Stephen Cauley1,2, Bryan Clifford3, Steffen Bollmann3, Thorsten Feiweier4, Berkin Bilgic1,2, Kawin Setsompop1,2,5, and Lawrence L. Wald1,2,5
1Department of Radiology, A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Siemens Medical Solutions USA, Boston, MA, United States, 4Siemens Healthcare GmbH, Erlangen, Germany, 5Harvard-MIT Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, MA, United States
Synopsis
We employ a compact
phase modeling strategy for accurate multi-shot echo-planar imaging (msEPI)
reconstruction. As an alternative to approaches that perform msEPI reconstruction
using strict low-rank constraints, we recast the problem as an iterative relative
phase estimation problem. This allows for us to utilize existing techniques
such as ESPIRiT, which are formulated for determining relative magnitude and
phase differences between multi-coil receive arrays. Through an iterative
search we jointly estimate an artifact-free combined image and the smooth relative
phase between each msEPI shot. We demonstrate the benefits of our approach for
clinical and highly-accelerated multi-shot diffusion-weighted acquisitions.
INTRODUCTION
Single-shot two-dimensional (2D) echo-planar imaging acquisition
methods are routinely used for both diffusion-weighted and functional magnetic
resonance imaging (fMRI) studies. Several methods for acceleration have been
explored to reduce the long pulse repetition time (TR) associated with
high-resolution full-brain coverage imaging. Accelerated 2D parallel imaging (PI)
techniques1–5 are used to remove phase-encoding
steps during an acquisition. These methods have many benefits, such as a
reduction of image distortion and T2* blurring. These
benefits are furthered through recent advances in model-based reconstruction,
where joint reconstruction of multiple highly accelerated EPI shots is
performed6–8. In prior works6,7, a low-rank representation for the block-Hankel
matrix across shots was motivated. The foundation of these methods is an assumption
that the smooth relative phase differences between the shots would live in the
null space of the block-Hankel matrix, which is comprised of shifted copies of
the k-space data. Although in certain scenarios these methods perform well, the
underlying assumption need not be satisfied.
We build an iterative reconstruction framework (PRIME) that jointly estimates an artifact-free
combined image and the relative phase between each msEPI shot. Throughout the PRIME algorithm we leverage
ESPIRiT9 to arrive at smooth relative phases between the shots. The compact phase
model provided by ESPIRiT relies on low-rank block-Hankel properties, locally
low-rank properties, and spatial continuity of phase. Figure 1 shows an
illustration of the method. Here, the initial phase of each shot is determined
through individual-shot SENSE reconstructions. These shot-specific phases are
then included in a SENSE reconstruction across all shots8. The combined image is then projected
through the phase estimates and then coil sensitivities. The original k-space
data is then integrated for data consistency. Coil combined shot images are fed
through ESPIRiT to determine the relative phase differences and the process repeats
until convergence. METHODS
PRIME was performed
using custom software written in MATLAB (Mathworks, MA, USA), and the data from
a healthy volunteer were collected using a prototype sequence (Figure 2) on a
3T system (MAGNETOM Prisma, Siemens Healthcare, Erlangen, Germany) in
accordance with the local IRB. EPI diffusion data were acquired at
1.2×1.2×4.0mm3, R=1, 224×224mm2 FOV, PF=6/8, 32 slices, b=1000
s/mm2, and BW = 1132Hz/px using a 32-channel coil. Here, we considered
fully sampled data arising from the use of 3 shots at [TR,TE] = [4800,94]ms,
and 8 shots at [TR,TE] = [5400,51]ms.RESULTS
The
accuracy of ESPIRiT for representing relative phase relationships in diffusion
data was tested on 3 segment data. This R=3, b=1000 s/mm2 data can be reliably
reconstructed using SENSE and it is used as the ground truth. Figure 3 shows
the magnitude and phase images from 2 of the single shot reconstructions. This
complex data was used as input to ESPIRiT, with 48×48 calibration size and 12×12 kernel size, and the
relative phase maps were extracted. The single shot SENSE and ESPIRiT phases
were each used within combined-shot R=1.5 SENSE reconstructions, and the data
consistency error was 39.4% and 40.6% respectively. Figure 4 shows a comparison
of PRIME and MUSSELS6 across
8 segments of fully-sampled data. To provide fair comparison between the
methods, we use a consistent kernel size (12×12) and limit MUSSELS to a desired rank of “1” to
match the “single” ESPIRiT maps. The singular value decay of the Hankel-matrix
constructed from each reconstruction is shown on the left of Figure 4. The average
magnitude image for MUSSELS and the combined-shot SENSE reconstruction for
PRIME are shown with their respective data consistency errors. In the zoomed
ROI we show the truncation of signal caused by the low-rank assumption of
MUSSELS. Finally, Figure 5 shows the applicability of the PRIME algorithm for
calibration-free removal of even/odd ghosts10. Here,
the vendor ghost correction was disabled, and the acquired segments were
separated into even and odd shots. Reconstruction improvement and data
consistency error reduction are shown.DISCUSSION AND CONCLUSION
We introduce an iterative approach for msEPI reconstruction
based upon shot-combined SENSE and ESPIRiT phase estimation. We show the
accuracy of the phase model using a standard clinical R=3 diffusion acquisition,
see Figure 3. Here, the PRIME algorithm converges in a single iteration and the
model accurately captures the significant phase variation to arrive within ~1% data
consistency compared to the ground-truth. In contrast, the MUSSELS algorithm takes
several steps (5-10) to impose the more stringent low-rank assumption.
