Chen Lin1, Stephanie Tensfeldt1, and Christopher Serago2
1Radiology, Mayo Clinic, Jacksonville, FL, United States, 2Radiation Oncology, Mayo Clinic, Jacksonville, FL, United States
Synopsis
To demonstrate the distortion
measurements based on automatic analysis of American College of Radiology (ACR)
weekly quality assurance (QA) scans is adequate for monitoring the geometric
distortion of a MR scanner for radiation therapy planning by comparing with
distortion measurements using a 3D grid phantom.
INTRODUCTION
Monitoring geometric distortion in MRI images used for radiation therapy planning and
other stereo-tactic procedures is critical in order to ensure the accuracy of
treatment delivery. Such quality assurance is typically done by scanning a
3D grid phantom and analyzing the spatial
displacement of control points from their known positions1. However,
adding a dedicated and time consuming geometric distortion QA procedure to the routine QA is challenging, especially for a large fleet of
MR scanners. On the other hand, ACR weekly QA (ACR QA) is already being
performed for MR scanners accredited by ACR2. If the basic geometric distortion
measurement based on ACR QA is sensitive and correlated with the more
comprehensive geometric distortion based on a 3D grid phantom3, it may serve a
surrogate to monitor the distortion and ensure that it is within the
appropriate limits for radiation therapy.METHODS
To test such hypothesis, distortion measurements
were performed on a Siemens 3T Vida scanner. Increasing amount of geometric
distortion from gradient calibration errors was simulated by systematically
changing the B0 shim in X and Y directions from the initial optimized shim setting. For
each shim setting, geometric distortion was quantified based on ACR QA and using
QUASAR™ GRID3D image distortion analysis system by Modus QA (GRID3D QA).
ACR QA were performed according to the
protocol and imaging parameters specified by ACR for weekly QA using a Siemens
20ch head&neck coil with receiver bandwidth (rBW) of 150 Hz/px.
The acquired ACR phantom images were analyzed automatically using an in-house analysis
software to measure the phantom length in Z and diameters in X and Y .
For GRID3D QA, a 3D fast gradient echo sequence was
used to acquire images of 1.0 mm isotropic resolution with rBW of 170Hz/pixel.
The 3D datasets were analyzed using Modus GRID3D software to provide
a full 3D distortion map and calculate mean displacement, maximum displacement
and percentage of control points with displacement > 1.5mm in X, Y
and Z directions as well as for the combined displacement vector.RESULTS
Figure 1 shows typical axial images of the ACR
phantom and Modus GRID3D
phantom. Automatic analysis of ACR
phantom was successful for all measurements. Likewise, GRID3D
analysis was also successful with 100% of control points found in each case.
Figures 2 – 4 show the measured absolute deviation of ACR phantom diameter in X direction (ACR Dx) of up to 6 mm with the change of shim settings. Also in the plots is the measured X displacement statistics for the same shim setting. The range of change of mean displacement
in X direction (GRID3D Dx[Mean]), maximum displacement in X direction (GRID3D
Dx[Max]) and percentage of control points with displacement > 1.5mm in
X direction (GRID3D Dx[%>1.5mm]) were 3.0mm, 0.8mm and 14%. Little change in
Y diameter of ACR QA and Y displacement in GRID3D QA was observed (not shown).
There are strong correlations between the GRID3D displacement
statistics and deviation of ACR phantom diameter with correlation coefficients of 0.915, 0.847 and 0.787 between GRID3D Dx[Mean], GRID3D Dx[Max], GRID3D Dx[%>1.5mm]
and ACR Dx.
For the first five shim settings where the change of
shim currents is between 0 and 40 a.u. and the distortion was relatively small, ACR Dx showed a gradual increase from
0.0 mm to 2.0 mm while relatively little change was observed in all three GRID3D
displacement statistics. With the further increase of shim current, ACR Dx continued to increase and displacement
statistics started to change as well. Such behavior suggests that the change of
ACR phantom diameter is more sensitive to small distortion than GRID3D displacement
statistics.DISCUSSION
Since the deviation of ACR phantom diameter is
correlated with the displacement statistics measured from a 3D grid phantom and
has high sensitivity to minor distortion, it’s a good indicator of overall
distortion and can be easily and objectively measured if there is an automatic
analysis program. However, given the size of ACR phantom, it may only be used to
monitor the distortion within similar FOV.
Although the same amount shim current change was
introduced in Y direction, it’s surprising that distortion in Y did not change
and remained small. It could be that the amount of gradient in Y was not as large
with same amount of current. But further investigation is needed to confirm.
Future validation will include experiments to demonstrate the
correlation between ACR QA and GRID3D QA when there is substantial amount of
distortion in all three directions.
Currently, a 1.0mm action criterion for the
deviation of ACR phantom length and diameters is used to monitor geometric
distortion in a fleet of 10 MR scanners. The effectiveness of such distortion
QA program is being evaluated by comparing with GRID3D QA quarterly.CONCLUSION
Limiting the ACR phantom diameter deviation to 1.0 mm
would limit the mean displacement within 0.5 mm,
maximum displacement within 2.0 mm and percentage of control points with
displacement > 1.5 mm less than 1% in 3D grid phantom in the same direction.Acknowledgements
References
- Torfeh
T et al. Characterization of 3D
geometric distortion of magnetic resonance imaging scanners commissioned for
radiation therapy planning, Magn Reson Imaging. 2016 Jun;34(5):645-53
- 2015
ACR MRI Quality Control Manual
- Stanescu
T. et al. Investigation of a 3D
system distortion correction method for MR images, Journal of Applied Clinical
Medical Physics, Vol. 11, No. 1, Winter 2010, pp 200-21