Marta Brigid Maggioni1, Martin Krämer1, and Jürgen R. Reichenbach1,2,3,4
1Medical Physics Group, Institute of Diagnostic and Interventional Radiology, Jena University Hospital - Friedrich Schiller University, Jena, Germany, 2Michael Stifel Center for Data-driven and Simulation Science Jena, Friedrich Schiller University, Jena, Germany, 3Abbe School of Photonics, Friedrich Schiller University, Jena, Germany, 4Center of Medical Optics and Photonics, Friedrich Schiller University, Jena, Germany
Synopsis
B1
mapping is a major challenge for reliable T1
quantification, and it is especially difficult for UTE sequences,
because most conventional B1
mapping methods are unable to perform on short T2*
tissues. Recently, an implementation of the actual flip angle imaging
(AFI) method was proposed for UTE sequences. The AFI method requires
ideal spoiling of residual transverse magnetization, which is
typically achieved by fine-tuning RF-spoiling phase increments as
well as the strength of spoiler gradients. In this work, we propose
an improved gradient spoiling scheme, which reduces the effect of the
RF-spoiling phase increment on the AFI results.
Introduction
The uncertainty in the estimation of the flip
angle caused by inhomogeneities of the B1
field is a major source of unreliable T1
quantification when using methods based on flip angle variation
(VFA). Although several methods have been proposed for directly
mapping the B1
distribution1,2, most
of them cannot be applied to tissues with short T2*
relaxation because such tissues require ultrashort echo-time (UTE)
imaging sequences in order to retrieve signal.
Recently, an actual flip angle (AFI) B1
correction method was proposed for UTE sequences3.
The AFI technique uses a sequence with two interleaved
repetition-times (TR). The signal from both TRs is then combined, to generate a B1
map. The AFI method relies on a pulsed steady-state regime, which,
however, has been shown to be highly sensitive to variations of the
RF-spoiling phase increment (φ)
as well as the amplitude and duration of the spoiler gradients4,5.
Various optimal values for φ
have been proposed in literature3,4,5 but they depend on the choice of TRs, and a small variation in φ
can also produce significant errors in B1
estimation4,5,6.
Therefore, very long spoiler gradients with high amplitudes have been
proposed as methods for eliminating the dependence on φ5. In
combination with 3D radial hard-pulse excitation-based UTE sequences
this comes at the cost of significant increases of the acquisition
time. In this work, we propose an optimized gradient spoiling scheme,
which reduces the dependency of UTE AFI based B1
mapping on φ and
allows a reduction of the spoiling gradient duration and thus the
total acquisition time.Methods
A hard-pulse 3D radial UTE sequence7
was extended by adding two adjustable interleaved TRs for AFI (Fig.
1) and further modified to allow for a wide range of φ
values in combination with adjustable durations and amplitudes of the
spoiler gradients. For the gradient spoiling scheme, constant
spoiling in z-direction was always used (after rephasing the
magnetization back to the center of k-space). In
addition, the sequence allowed for random spoiler gradients in the x-
and y-directions. The purpose of the latter was to further “scramble”
the transverse magnetization to achieve, on average, a more pure
steady-state signal.
The sequence was applied to a 12cm diameter
spherical phantom, filled with water and superabsorber polymer for
stabilization. Different TR combinations as well as amplitudes and
durations of the gradients spoilers were chosen (Table 1) and tested
on 180 phase increment values ranging from 1° to 180°,
with all other acquisition parameters kept
identical: flip angle 45°, isotropic resolution (3.5×3.5×3.5)mm3.
All measurements were performed with a
clinical 3T MRI scanner (PRISMA, Siemens) using the vendor supplied
single-channel transmit/receive knee
coil.
