Xiaoyin Wang1, Qizhi Yang2, Hongjian He1, Congbo Cai2, Yi-Cheng Hsu3, and Jianhui Zhong1,4
1Center for Brain Imaging Science and Technology, Department of Biomedical Engineering, Key Laboratory for Biomedical Engineering of Ministry of Education, Zhejiang University, Hangzhou, China, 2Department of Electronic Science, Fujian Provincial Key Laboratory of Plasma and Magnetic Resonance, Xiamen University, Xiamen, China, 3MR Collaboration, Siemens Healthcare Ltd., Shanghai, China, 4Department of Imaging Sciences, University of Rochester, Rochester, NY, United States
Synopsis
Magnetic Resonance parametric
mapping can provide quantitative information to characterize tissue properties.
Recently, a single-shot T2 mapping method based on Multiple Overlapping-Echo
Detachment (MOLED) planar imaging was proposed. However, limited echo time
ranges still affected the reconstruction accuracy of the T2 values, especially
when large T2 value
ranges were present. In this abstract, MOLED was expanded through
multiple-echo-train acquisitions that achieved high accuracy and better
texture. The deep convolution neural network was used to reconstruct T2 maps, B1 maps and
spin densities in synchrony. The sequence efficiencies were demonstrated in digital-brain,
phantom and human-brain experiments.
Introduction
Magnetic Resonance
parametric mapping can provide quantitative information to characterize tissue properties.1,2 However, slow data acquisition speeds usually hinder real-time magnetic
resonance (MR) parameter mapping. In this work, a novel single-shot T2 mapping method, which can be acquired in hundreds of milliseconds, was
proposed based on multiple overlapping-echo detachment (MOLED) with a planar
imaging acquisition scheme, termed the multi-train MOLED (MMOLED) method. Multiple
overlapping echo signals from different echo trains with the different T2-weighting were obtained simultaneously.3 A convolution neural
network (CNN) was proposed to reconstruct T2 mapping. The robustness and
efficiency of the MMOLED sequence were demonstrated using numerical brains,
phantom experiments and human brains.Methods
The novel single-shot MMOLED T2 mapping sequence designed is shown in Fig.1. The
current strategy uses sequences that have three echo trains, and for each echo
train, four echo signals with different evolution time are generated in turn by
four excitation pulses that hold a same small flip angle of α.
Four echo-shifting gradients (G1,G2,G3,G4) were used to shift the echoes from the k-space
center along the phase-encoding and frequency-encoding directions.
To compare the reconstruction results
of different echo train numbers during post-processing, we used first echo
train (dubbed MMOLED-1 train), the first two echo trains (dubbed MMOLED-2
trains) and the first three echo trains (dubbed MMOLED-3 trains) as the input
of the convolution neural network (CNN). The pure synthetic training samples
were utilized, the detail can be seen in the previous paper.4-7 The
phantom and human brain validation experiments were performed on a whole-body
3T scanner ( MAGNETOM Prisma, Siemens Healthcare, Erlangen, Germany). For the
single-shot MMOLED-1, MMOLED-2, and MMOLED-3 trains sequence, the scan time was 118 ms, 185 ms and 252 ms respectively with acquisition matrix 128×128 and FOV 22×22 cm2. The conventional SE sequence was used for
comparison.Results and Discussion
A numerical Brain pattern
(including T2 , B1 and proton density templates) was constructed.
The numerical simulation results are shown in Fig.2. The images before
reconstruction (Fig. 2(a)) had visible cross stripes, since they were composed
of multiple images with different linear phase ramps. A reference T2 map (Fig.
2(b)) was used as a templet for the simulation. We enlarge a region in
numerical brain for comparison. The reconstruction results of the three input types
mentioned are also shown in Fig. 2(c-e). The T2 mapping textures
from the MMOLED-2 (Fig. 2(d)) and MMOLED-3 train3 (Fig. 2(e)) were most
consistent with the T2 mapping
reference, especially in regions with relatively large T2 values,
as indicated by the red arrows. The red circle shown in Fig. 2(a) denotes a
trace for the comparison of the T2 values
from MMOLED-1, MMOLED-2, MMOLED-3 trains and reference T2 mapping.
