Jie Wen1, Dmitriy A Yablonskiy2, Alexander L Sukstanskii2, Feiyan Zeng1, and Ying Liu1
1Radiology, The First Affiliated Hospital of USTC, Hefei, China, 2Radiology, Washington University in St. Louis, St. Louis, MO, United States
Synopsis
Macroscopic B0 magnetic field inhomogeneity adversely affects Gradient-Recalled-Echo (GRE) MRI images. A previously introduced voxel spread function (VSF) method has been proposed to correct for these adverse effects if data are acquired using multi-GRE (mGRE) sequences. The method accounts for both, through-slice and in-slice field inhomogeneities and accounts for between-voxel signal propagation effects. The direct voxel-wise VSF calculation requires long computation time. We propose a library-driven approach for the VSF calculation that significantly reduces computation time, thus providing a platform for the applications of mGRE-based quantitative techniques in clinical practices.
Introduction
It
is well-known that macroscopic B0 magnetic field inhomogeneities adversely
affect quantitative measurements of relaxation parameters obtained with multi-Gradient-Recalled-Echo
(mGRE) MRI 1-4
if not accounted for. It was demonstrated recently that a voxel spread function
(VSF) method of correcting field inhomogeneities 4
incorporated in a quantitative mGRE-based 5 approach
provides quantitative information on brain tissue cellular structure 6
and its alterations under healthy aging 7, and different neurological diseases 8-11. VSF method takes advantage of utilizing both magnitude and phase of mGRE data
from a single scan, hence it saves acquisition time and avoids misregistration errors.
In this study, we propose a library-driven approach for VSF calculation that
significantly decreases computation time as compared with direct calculation. We
also show that image contrasts are improved by incorporating signal decay in
VSF calculation.Methods
As described previously 4, the signal of mGRE-MRI can be expressed as: $$S_{n}\left(TE\right)=\sum_m\psi_{nm}\left(TE\right)\cdot\sigma_m\left(TE\right)\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\left(1\right)$$where $$$S_{n}\left(TE\right)$$$ is the complex image, $$$\sigma_m\left(TE\right)$$$ is the “ideal” signal from the mth voxel. $$$\psi_{nm}\left(TE\right)$$$ is a voxel-dependent convolution kernel, which in 3D case:$$\psi_{nm}\left(TE\right)=\exp\left(i\omega_m\cdot TE+i\phi_0\right)\cdot\eta_x\cdot\eta_y\cdot\eta_z\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\left(2\right)$$where voxel indices n and m are considered as 3D vectors: $$$\left(n_x, n_y,n_z\right)$$$ and $$$\left(m_x, m_y,m_z\right)$$$, $$$\omega_m$$$ represents the frequency shifts. $$$\phi_0$$$ is the signal phase at TE=0. The function $$$\eta$$$ for each dimension is:$$\eta=\sum_qsinc\left(q-q_m\left(TE\right)\right)\cdot\exp\left(2\pi iq\left(n-m\right)\right)\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\left(3\right)$$where q is the 1D normalized k-space defined in the range [-0.5, 0.5]. Parameter $$$q_m$$$ represents the phase dispersion across a voxel:$$q_m\left(TE\right)=\frac{1}{2\pi}\left(g_m\cdot TE+G_m\right)\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\left(4\right)$$where $$$g_m\left(j\right)=\frac{1}{2}\left[\omega_m\left(j+1\right)-\omega_m\left(j-1\right)\right]$$$ and $$$G_m\left(j\right)=\frac{1}{2}\left[\phi_{0m}\left(j+1\right)-\phi_{0m}\left(j-1\right)\right]$$$ are the gradients for the jth voxel calculated from the spatially unwrapped $$$\omega$$$ and $$$\phi_0$$$, respectively. To solve Eq. (1), similarity approximation was introduced previously 4:$$\sigma_m\left(TE\right)\Rightarrow\sigma_n\left(TE\right)\frac{|S_m\left(0\right)|}{|S_n\left(0\right)|}\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\left(5\right)$$In this case, Eq. (1) becomes:$$S_n\left(TE\right)=\sigma_n\left(TE\right)\cdot F_n\left(TE\right)\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\left(6\right)$$and the F-function, which describes the influence of B0 field inhomogeneity on the image:$$F_n\left(TE\right)=\frac{1}{S_n\left(0\right)}\sum_m{\psi_{nm}\left(TE\right)\cdot |S_m\left(0\right)|}\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\left(7\right)$$Since the summation in Eq. (7) extends over many neighboring voxels 4, the direct calculation of the F-function for all the imaging voxels requires significant computation time. Herein we propose to use a pre-calculated library for function η in Eq. (3) that is subsequently used in Eq. (7).We also propose to further improve a similarity approximation in Eq. (5) by incorporating into it the signal decay rate R2*:$$\sigma_m\left(TE\right)\Rightarrow\sigma_n\left(TE\right)\frac{|S_m\left(0\right)|\cdot\exp\left(-R2_m^*\cdot TE\right)}{|S_n\left(0\right)|\cdot\exp\left(-R2_n^*\cdot TE\right)}\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\space\left(8\right)$$The main steps of our approach are the following:
(1) Calculation of the 1st-run F-function (F1) using Eq. (5).
(2) Mono-exponential fitting of the complex signal taking into account the function F1 to get $$$|S\left(0\right)|$$$ and the relaxation rate constant R2*.
(3) Calculation of the 2nd-run F-function (F2) using Eq. (8).
