Xiaoyang Liu1,2 and Paul Bottomley1,2
1Department of Electrical and Computer Engineering, Johns Hopkins University, Baltimore, MD, United States, 2MR Research Division, Department of Radiology, Johns Hopkins University, Baltimore, MD, United States
Synopsis
Parameterized
image synthesis is desirable for region-specific image contrast and protocol
optimization. While this could arguably be performed with quantitative mapping
and Bloch equation simulations, it is not well-suited for real-time cine MRI or
MRI endoscopy in a real-time interventional application with views changing
from frame-to-frame. Here we propose to use empirical data to calibrate signal
behavior patterns from a specific MRI endoscopic device and pulse sequences to fit
new acquisitions from which extra images and ultimately guide automated local
contrast optimization.
Purpose
Intravascular MRI endoscopy
uses a micro-MRI transmit/receive coil moving through blood vessels(1–3), with a goal of facilitating the detection and
diagnosis of vascular disease or for guiding and monitoring therapy delivery such
as the treatment of tumors involved with blood vessels(4,5). Recently the speed of MRI endoscopy was
accelerated from ~2 frames-per-second (fps) to ~10fps using highly under-sampled
nonlinear reconstruction and graphics processing units to achieve real-time display
and feedback(6).
During MRI endoscopy from the viewpoint of the
coil at the probe-head, the surrounding tissue is ever-varying. A single set of
protocol parameters is generally not optimal for all locations or even a
specific region-of-interest. Adjusting the parameters on-site is time-consuming,
requiring user experience, and knowledge of the MRI physics and properties of
the local environment. Protocol parameter-based image synthesis is desirable as
a first step towards automatic optimization of image contrast during high-speed
acquisitions. Conventionally image synthesis is done by Bloch equation
simulation, but this requires additional quantitative mappings that may not account
for all contrast mechanisms (the true contrast in MRI endoscopy is a
convolution of through-plane signals that reflect the highly nonuniform coil profile
and the adiabatic excitation) and takes time that compromises the real-time
nature of the signal stream. Here we propose a method that learns the contrast
behavior from empirical data as a calibration set for a specific endoscopy coil.
This can then be applied to synthesize image contrasts for unacquired parameter
sets based on a fit to a few acquired images at updated locations.Methods
A loop MRI coil catheter fitted
to a Siemens clinical 3T Prisma
MRI scanner was used to acquire endoscopic images in post-mortem human artery
samples acquired from our pathology department. These vessels had
atherosclerosis plaques including calcifications. Images with a range of protocol
parameters–repetition time, TR; flip angle, FA; echo time, TE–were acquired from
different vessels, initially constraining the adjustment dimensions to FA and
TE. 400 images were acquired with a FLASH sequence modified for MRI endoscopy by
replacing the excitation pulse with an adiabatic ‘BIR-4’ pulse(1,2) (MRI parameter sets: 20 FA x 20 TE, FA=5-100°,
TE=5.6-13.2ms, TR=20ms, time of acquisition, TA=3.6s, field of view, FOV=40mm,
resolution=200µm) to create a “parameterized image library.” The same procedure
was repeated in other vessels for testing.
After acquisition, the signals
at each pixel location of the endoscopy foreground (91x91 pixels) were
extracted to form 8281 signal patterns in FA-TE space. The patterns were used to
train a variational autoencoder (VAE) (encoder =2 fully-connected layers;
decoder =3 fully-connected layers) with 50 epochs(7). The latent code space dimension of the
autoencoder was set to 10 to constrain the propagation of information and force
the output to retain only the strong signal patterns, as a denoising step.
These signal patterns were treated as a calibration set to use as a basis for
synthesizing new endoscopy images that could be used to automatically optimize
contrast with minimal new input data.
The application was tested by arbitrarily selecting
several protocol settings (3 acquisitions: FA/TE= 20/10.8, 45/8.8, 70/6.8; or 5
acquisitions: FA/TE=20/10, 50/8, 60/12.4, 70/12.8, 95/11.2) from the set of test
images. For each pixel location of the test image, the signals were matched based
on their ‘cosine similarity’ to the most probable filtered signal patterns in
the calibration set, to fill the signal in the rest of the 400 parameter
locations. After rearrangement, the upshot was a set of images representing a
best estimate of the test image contrast at 400 parameter combinations. This
can be used to guide subsequent selection of acquisition parameters. For
reference, the testing procedure was run 100 times with 3-10 random parameter
selections and the average cosine similarity of the estimated images as
compared to ground truth was calculated.Results
Figure
1 shows examples of endoscopic human vessel images in the calibration set
acquired with different parameters. The dark discoidal regions in the center
are the endoscopy coils. Note the difference in visibility of the attached
tissue (white arrows) and the differentiation of calcifications in the vessel
wall (yellow arrows). Figure 2 shows examples of signal patterns of pixels at
different locations. The raw acquisition and filtered data illustrate the
network’s retention of FA-TE contrast features and suppression of noise
features. Figure 3 shows synthesized images of the testing sample at unseen
parameter settings from 3 acquisitions and 5 acquisitions, as well as the
ground truth. Note the visibility and differentiability of attached tissues,
calcifications and vessel walls (arrows). Box plots in Figure 4 shows the
distribution of average similarity between the synthesized images and ground
truth as a function of the number of measurements.Discussion
This
work demonstrates the feasibility of using empirical data for parameterized
image synthesis in MRI endoscopy, yielding comparable image quality. The cosine
similarity evaluation is used for reference here, but may not be the ultimate criterion
for perception quality. Such synthesis has the potential to guide automatic localized
contrast optimization in a real-time in
vivo setting such as the interventional applications intended here. Further
investigation to extend the process to higher dimensional parametric spaces and
to conventional MRI acquisition settings is underway.Acknowledgements
Supported by NIH grant R01 EB007829
and the Russell H Morgan Chair in Radiology.References
1.
Sathyanarayana S, Bottomley PA. MRI endoscopy using intrinsically localized
probes. Med. Phys. 2009;36:908–919 doi: 10.1118/1.3077125.
2.
Sathyanarayana S, Schär M, Kraitchman DL, Bottomley PA. Towards Real-Time
Intravascular Endoscopic Magnetic Resonance Imaging. JACC Cardiovasc. Imaging
2010;3:1158–1165 doi: 10.1016/j.jcmg.2010.08.014.
3. Hegde
SS, Zhang Y, Bottomley PA. Acceleration and motion-correction techniques for
high-resolution intravascular MRI. Magn. Reson. Med. 2015;74:452–461 doi:
10.1002/mrm.25436.
4. Wang
G, Zhang Y, Hegde SS, Bottomley PA. High-resolution and accelerated
multi-parametric mapping with automated characterization of vessel disease
using intravascular MRI. J. Cardiovasc. Magn. Reson. 2017;19:89 doi:
10.1186/s12968-017-0399-6.
5. Liu X,
Ellens N, Williams E, et al. High-resolution intravascular MRI-guided
perivascular ultrasound ablation. Magn. Reson. Med. 2020;83:240–253 doi:
10.1002/mrm.27932.
6.
Xiaoyang Liu, Parag Karmarkar, Dirk Voit, Jens Frahm, Paul Bottomley. Real-time
MRI endoscopy at up to 10 frames/sec. In: Proc. Intl. Soc. Mag. Reson. Med. 26
(2018). Paris, France; 2018. p. 0603.
7. Kingma DP, Welling M.
Auto-Encoding Variational Bayes. ArXiv13126114 Cs Stat 2013.