In Figure 4, we have also shown the application of PRIME to
highly-accelerated diffusion-weighted acquisitions. We achieve comparable data
consistency error to MUSSELS with matched parameters and show better stability
across the reconstructed shots (flatter singular value decay). In addition, the
relaxed constraints afforded by the PRIME algorithm can also be clearly observed
with the retention of signal near the skull. The ability of PRIME to model
phase variation across even and odd EPI lines was shown in Figure 5. The data
consistency error decreased, and the ghosts were substantially mitigated.
Finally, this strategy aligns well with current ML guided msEPI reconstruction
approaches as the shot-combined SENSE reconstruction can be integrated with ML
priors.Acknowledgements
This work was
supported in part by NIH research grants: R01EB020613, U01HD087211, R01EB006847, R01EB019437,
R24MH106096, P41EB015896, and the shared instrumentation grants: S10RR023401, S10RR019307,
S10RR019254, S10RR023043References
1. Griswold, M. A. et al.
Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA). Magn.
Reson. Med. 47, 1202–1210 (2002).
2. Pruessmann, K. P., Weiger, M.,
Scheidegger, M. B. & Boesiger, P. SENSE: Sensitivity encoding for fast MRI.
Magn. Reson. Med. 42, 952–962 (1999).
3. Lustig, M., Donoho, D. & Pauly, J.
M. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn.
Reson. Med. 58, 1182–1195 (2007).
4. Sodickson, D. K. & Manning, W. J.
Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with
radiofrequency coil arrays. Magn. Reson. Med. 38, 591–603 (1997).
5. Liang, D., Liu, B., Wang, J. & Ying,
L. Accelerating SENSE using compressed sensing. Magn. Reson. Med. 62,
1574–1584 (2009).
6. Mani, M., Jacob, M., Kelley, D. &
Magnotta, V. Multi-shot sensitivity-encoded diffusion data recovery using
structured low-rank matrix completion (MUSSELS). Magn. Reson. Med.
(2017). doi:10.1002/mrm.26382
7. Haldar, J. P. Low-Rank Modeling of Local
k-Space Neighborhoods (LORAKS) for Constrained MRI. IEEE Trans. Med. Imaging
(2014). doi:10.1109/TMI.2013.2293974
8. Chen, N. kuei, Guidon, A., Chang, H. C.
& Song, A. W. A robust multi-shot scan strategy for high-resolution
diffusion weighted MRI enabled by multiplexed sensitivity-encoding (MUSE). Neuroimage
(2013). doi:10.1016/j.neuroimage.2013.01.038
9. Uecker, M. et al. ESPIRiT - An
eigenvalue approach to autocalibrating parallel MRI: Where SENSE meets GRAPPA. Magn.
Reson. Med. 71, 990–1001 (2014).
10. Lobos, R. A., Javed, A., Nayak, K. S.,
Hoge, W. S. & Haldar, J. P. Robust autocalibrated loraks for EPI ghost
correction. in Proceedings - International Symposium on Biomedical Imaging
(2018). doi:10.1109/ISBI.2018.8363661