Bilal Tasdelen1,2, Alireza Sadeghi-Tarakameh1,2, Ugur Yilmaz2, and Ergin Atalar1,2
1Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey, 2National Magnetic Resonance Research Center (UMRAM), Ankara, Turkey
Synopsis
In simultaneous
transmission and reception (STAR), injection of noise and spurs stemming from transmit
chain is a problem as well as the leak power. Investigation of noise and spurs
are important in order to determine minimum necessary isolation to achieve
similar noise levels compared to conventional imaging. Noise sources during
STAR are investigated and compared with experiments. It has been shown that
using an active cancellation method where dominant noise source is transmit
chain, with combination of passive cancellation can reduce noise and spurs.
Necessary isolation is shown to depend on input noise and total gain in
transmit chain.
Introduction
In standard MRI,
excitation and acquisition is interleaved which eliminates necessity of
electrical isolation between receive and transmit coil. Ability of simultaneous
transmission and reception (STAR) on the other hand, requires high
isolation (>100 dB) due to small MR
signal amplitude, but it has several benefits such as signal acquisition from
short T2 materials and less RF power consumption1. However, this
method introduces additional considerations that need to be taken care of,
mostly originating from transmit interference to receive chain, such as dynamic
range of receive chain, injected noise from transmit chain to receive chain and
bandwidth of decoupling. Many works have been done to tackle some of these
issues, particularly focused around active and passive cancellation of
interference2,3. Although it is possible to reach isolation
values more than 85 dB, this metric alone may not be the only
factor that affects image quality. Previous studies so far have not
investigated how much isolation is enough and what other parameters are of
importance in order to get similar image quality and SNR as in conventional
imaging. In this work, these considerations are analyzed with a high
emphasis on noise, to determine necessary hardware specifications and
parameters for a simultaneous transmit and receive enabled MRI scanner.Methods
System is modeled in terms of gain and noise, considering each part separately as drawn in Figure 1. 10 ms RF pulses with known amplitudes are applied and output is measured with scanner when different combinations of components are connected, in order to determine gain and noise contribution coming from each module. To demonstrate effect of transmit amplifier parameters on the noise, three different setups are used. First setup referenced as PA1, is a custom PLNA (GRF5020), second setup PA2 is the cascade of these two amplifiers and last setup PA3 is the system’s own amplifier(Analogic AN8135). Measured gains and noise contributions of these amplifiers, as well as the other components are given in Figure 2.
Thermal noise is measured by setting RF input as 0 V and
computing the Power Spectral Density (PSD) of the output. Measurements are
compared with theoretical expectations calculated from the formula $$$P_{noise} (dBm/Hz) = -174 + G_{sys} + NF_{sys}$$$ . Output referred noise power ($$$Na$$$)
of the each device is also calculated and used in power calculations. In addition to thermal noise, phase
noise, non-linear distortions and spurs coming from transmit are also
investigated4,5,6.
Signal/noise power is investigated with passive decoupling only and with the combination of passive and active decoupling. For active
decoupling, recently proposed method7 is used. Sampling rate is 200 kHz and total acquisition time is 1.7 seconds. Receive
chain noise figure is neglected since it is known to be small (~1 dB) and LNA
noise figure will be the determining factor.Results
A summary of experiments and associated thermal noise measurements with PA1 setup can be seen in Figure. 3. Minimum achievable noise
power is calculated as -144 dBm/Hz, which is in agreement with experiment result in
Figure 3.1. With the combination
of active and passive cancellation, noise floor can be reduced down to the
minimum achievable case.
In Figure 4, effect of cancellation can be seen when RF power is present. It is clear from Figure 4.c., bandwidth of the cancellation plays an important role in determining imaging bandwidth in which
there is no significant SNR loss. Bandwidth of the active cancellation and
an image acquired within feasible bandwidth is given in Figure 5.Discussion
Experiments show that when proposed active and passive cancellation combined, not only noise level is reduced, also non-linearities and spurs are eliminated, as long as the noise source is transmit chain. From these results, to determine how much decoupling is adequate, important parameter is the transmit gain, along with imaging bandwidth. To use a higher gain and noise figure amplifier, one must either increase isolation, or decrease imaging bandwidth, which has a similar effect due to band-stop filter nature of the active cancellation.
Decrease in thermal
noise level was not on par with decrease in leak signal, phase noise and spurs.
One reason can be an external noise source stems from the transmit chain picked up
by coil. Additional experiment suggests the presence of such a noise source,
however investigation and elimination of the source requires further
experiments.
It is possible to use
commercial amplifiers bundled with MRI scanners for STAR, given that necessary
isolation is achieved in the imaging band, to preserve SNR. For most cases,
gain of amplifier is the dominant
factor.
Considering most commercial amplifiers have
more than 60dB gain, bandwidth
limitation decreases available
imaging bandwidth well below 70 kHz for our method. In this extent, increasing isolation
bandwidth can be the next improvement.Conclusion
It is shown that there is a
trade-off between these four parameters; transmit amplifier gain, total
isolation, isolation bandwidth and imaging bandwidth, to get the same SNR. As
long as the transmit noise is eliminated in imaging band, remaining leak signal
can be removed with digital subtraction. This trade-off holds for active cancellation, as long as the circuit itself does not
generate significant noise
and there is no independent noise source picked up by
receiver.Acknowledgements
No acknowledgement found.References
1.
D. J. Tyler, M. D. Robson, R. M. Henkelmann, I. R. Young, and G. M. Bydder,
“Magnetic resonance imaging with ultrashort TE (UTE) PULSE sequences: Technical
considerations,” J. Magn. Reson. Imaging, vol. 25, no. 2, pp. 279-289,
Feb. 2007.
2.
A. C. Özen, E. Atalar, J. G. Korvink, and M. Bock, “In vivo MRI with Concurrent
Excitation and Acquisition using Automated Active Analog Cancellation,” p. 3.
3. S.-M. Sohn, J. T. Vaughan, R. L. Lagore, M. Garwood, and D.
Idiyatullin, “In vivo MR imaging with simultaneous RF transmission and
reception,” Magn Reson Med, vol. 76,
no. 6, pp. 1932–1938, 2016.
4. C.
D. Motchenbacher and J. A. Connelly, Low-Noise Electronic System Design. 1993.
5. J.
Phillips and K. Kundert, “Noise in mixers, oscillators, samplers, and logic an
introduction to cyclostationary noise,” in Proceedings of the IEEE 2000 Custom
Integrated Circuits Conference (Cat. No.00CH37044), Orlando, FL, USA, 2000, pp.
431–438.
6. M. J. Flanagan and G. A.
Zirnmerman, “Spur-Reduced Digital Sinusoid Synthesis,” p. 14, Nov. 1993.
7. Tasdelen, Bilal, et al. "Dynamic Decoupling for
Simultaneous Transmission and Acquisition in MRI" Proc. Intl. Soc. Mag.
Reson. Med. Vol. 27, 2019.
8. D. Idiyatullin, S. Suddarth, C. A. Corum, G. Adriany, and M. Garwood, “Continuous SWIFT,” Journal of Magnetic Resonance, vol. 220, pp. 26–31, Jul. 2012.