Ignacio Xavier Partarrieu1, Elliot Fox1,2, Frederic Brochu1, Matt Cashmore1, Neha Koonjoo3, Sheng Shen3, David Sinden1, Matthew Rosen3, and Matt G. Hall1,4
1National Physical Laboratory, Teddington, United Kingdom, 2Durham University, Durham, United Kingdom, 3Harvard, Boston, MA, United States, 4University College London, London, United Kingdom
Synopsis
The large expense and running costs of traditional 1.5T scanners
have meant that as demand for scans has grown supply has not been able to keep
up. Novel, ultra-low field systems aim to address this problem by producing
images with some resolution losses as a trade-off for reduced costs.
However, access to such devices is currently restricted, limiting the possible research
output. In this work we develop and test a simulation model of 6.5 mT b-SSFP
acquisitions, which we compare to actual scanner acquisitions. We find that
such simulations produce qualitatively and quantitatively similar images to real
scans.
Introduction
There is a newly emergent interest in ultra-low field (ULF) MRI, driven
by recent successes in the field and cost-cutting pressures in healthcare. In
particular, scanners with a cost of less than $50 000 have been reported [1], which is within the reach of consortia of
general practitioners. This means that such devices could potentially be used
as a decision tool for the triage of patients, identifying those which might
need higher resolution scans and those for which no follow-up is necessary.
This would reduce some of the throughput pressures currently faced by hospitals.
Simulations of ULF MRI scanners would be useful in helping to determine whether
they can provide the information necessary for this and permit multiple
researchers to carry out research at once, without needing access to the highly
specialized equipment. In this work we have developed and tested software for creating
such simulations and found them to be representative of ultra-low field acquisitions.Methods
The ULF scanner described by Sarracanie [1] was chosen as the
basis of this simulation, which was implemented using MRiLab [2]. The b-SSFP sequence described within was
replicated by adapting the FIESTA sequence within the
software to have a TR of 22.5 ms as described in the paper, with a voxel size
of 2.5×3.5×11.5 mm3 . The MRiLab GUI was
bypassed, with changes made directly to the gradient codes to avoid some
undesired default assumptions, such as identical rise times for all pulses within
a gradient sequence. Additionally, the ULF scanner uses a bespoke wound litz
coil to improve signal. This atypical coil presents several advantages in the
ULF environment and was modelled by coding in an Archimedean spiral coil
element for MRiLab. The ULF k-space
acquisition scheme, with Gaussian sampling of half of the k-space, zero
padded to be a 96×96
matrix, was also simulated, though the Gaussian sampling was applied post-simulation. The reconstructed simulation image was filtered using anisotropic
diffusion through the Perona-Malik filter as implemented in MATLAB, with a
gradient threshold of 10% and only two iterations. The change of spin
values at ULF was also considered, and values were adjusted using extrapolation
based on historical MR data [3], though it is unsure that these rules hold at ultra-low
fields, as there are limited reports of T1 and T2 measurements
at this field strength. An acquisition of a resolution phantom on the ULF MRI scanner was used
for qualitative and quantitative comparison. The latter was performed using the
signal-to-noise ratio (SNR), defined as mean over the phantom area vs. standard
deviation of the background.
Results
A slice of the ULF acquisition of the resolution phantom may be seen in
figure 1, and a slice of the simulated brain phantom is shown in figure 2. Qualitative
inspection of these shows similar intensity and noise properties, though some
of the finer structures normally visible in the brain phantom have been
blurred. The SNR of the brain phantom is 21.9, which is slightly higher than
the value calculated for the resolution phantom, which is 19.2.Discussion
MRiLab is capable of simulating
ULF MRI acquisitions accurately, though care is required to ensure that
physical parameters such as T1, T2 and T2* are still representative. Though visual
inspection and SNR values suggest that the simulation is accurate, further
validation is needed by simulating and measuring known quantitative objects or
simulating and measuring the same standard object. The slight difference in SNR
could be due to either a lower noise contribution in the simulation than could
be expected, or differences in the implementation of the filtering process such
as in the gradient threshold or the number of iterations. A next step in the
validation of the model would be to use simulations and acquisitions of the
same known object to investigate both SNR and contrast-to-noise metrics, in
order to ensure that the information content is similar and that the
simulations do not lose or gain contrast from discretisation effects.Conclusion
Simulations using MRiLab appear to provide value even for ultra-low
field investigations. This can be used for the testing of hypotheses in less
time and more cheaply than for physical investigations and provide access to
ULF testing to researchers without their own ULF scanner. Further validation
work involving the simulation of various objects will provide more detail on
the limitations and advantages of this method of investigation.Acknowledgements
This work was funded by CRUK grant C69862/A29020References
[1] Sarracanie,
M. et al. Low-Cost High-Performance MRI. Scientific
Reports 5, 15177 (2015).
[2] Liu, F., Velikina, J. V., Block, W. F.,
Kijowski, R. & Samsonov, A. A. Fast Realistic MRI Simulations Based on
Generalized Multi-Pool Exchange Tissue Model. IEEE Transactions on Medical
Imaging 36, 527–537 (2017).
[3] Rooney, W. D. et al. Magnetic
field and tissue dependencies of human brain longitudinal H2O relaxation in
vivo. Magn. Reson. Med. 57, 308–318 (2007).