Diana Bencikova1,2, Stephan Kannengiesser3, Gert Reiter4, Ahmed Ba-Ssalamah1, Siegfried Trattnig1,2, and Martin Krššák2,5
1Department of Radiology, Medical University Vienna, Vienna, Austria, 2Christian Doppler Laboratory for Clinical Molecular Imaging, MOLIMA, MUW, Vienna, Austria, 3Siemens Healthcare GmbH, Erlangen, Germany, 4Siemens Siemens Healthcare Diagnostics GmbH, Graz, Austria, 5Division of Endocrinology and Metabolism, Department of Medicine III, Medical University of Vienna, Vienna, Austria
Synopsis
Radial T2 mapping
of the liver is fast and motion insensitive, which is ideal for clinical
applications. Here we tested a prototype radial turbo-spin-echo sequence for T2 mapping
of the liver in phantoms and in-vivo in patients. In the phantoms we compared
it to conventional Cartesian multi-SE sequence and tested the effect of fat
suppression by comparing fat-suppressed and fat-unsuppressed values. From the
patient evaluation, comparing it to single-voxel multi-echo MRS, we proved it
to be suitable for clinical applications and a promising tool for
characterization of diffuse liver disorders, but there are systematic
differences between different methods.
Introduction
Magnetic resonance is a non-invasive modality
that was established as a reliable method for diagnosis of many liver disorders1,2,3. Hepatic T2
relaxation times change with many pathological conditions, but clinical
application of conventional techniques is limited due to breathing motion in
the abdomen and long scan times. A fast radial turbo-spin-echo (rTSE) sequence
was proposed to overcome these problems4. Crossing the center of the
k-space with each acquisition results in oversampling and averaging of low
spatial frequencies and to motion insensitivity. In combination with tiered
echo sharing and pseudo golden angle ratio reordering, the scan time is
reduced, which makes it acceptable in clinical settings5. Therefore,
the goal of this study was to test a fast prototype rTSE T2 map in phantoms and
in-vivo.Workflow & methods
A prototype
rTSE T2 map (TR = 1500 ms, turbo factor = 29, TE range = 9.2 – 266 ms, FOV = 400×400
mm2, matrix = 256×256, FA = 180˚, ST = 6 mm, total acquisition time
= 17 sec) was compared to conventional Cartesian multi-SE based T2 map (TR =
5000 ms, turbo factor = 32, TE range = 11.5 – 380 ms, FOV = 250×250 mm2, matrix
= 256×256, FA = 180˚, ST = 6 mm, total acquisition time > 20 minutes) in the
NIST phantom with a T2 array (range 8 – 800 ms)6 at a 3T MRI scanner
(MAGNETOM Prisma Fit, Siemens Healthcare, Erlangen, Germany), equipped with a 18-channel
body coil and 32-channel spine coil. Secondly, to test the effect of fat
suppression on the T2 values obtained by the rTSE, fat-suppressed (FS) and
fat-unsuppressed (noFS) acquisitions were compared in phantoms with variable
fat fraction (FF = 0-30%). T2 values were calculated inline using a
mono-exponential signal model and noise floor subtraction; for rTSE, the first
echo was excluded from the fit to reduce the effect of stimulated echoes. For
comparison, single voxel multi-echo STEAM spectroscopy (HISTO, 5 TEs = 12, 24,
36, 48 and 72 ms, TR = 3 s)7, which allows separate water and fat
relaxometry, was measured in this case. Finally, T2 values from fat-suppressed
rTSE T2 map and HISTO MRS-based acquisition were evaluated in a heterogeneous
group of 75 patients appointed to clinical abdomen examinations. The presence
of streaking artifacts, to which radial acquisitions are prone, was
subjectively evaluated, and the BMI was recorded for each patient. Patients
were divided into 3 groups: normal, overweight and obese. T2 values were
compared, and the effect of BMI on streaking artifacts was evaluated.Results
From the NIST phantom measurement, rTSE T2 values
were capped at 400 ms by the fitting algorithm. There was an excellent
correlation between the rTSE and Cartesian values in the range of T2s ≈ 22 –
260 ms, but rTSE gave systematically higher values (Figure 1).
In the phantoms containing fat, the application
of FS in rTSE decreased the T2 values in samples with FF over 10% (Figure 2(A)).
From the HISTO results, the T2s of fat are longer than water T2s, and with
increasing FF the difference is also increasing (Figure 2(B)). Moreover,
correlation between rTSE T2s and MRS-T2s of water was improved with the
application of FS (r=0.722, P=0.043 without, r=0.849, P=0.008 with FS).
