Ke Jiang1, Jiazheng Wang1, and Xiaofang Xu1
1Philips Healthcare Greater China, Beijing, China
Synopsis
T1-weighted imaging is a necessary
component of clinical brain MR to provide anatomical information at high
spatial resolution. T1w brain image in clinical practice is regularly acquired
with two-dimensional turbo-spin-echo sequence, which suffers from the pulsation
artifacts in the phase-encode direction from superior sagittal sinus, the straight
sinus, and the sigmoid sinus. This study exploits the spiral trajectory for the
spin-echo sequence to achieve the T1 weighting and to avoid the pulsation
artifacts.
Introduction
T1-weighted imaging is a necessary component of
clinical brain MR to provide anatomical information at high spatial resolution.
T1w brain image in clinical practice is regularly acquired with two-dimensional
turbo-spin-echo sequence, which suffers from the pulsation artifacts in the
phase-encode direction from superior sagittal sinus, the straight sinus, and
the sigmoid sinus [1]. In contrast, the
spiral trajectory, due to its fast imaging nature, is less sensitive to motion,
and the absence of a phase-encode direction prohibits the formation of motion
artifacts in a certain direction. Despite the intensive research on spiral
trajectory for ultra-fast imaging [2], here we demonstrate the capability of
spiral acquisition in T1w sagittal brain imaging to avoid pulsation artifacts.Method
The spiral acquisition was implemented in a spin-echo
(SE) sequence such that a spiral-out trajectory starts at the echo time, a
schematic diagram of the sequence shown in Figure 1. The whole k-space was
acquired using multiple interleaves, with an illustration given in Figure 2. The
SE-spiral and the traditional TSE images, both with T1 weighting, were
collected at 3.0 T (Ingenia Elition, Philips healthcare, Best, the Netherlands)
on the brain of healthy volunteer with 32 channel head coil used. Informed
consent was obtained from each volunteer and the study was approved by a local
IRB. Isotropic FOV and voxel size were
acquired for both sequences: 230mm and 0.8mm. Other parameters for SE-Spiral:
TR/TE = 600/9.5ms, acquisition window (the duration of a single spiral) = 10ms,
interleaves = 33, NSA = 2, scan time = 2.7min. For TSE: TR/TE = 600/19ms, NSA
=1, TSE factor = 4, scan time =2.1min. The TR was kept the same for the two
sequences to yield image contrasts as similar as possible.Result
Figure 3 shows the T1 weighted brain images
obtained from both SE-Spiral and the traditional TSE sequences. Pulsation
artifact appeared only on TSE image which was indicated by red arrows. SNR/CNR
for SE Spiral and TSE was 42/23.8 and 29/11.2, respectively.Discussion
We demonstrated the feasibility of SE Spiral
in avoiding pulsation artifact for human brain imaging at 3T system. Longer acquisition
window can lead to less scan time but introduce more blur. In this study acquisition
window was set to 10ms and interleaves was set to 33 to achieve the balance between
image quality and scan time. The slightly longer TA for SE Spiral compared to TSE was due to NSA (=2) which was implemented to obtain higher SNR.Conclusion
For T1 weighted human brain imaging, SE Spiral acquisition
is capable of alleviating pulsation artifact at 3T systemAcknowledgements
No acknowledgement found.
References
[1] Katarzyna Krupa,
et al. Pol J Radiol.
2015; 80: 93–106.
[2] Curtis
L. Johnson, et al. Magnetic Resonance in Medicine. 2013; 70:404–412.