All studies were approved by the local IRB. Images were collected from one healthy volunteer (25 years old, male) using a 3T MRI scanner (GE Discovery 750w) equipped with a 12-channel phased-array head coil. High resolution datasets with a voxel size of 1×1×2 mm3 were acquired using a 3D mGRE sequence (ESWAN on GE scanners) with a flip angle of 20° and TR = 53.2 ms. For each acquisition, 10 echoes were collected with first echo time TE1 = 3.3 ms and echo spacing ΔTE = 3.3 ms. The total acquisition time was around 11 min. Complex images from different channels were combined using a previously introduced method 12. The combined mGRE datasets were then used for calculation of the VSF and R2*. All computations were done in Matlab (2019a, The MathWorks, Inc.) on a computer with standard configurations: 3.2 GHz CPU and 16 GB RAM.Results
Comparisons
of F-functions calculated using direct
and new procedure are shown in Fig 1. The direct calculations of F1 and F2 require 5 hours
20 minutes while the new method requires only 3 minutes. Fig 2
shows the comparisons of fitted R2* maps using no F-funtion, F1 and F2. It
clearly shows that F-function helped with reducing artifacts on R2* maps, as
indicated by red arrows in Fig 2.
By using a two-step procedure, the sharpness of R2* is visually enhanced. Fig 3
compares R2* maps calculated using F-functions with different neighbors (4 and
16 neighbors). When F1 was used, the differences between R2* maps are
relatively large. When F2 was used, the differences were substantially reduced.Discussion
Quantitative
parameters that are evaluated using mGRE-based techniques are usually affected
by magnetic field inhomogeneity effects. These effects cause bias and often
corrupt the quantitative measurements, thus hinder the clinical application of
these techniques. A VSF method 4
allows correction for these effects. In this study, we presented a library-driven
approach for fast VSF calculation. By using a pre-calculated library, the
F-function calculation time has been reduced from hours (when direct
calculation is used) to a few minutes. Additionally, we also incorporated a
two-step iterative procedure for VSF calculation that increases the contrast
(sharpness) of R2* maps. Our results also showed that by using two-step
iterative procedure, the influence of number of neighboring voxels used in
calculating F-function for non-hanning filtered data becomes less important (Fig 3).
One can always use 4 neighbors to calculate F-function, which helps to further
reduce computing time when non-hanning filtered datasets are used.Conclusion
We
presented a library-driven approach for fast two-step iterative VSF
calculation. The new procedure provides a fast way to correct for background
field inhomogeneity artifacts, thus can facilitate the applications of mGRE-based
quantitative techniques in clinical practices.Acknowledgements
This
work was supported by “the Fundamental Research Funds for the Central
Universities” and “the Research Start-up Funds” from the
First Affiliated Hospital of USTC.References
1. Yablonskiy, D.A., Quantitation of intrinsic magnetic susceptibility-related effects in a tissue matrix. Phantom study. 1998. 39(3): p. 417-428.
2. Fernández-Seara, M.A. and F.W. Wehrli, Postprocessing technique to correct for background gradients in image-based R*2 measurements. 2000. 44(3): p. 358-366.
3. Hernando, D., K.K. Vigen, A. Shimakawa, and S.B. Reeder, R*(2) mapping in the presence of macroscopic B(0) field variations. Magn Reson Med, 2012. 68(3): p. 830-40.
4. Yablonskiy, D.A., A.L. Sukstanskii, J. Luo, and X. Wang, Voxel spread function method for correction of magnetic field inhomogeneity effects in quantitative gradient-echo-based MRI. 2013. 70(5): p. 1283-1292.
5. Ulrich, X. and D.A. Yablonskiy, Separation of cellular and BOLD contributions to T2* signal relaxation. Magn Reson Med, 2016. 75(2): p. 606-15.
6. Wen, J., M.S. Goyal, S.V. Astafiev, M.E. Raichle, and D.A. Yablonskiy, Genetically defined cellular correlates of the baseline brain MRI signal. Proc Natl Acad Sci U S A, 2018. 115(41): p. E9727-E9736.
7. Zhao, Y., J. Wen, A.H. Cross, and D.A. Yablonskiy, On the relationship between cellular and hemodynamic properties of the human brain cortex throughout adult lifespan. Neuroimage, 2016. 133: p. 417-429.
8. Wen, J., D.A. Yablonskiy, A. Salter, and A.H. Cross, Limbic system damage in MS: MRI assessment and correlations with clinical testing. PLoS One, 2017. 12(11): p. e0187915.
9. Xiang, B., J. Wen, A.H. Cross, and D.A. Yablonskiy, Single scan quantitative gradient recalled echo MRI for evaluation of tissue damage in lesions and normal appearing gray and white matter in multiple sclerosis. J Magn Reson Imaging, 2019. 49(2): p. 487-498.
10. Zhao, Y., M.E. Raichle, J. Wen, T.L. Benzinger, A.M. Fagan, J. Hassenstab, A.G. Vlassenko, J. Luo, N.J. Cairns, J.J. Christensen, J.C. Morris, and D.A. Yablonskiy, In vivo detection of microstructural correlates of brain pathology in preclinical and early Alzheimer Disease with magnetic resonance imaging. Neuroimage, 2017. 148: p. 296-304.
11. Astafiev, S.V., J. Wen, D.L. Brody, A.H. Cross, A.P. Anokhin, K.L. Zinn, M. Corbetta, and D.A. Yablonskiy, A Novel Gradient Echo Plural Contrast Imaging Method Detects Brain Tissue Abnormalities in Patients With TBI Without Evident Anatomical Changes on Clinical MRI: A Pilot Study. Mil Med, 2019. 184(Suppl 1): p. 218-227.
12. Luo, J., B.D. Jagadeesan, A.H. Cross, and D.A. Yablonskiy, Gradient echo plural contrast imaging--signal model and derived contrasts: T2*, T1, phase, SWI, T1f, FST2*and T2*-SWI. Neuroimage, 2012. 60(2): p. 1073-82.