The method was further validated to correct the T1
quantification results on a knee of a
healthy 35-year-old male volunteer. The T1
map was calculated by using a
variable flip angle (VFA) approach (FAs:
5°,8°,10°,12°,15°,20°,25°,30°,35°,40°,
TR: 5 ms, (1.4×1.4×3.0)mm³
anisotropic resolution), and corrected by using an AFI sequence with
the same resolution, FA:45°,TR:15 ms
and AG1/AG2
of 65/325, where AG is the spoiler gradient
area expressed in [mT·ms/m].Results
Figure 2 shows the calculated flip angle
correction factor for a spherical region of interest in the centre
of the phantom for the 180 RF-increments φ
with different spoilers and TR
combinations. Even the measurement with the largest spoiler gradient
amplitude and duration, but spoiling only in the z direction (red
curve) still showed a dependency on φ,
that is only increased with the application of smaller gradient
spoilers (blue curve). The introduction of randomized spoiling in the
x and y direction, however, not only is able to completely eliminate
the dependency on φ,
but is also able to achieve comparable correction factors, even with
a reduction of the amplitude of the gradients and the TRs (yellow,
violet and green curves).
Figure 3 shows the T1
VFA map before and after correcting
for B1
inhomogeneities with the AFI correction, the results of T1
quantification in regions of interest
(ROI) in the tendon, bone marrow and muscle are shown in
table 2. Discussion and Conclusion
By adding randomized x and y spoiling to a 3D UTE
AFI sequence we demonstrated that the dependency on the RF-spoiling
phase increment can be highly reduced, potentially making UTE AFI B1
mapping more robust. Investigating why the added randomized x and y
gradient spoiling stabilizes the AFI results would require full Bloch
simulations of the AFI sequence, including the initial phase where
the steady-state is reached. We suspect that the added randomized
spoiling effectively adds a scrambling of the residual transverse
magnetization to the sequence which then reduces potential for
refocusing of old magnetization (that would disturb the steady-state
signal required for AFI). This, in combination with the strong k-space
center averaging in a 3D center-out radial acquisition of the UTE
sequence, ultimately stabilizes the steady-state signal. The
proposed improved gradient spoiling scheme, furthermore, has the
advantage of allowing a reduction in the duration and amplitude of
the spoiler gradients and thus also TRs, ultimately leading to a
reduction of the total acquisition time. Furthermore, the in-vivo
results of the proposed AFI correction lead to T1
results that are in line with
literature8,9.Acknowledgements
No acknowledgement found.References
-
Cunningham
CH, Pauly JM, Nayak KS. Saturated double-angle method for rapid
B1+ mapping. Magn Reson Med 2006;55:1326–1333.
-
Chung S, Kim
D, Breton E, et al,Rapid B1+ mapping using a preconditioning RF
pulse with TurboFLASH readout, Magn
Reson Med.
2010;64(2):439–446.
-
Nehrke K, On
the steady‐state properties of actual flip angle imaging (AFI).
Magn. Reson. Med., 2009; 61: 84-92.
-
Yarnykh V L,
Actual flip‐angle imaging in the pulsed steady state: A method for
rapid three‐dimensional mapping of the transmitted radiofrequency
field. Magn. Reson. Med., 2007; 57: 192-200
-
Yarnykh V L,
Optimal radiofrequency and gradient spoiling for improved accuracy
ofT1and
B1measurements
using fast steady‐state techniques. Magn. Reson. Med., 2010; 63:
1610-1626
-
Ma
Y, Lu X, Michael C, et al. Accurate
T1 mapping of short T2 tissues using a three‐dimensional
ultrashort echo time cones actual flip angle imaging‐variable
repetition time (3D UTE‐Cones AFI‐VTR) method. Magn. Reson. Med,
2018; 80-89.
-
Herrmann KH,
Krämer M, Reichenbach JR Time Efficient 3D Radial UTE Sampling with
Fully Automatic Delay Compensation on a Clinical 3T MR Scanner, PLOS
ONE, 2016, 11(3)
-
avala
Bojorquez J, Bricq S, Acquitter C, et al, What are normal relaxation
times of tissues at 3 T?, Magnetic Resonance Imaging, 2017; 35, 69-80.
-
Ma
Y,
Zhao
W,
Wan
L,
et al.Whole
knee joint T1
values measured in vivo at 3T by
combined 3D
ultrashort echo time cones actual flip angle and variable flip angle
methods.
Magn
Reson Med.,
2019;
81:1634–1644