJV, the new BiCal method significantly outperformed the N4 method in the challenging scenario of 7T brain MRI and abdominal GRASP datasets. These two applications had higher original CJV than breast images. This suggests that BiCal is the method of choice for high-field and fast MRI acquisitions.
Acknowledgements
This
work was supported by NIH/NIBIB grant U24 EB028980.References
1. Sled JG, Zijdenbos AP, Evans AC. (1998). A nonparametric
method for automatic correction of intensity nonuniformity in MRI data. IEEE
Trans Med Imag 17(1), 87–97.
2. Tustison NJ, Avants BB, Cook PA, Zheng Y, Egan A,
Yushkevich PA, Gee JC. (2010). N4ITK: improved N3 bias correction.IEEE Trans
Med Imag 29(6), 1310-20.
3. Pujara AC, Mikheev A, Rusinek H et al. (2017). Clinical
applicability and relevance of fibroglandular tissue segmentation on routine T1
weighted breast MRI. Clinical Imaging 42:119-125.
4. Fujimoto K, Feng L,
Otazo R, Block K, Rusinek H, Wake N, Chandarana Hersh. (2017). GRASP with
Motion Compensation for DCE-MRI of the Abdomen. Program#2009, ISMRM 2017.
5. Belaroussi B, Milles J, Carme S, Zhu Y-Ma, Benoit-Cattin
H. (2006). Intensity non-uniformity correction in MRI: Existing methods and
their validation. Medical Image Analysis 234–246.