Natalie Bnaiahu1, Ella Wilczynski1, Shir Levy2, Noam Omer1, Tamar Blumenfeld-Katzir1, and Noam Ben-Eliezer1,3,4
1Biomedical Engineering, Tel Aviv University, Tel Aviv, Israel, 2School of Chemistry, Tel Aviv University, Tel Aviv, Israel, 3Sagol school of Neuroscience, Tel Aviv University, Tel Aviv, Israel, 4Center for Advanced Imaging Innovation and Research, New York, NY, United States
Synopsis
On high-field scanners, strong imaging
gradients induce spurious diffusion signal decay, which distorts qT2
calculation. By modeling the pulse
sequence gradients and coherence pathways, an effective b value was calculated
to assess signal attenuation caused
by diffusion. ADC values were measured, extensive phantom scans and
in-vivo experiments were performed using the MSME protocol with varied voxel
sizes to examine the effect and the solution. Before correction T2
values were short with high variability across different voxel sizes, after correction T2 values increased and became more accurate and consistent.
Tissue properties also influenced the correction as their ADC varies.
Introduction
High-field
preclinical scanners employ considerable imaging gradients to achieve suitable
spatial resolutions. These high gradient values result in spurious diffusion
weighting and attenuation of the signal, particularly in multi spin-echo (MSE)
sequences where the effect accumulates along the echo train. Thus, accurate T2
mapping involves tackling both diffusion-related underestimation of T2 values,
and the acknowledged problem of stimulated and indirect echoes, which distort
MSE signals. In addition, different scan parameters, especially those relating
to voxel size, will produce different diffusion and stimulated-echoes signal
bias, leading to highly inconsistent results. Previously we introduced a
preliminary, formula-based design, for correcting diffusion bias, while
employing the EMC algorithm1 to simulate the effect of
stimulated echoes on the T2 decay curve.
In this work, we
present a significant improvement to the diffusion correction, which is now
applied to each of the coherence pathways which make up the signal. This
improvement offers higher accuracy and reproducibility in T2 values,
even for long T2 values and extremely small voxel sizes.Methods
An effective b-value2,3 was calculated
for MSE signals based on the pulse sequence scheme and its specific imaging
gradients. b-value was evaluated per echo based on the subset of coherence pathways
that contributed to the signal4, thus incorporating the effects of
stimulated echoes in the assessment of diffusion attenuation.
$$b-value=γ^{2}\int_{0}^{t}\left(\begin{array}{c}\int_{0}^{t'} g_{x,y,z}^{*}t''dt''\end{array}\right)^{2}dt'$$
A phantom, containing 7 concentrations of MnCl2 ranging from
0.02 to 0.5 mM, was scanned on 9.4T Bruker Biospin. Spectroscopic (Hahn)
spin-echo was performed to achieve ground truth T2 values. Conventional
single SE (SSE) and MSE sequences were used to image the phantom with varying parameters, especially
in-plain resolution and slice thickness. Diffusion coefficient was found to be
close to diffusion coefficient of water at 25°C, 2.29 x 10-5 cm2/s.
SSE T2 values were fitted exponentially and MSE data was fitted using the
EMC algorithm – before and after applying diffusion correction.
An in-vivo scan
of a healthy mouse brain was imaged on a 7T Bruker Biospec using conventional
DTI-EPI and MSME protocols. ADC values along different directions were extracted. A series of MSE scans was performed with different
voxel sizes: 64x64x800, 80x80x800, 100x100x800, 125x125x800, 150x150x800, 125x125x300
and 200x200x300 μm3.
Three brain segments were examined: the cortex, corpus callosum and the
hippocampus. Results
Figure 1 shows the MRS T2 values and the
various MSE averaged T2 values and standard deviations, representing the variance across
scans with different parameters (mainly resolution and slice thickness) per
tube. For example, in the 52 μm pixel scan, the highest
T2 error was 70%, found in the lowest concentration tube. In the physiological
range, T2 errors were as high as 50%. After diffusion correction, T2
values of the MSE and MRS were in agreement with maximal error of 4%.
In
vivo results, before diffusion correction, varied in accordance with the chosen resolution / slice thickness.
Thin slices voxels showed greater signal loss due to diffusion than voxels with
thicker slices, even when the in plain resolution was significantly higher. The
highest deviation, more than 20%, was found in the narrowest voxel and not in the
overall smallest voxel. Figure 5 shows the T2 error (compared to
the uncorrected value) across different voxels.Discussion
Diffusion effect leads to significant underestimation
of T2 values on preclinical scanners. Moreover, this bias is subjected
to the chosen resolution, slice thickness, bandwidth, refocusing RF pulses and
the sample’s diffusion coefficient along different directions. Because the effect of diffusion is accumulative, it increases
with each echo and has higher influence on longer T2 values. Our
results show that decreasing slice thickness leads to greater signal loss than increasing
in-plain resolutions. The error also varies among tissues: narrow slices had
the most significant error in T2 values in the hippocampus, whereas
in voxels with high in-plain resolution the most extensive error was found in
the CC. That is because both tissue's composition and geometry influence ADC
values, in addition to gradients amplitudes, different tissues would undergo different
T2 corrections. Conclusion
Correction
is necessary for high-field / high-resolution qMRI as imaging gradients
intensify the influence of diffusion, which attenuate the signals and lead to miscalculations
of qT2. Effective b-value was modeled
and calculated based on the MSME pulse sequence and refocusing RFs’ coherence
pathways. This method is
generalized and can be applied to different pulse sequences. The
suggested solution improves accuracy, reduces the variability across scan
settings, and provides reproducible T2 values. Acknowledgements
ISF Grant 2009/17References
1. Ben-Eliezer N, Sodickson DK, Block
KT. Rapid and accurate T2 mapping from multi-spin-echo data using
bloch-simulation-based reconstruction. Magn Reson Med. 2015;73(2):809-817.
2. Abragam A. The Principles of
Nuclear Magnetism. Oxford: Oxford Press; 1961.
3. Torrey HC. Bloch Equation with
Diffusion Terms. T Rev. 1956;104(3):563-565.
4. Hennig J. Multiecho imaging
sequences with low refocusing flip angles. J Magn Reson. 1988;78(3):397-407.