Naveen Bajaj1, Jaladhar Neelavalli1, Rupesh VK1, Karthik Gopalakrishnan1, and Marc Van Cauteren2
1Philips Healthcare, Bangalore, India, 2Philips Healthcare, Tokyo, Japan
Synopsis
A novel approach for mitigating artifacts from localized motion of the
eye, during interleaved multi-shot diffusion weighted imaging, is presented. A 2D cylindrical spatial
saturation pulse is used to suppress signal from the orbit region, thereby significantly suppressing the ghost artifacts
from eye movement in low b-value images. The saturation being localized to the orbit region does not
influence the surrounding tissues. Placement of such a cylindrical pulse on the orbit region is relatively straight forward. In
addition, the total scan time is not significantly affected by the cylindrical
pulse saturation.
Introduction
Diffusion weighted imaging (DWI) is one of the primary techniques used
in MRI for evaluating tissue damage in-vivo. Single-shot echo-planar imaging
(SSH EPI) methods are typically used for acquiring diffusion weighted images, due to short
acquisition times. SSH EPI methods, however, suffer from geometric distortion,
adversely affecting their diagnostic quality. For distortion, while shorter
readout times help in reducing distortion in EPI, this in-turn limits the
spatial resolution that can be achieved. To overcome this, multi-shot imaging
techniques were introduced which allow collection of the required k-space in multiple
segments, with each segment having a short readout time. However, with data
acquisition in multiple shots, data consistency from shot-to-shot becomes
an important parameter to address[1], for obtaining artifact free images.
Navigator based methods[2] or self navigated methods[3] have
been used to correct for shot-to-shot variations arising from bulk patient
motion. However, effects of involuntary/voluntary motion of local tissue
between the shots, like that of the eye, can lead to anatomy-specific data
inconsistencies, leading to ghost artifacts in interleaved multi-shot techniques. This is especially true in low
diffusion weighting (low b-value) images where there is sufficient signal from the
vitreous humor of the eye. Data consistency based reconstruction approaches may
help in mitigating artifacts related to localized tissue motion[4]. However,
this can lead to SNR penalty. Conversely, spatial saturation of the signal from the eye could help in mitigating ghosts
from the eye[5]. However, due to the rectilinear nature of the spatial
1D saturation bands, it could suppress signal from other regions due to the
curvature of the brain. Hence, a more localized spatial saturation of the orbit
region, using a two dimensional (2D) cylindrical spatial saturation pulse[6], was evaluated for
artifact mitigation from eye-motion in an interleaved multi-shot DWI technique.Materials and Methods
Two
adult healthy volunteers underwent neuroimaging on a Philips 1.5T Prodiva
system after obtaining an informed consent. The volunteers were instructed to
move their eyes intermittently during the scan session. Data was acquired using
a navigator based interleaved multi-shot DWI sequence without and with cylindrical spatial
saturation of the eye-balls. Figure 1 shows the scheme of 2D cylindrical spatial
saturation method. The cylindrical spatial saturation is based on a 2D spiral radio-frequency pulse with gradients in both frequency
and phase directions and a spoiler gradient in the slice direction that enables
a cylindrical saturation along the right-left anatomical direction. The cylindrical spatial saturation pulse is played between the fat suppression pre-pulse and the excitation of each shot. Following
imaging parameters were used: TR/ TRwithSaturation/TE – 4063/4787/105
ms, 2 shot, 1.5 x 1.5 x 5 mm3, b=0 and 1000, 22 slices, scan-time -1:38
min, scan-timewithSaturation -1:55 min. A 2D cylindrical spatial
saturation pulse (9.2ms duration) with diameter of 25mm and flip angle 90o
was used in the orbit region. Figure 2 shows the placement of the
cylindrical pulse on the eye-balls. Image quality was assessed for artifacts in
both acquisitions.Results
Ghost
artifacts from eye-movement were clearly visualized in multi-shot DWI without
cylindrical spatial saturation of the eyeballs. The ghosts were significantly suppressed
using 2D cylindrical spatial saturation (Figure 3).Discussion
In this work we found that a 2D cylindrical spatial saturation pulse successfully mitigates the artifacts from eye-movement in interleaved multi-shot diffusion weighted imaging of brain. The 2D cylindrical pulse does not cause any significant impact on the scan time. Since the saturation is localized to the orbit region, this solution does not influence the other tissue regions, thereby the image quality remains unchanged relative to a no-motion case. Placement of such a 2D cylindrical saturation pulse is relatively straight forward compared to the spatial 1D saturation bands. Even though the profile of the cylindrical spatial saturation is susceptible to magnetic field inhomogeneity, the pulse does not seem to negatively affect the image quality. The usage of the 2D cylindrical spatial saturation pulse for interleaved multi-shot diffusion may be extended to other anatomies like suppression of cardiac pulsation in breast diffusion or urinary bladder motion in prostate diffusion.Acknowledgements
No acknowledgement found.References
[1]
Wu W, Miller KL. Image formation in diffusion MRI: A review
of recent technical developments. J Magn Reson Imaging. 2017 Sep;46(3):646-662.
[2]
Jeong HK, Gore JC, Anderson AW. High-resolution human
diffusion tensor imaging using 2-D navigated multishot SENSE EPI at 7 T. Magn
Reson Med. 2013 Mar1;69(3):793-802.
[3] Chen
NK, Guidon A, Chang HC, Song AW. A robust multi-shot scan strategy for high resolution
diffusion weighted MRI enabled by multiplexed sensitivity-encoding (MUSE).
Neuroimage. 2013 May 15;72:41-7.
[4] Winkelmann
R, Börnert P, Dössel O. Ghost artifact removal using a parallel imaging
approach. Magn Reson Med. 2005 Oct;54(4):1002-9.
[5]
Chou MC, Wang CY, Liu HS, Chung HW, Chen CY. Pseudolesions
arising from unfolding artifacts in diffusion imaging with use of parallel
acquisition: origin and remedies. AJNR Am J Neuroradiol. 2007
Jun-Jul;28(6):1099-101.
[6] Yoneyama
M, Nakamura M, Ozawa Y, Obara M, Okuaki T, Tatsuno S, Sashi R, Sawano S, Van
Cauteren M. Cylindrical Inversion Pulse for the Reduction of Cardiac Motion
Artifacts in Contrast-enhanced Breast MRI. Magn Reson Med Sci. 2018 Jan10;17(1):80-85.