The T2 profiles
along the red trace from the different T2 mapping schemes
were drawn in Fig. 2(f), which showed that in the cerebrospinal fluid (CSF)
regions where high T2 values
usually exist, the results from multi-train OLED including MMOLED-2 trains and
MMOLED-3 trains had better agreement with the reference than MMOLED-1 train.
The Structural Similarity Index (SSIM) for the entire image results of MMOLED-1,
MMOLED-2, and MMOLED-3 trains were 0.775, 0.991 and 0.988, respectively.
For the phantom validation, the
mean T2 values ranged
from 15 ms to 180 ms (Fig. 3). We found that T2 mapping
reconstructed from MMOLED-2 and MMOLED-3 trains performed better when T2 value was
a relatively large value (>75ms). Moreover, the maximum compartment 16 deviation
for MMOLED-1 train was about 11.1 ms, while that of the MMOLED-2 and MMOLED-3
trains was about 6.2 ms and 5.7 ms, respectively.
In Fig.4, the mean T2 values
and standard deviations (SDs) for 17 regions of interest (ROIs) marked in Fig.
4(a) were calculated and the results were given in Fig. 4(b). The 17 ROIs were
chosen manually to represent different regions of the brain, including white
matter, gray matter, and deep gray matter with different T2 values. All
of the reconstructed T2 mapping
results were compared with the T2-mapping reference obtained from SE sequence
using the FNIRT tool (FMRIB’s Nonlinear Image Registration Tool) of FSL
(Oxford, UK). Most results agreed well with SE result, but in ROI10 where T2 value was
relatively large (>75ms), the reconstructed results from MMOLED-2 and
MMOLED-3 trains performed better than those of the MMOLED-1 train. We found that
the reconstructed T2 mappings
from MMOLED-2 and MMOLED-3 trains (Fig. 4(a)) were closer in texture details to
the T2 reference
compared with the mapping from MMOLED-1 train. The B1 and M0 maps could
also be reconstructed through the convolution neural network output from three
different methods as shown in Fig. 5.8 As one can see, The reconstructed B1 mapping
from MMOLED-2 and MMOLED-3 trains was also enhanced than from the MMOLED-1
train.Conclusion
The MMOLED-2 and MMOLED-3
trains provided better single-shot T2 mapping
and B1 mapping than
did the MMOLED-1 train, and all could be acquired within milliseconds while maintained high image quality, which is useful for dynamic imaging. The
successful application of deep learning method and synthetic training dataset
proved their great potential in making complex MRI sequences available.Acknowledgements
No acknowledgement found.References
1. Ma D, Gulani V, Seiberlich
N, et al. Magnetic resonance fingerprinting. Nature.2013;495:187-192.
2. Zhao B, Lam F, Liang Z.
Model-based MR parameter mapping with sparsity constraints:parameter estimation and performance bounds. IEEE Trans Med Imaging.
2014;33:1832-1844.
3. M.-L. Chu, H.-C. Chang, K.
Oshio, and N.-K. Chen. A single-shot T2 mapping
protocol based on echo-split gradient-spin-echo acquisition and parametric
multiplexed sensitivity encoding based on projection onto convex sets
reconstruction. Magnetic Resonance in Medicine. 2018;79(1):383-393.
4. Ma L, Cai C, Yang H, et al.
Motion-tolerant diffusion mapping based on single-shot overlapping-echo
detachment (OLED) Planar Imaging.
Magnetic Resonance in Medicine. 2018;80:200-210.
5. Cai C, Zeng Y, Zhuang Y, et
al. Single-shot T2 mapping
through overlapping-echo detachment (OLED) planar imaging. IEEE Transactions on
Biomedical Engineering. 2017;64(10):2450-2461.
6. Cai C, Wang C, Zeng Y, et al.
Single-shot T2 mapping using overlapping-echo
detachment planar imaging and a deep convolutional neural network. Magnetic
Resonance in Medicine. 2018;80(5):2202-2214.
7. Zhang Jun, Wu Jian, Chen Shaojian, et al.
Robust Single-Shot T2 Mapping via Multiple
Overlapping-Echo Acquisition and Deep Neural Network. IEEE Transactions on
Medical Imaging, 2019;38(8):1801-1811.
8. Cunningham, C.H., J.M. Pauly
and K.S. Nayak, Saturated double-angle method for rapid B1+ mapping. Magnetic Resonance in Medicine.
2006;55(6):1326-1333.