In the patient group, there were 42 women (age
= 58.9 ± 18.2y) and 34 men (age = 63.2 ± 11.4y). rTSE T2 values ranged from 23.8
to 61.9 ms (mean = 39.9 ms). HISTO T2 water values ranged from 18.6 to 45.9 ms (mean
= 28.2 ms). There was a significant correlation between rTSE and HISTO T2
values (r = 0.7332, P < 0.0001, Figure 3), but there is a systematic
difference between the two.
There
were 41 patients with normal BMI, 25 overweight and 10 obese patients. All the
T2 maps were fit to be evaluated (representative T2 maps without and with
streaking artifacts present are shown in Figure 4). The streaking artifacts
score for each group is given in Figure 5. There was a significant difference
between normal and overweight and between normal and obese (Games-Howell test,
P = 0.002 and P = 0.02, respectively), but no difference between overweight and
obese patients (P = 0.756, Figure 5).
Discussion
There
are systematic differences in phantom T2 values between rTSE and Cartesian; potential
reasons include differences in sequence parameters and the parameter fitting.
Going forward, advanced algorithms should be considered8. The
discrepancy between rTSE and HISTO in our results is in agreement with a
previously reported study that compared multiple single-spin-echo imaging with HISTO9.
There, a non-unity slope (0.548) and an intercept of 16.29 was reported from regression
analysis. In our study the slope is similar (0.462) and the intercept is 9.767
which reflects the fact that rTSE yields higher values. The presence of
streaking artefacts, which increases with increasing BMI, results in limited
measurable ROI areas. In addition, as indicated by the phantom results, fat
signal influences the T2 values, therefore fat suppression appears necessary. A
limitation of this study is the missing direct comparison to a reference
standard. Conclusion
rTSE
T2 mapping is practical for in-vivo application, making it a promising tool for
characterizing diffuse liver disorders.Acknowledgements
This study was funded by the Christian
Doppler Laboratory for Clinical Molecular MR Imaging (Austrian Federal Ministry
for Digital and Economic Affairs and the National Foundation for Research,
Technology and Development).
Word of thanks belongs to Yutaka Natsuaki, who is the original developer of the rTSE T2 mapping prototype.
References
1. Zhan, Chenyang, et al. Detection of hepatic steatosis and iron content at 3 Tesla: comparison of two-point Dixon, quantitative multi-echo Dixon, and MR spectroscopy. Abdominal Radiology, 2019, 44.9: 3040-3048.
2. Dekkers, Ilona A.; LAMB, Hildo J. Clinical application and technical considerations of T 1 & T 2 (*) mapping in cardiac, liver, and renal imaging. The British Journal of Radiology, 2018, 91.1092: 20170825.
3. Banerjee, Rajarshi, et al. Multiparametric magnetic resonance for the non-invasive diagnosis of liver disease. Journal of hepatology, 2014, 60.1: 69-77.
4. Altbach, Maria I., et al. Radial fast spin‐echo method for T2‐weighted imaging and T2 mapping of the liver. Journal of Magnetic Resonance Imaging: An Official Journal of the International Society for Magnetic Resonance in Medicine, 2002, 16.2: 179-189.
5. Natsuaki, Yutaka, et al. Flexible and Efficient 2D Radial TSE T2 Mapping with Tiered Echo Sharing and with “Pseudo” Golden Angle Ratio Reordering. In: Proceedings of the 25th ISMRM annual meeting, Honolulu, HI. 2017.
6. NIST phantom: https://collaborate.nist.gov/mriphantoms/bin/view/MriPhantoms/MRISystemPhantom. ISMRM/NIST MRI System Phantom. Accessed October 5, 2019.
7. Pineda, Nashiely, et al. Measurement of hepatic lipid: high-speed T2-corrected multiecho acquisition at 1H MR spectroscopy—a rapid and accurate technique. Radiology, 2009, 252.2: 568-576.
8. Huang, Chuan, et al. T2 relaxometry with indirect echo compensation from highly undersampled data. Magnetic resonance in medicine, 2013, 70.4: 1026-1037.
9. Lin, Huimin, et al. Quantitative analysis of hepatic iron in patients suspected of coexisting iron overload and steatosis using multi‐echo single‐voxel magnetic resonance spectroscopy: Comparison with fat‐saturated multi‐echo gradient echo sequence. Journal of Magnetic Resonance Imaging, 2018, 48.1: 205-213.