Michael Nicholas Hoff1, Nathan M Cross2, Qing-San Xiang3, Daniel S Hippe2, Charles G Colip2, and Jalal B Andre2
1Radiology, University of Washington, Seattle, WA, United States, 2Radiology, University of Washington, SEATTLE, WA, United States, 3Radiology, University of British Columbia, Vancouver, BC, Canada
Synopsis
Widespread use of phase-cycled bSSFP imaging is hampered by
problematic sensitivity to artifacts caused by susceptibility and motion. The geometric solution (GS), which can
eliminate susceptibility-related banding and signal modulation, has previously
show relative insensitivity to motion.
Here, GS-bSSFP is evaluated using a phantom and in humans, imaging the
skull base. The GS mitigates motion
artifact in both paradigms and is particularly resilient when one of its four
phase cycles is corrupted by motion.
Introduction
Balanced steady state free precession (bSSFP) imaging boasts
high SNR efficiency but is plagued by dark artifacts. While
susceptibility-induced banding and signal modulation are primary concerns,
other darkening artifacts including those caused by motion can be
problematic. The geometric solution1
(GS) can demodulate bSSFP of banding, and recent simulations and in vivo
work have shown an insensitivity to motion2,3. Here we evaluate the GS performance using a
water-flow phantom and in vivo while imaging the human skull base (including
the internal auditory canal (IAC) and orbits, where aqueous/vitreous humor
flow, ocular motion, cerebrospinal fluid flow, and cisternal vessel pulsation
can occur) to demonstrate that the GS can mitigate motion artifacts. Methods
A plastic bottle was wrapped with ¼” inner diameter plastic
tubing and placed in the MRI scanner. Water was pumped through the tube at
controlled velocity and duration. Figure 1 details the 0°, 90°, 180°, and 270°
phase-cycled (PC) bSSFP images that were acquired 1: without flow, during 1PC
(0°) at velocities 2: 40ml/15s, 3: 40ml/30s, and 4: 40ml/60s, and during all
4PC at velocities 5: 40ml/15s, 6: 40ml/30s, and 7: 40ml/60s. The Complex Sum
(CS) and GS were computed in each of the seven scenarios as previously
discussed1,4.
In order to gauge flow artifact in flow scenario X, residual flow artifact was computed by the absolute value of its complex
difference from the 1: non-flow map, say |GSx – GS1|, as
at the bottom of Fig. 1. A quantitative estimate of the total flow
artifact may then be made via the total energy of each difference map:
$$En=\sum_{all pixels}(GS_X-GS_1)^2$$
En energy was computed for each flow/reconstruction
scenario, normalized to the largest value, and listed on its corresponding
difference map.
An institutional review board approved 21
patients undergoing a clinically indicated temporal bone (IAC protocol) MRI of
the skull base for evaluation of treated vestibular schwannoma to received
added bSSFP sequences acquired with PC = 0°, 90°, 180°, 270° in both
sagittal and axial orientations on Philips Ingenia 3T scanners. Scan parameters were FA/TR/TE = 30°-45°/4.8-5.5ms/1.9-2.2ms,
316/314-316/42-100 matrix size and 0.57/0.57/1mm voxel size along
frequency/phase/slice directions, respectively.
The GS and CS were computed, and signal-normalized to avoid bias. Three fellowship-trained neuroradiologists
blinded to sequence type, imaging parameters, clinical presentation, and
patient disposition evaluated standard bSSFP, GS-bSSFP, and CS-bSSFP for the
degree of dark artifact in the IAC and orbits of each patient using a 5-point
Likert scale (1 = significant to 5 = none).
Prevalence of dark artifact was compared between techniques using the
Wilcoxon rank-sum test or the Sign test, clustered by subject to account for
multiple readers.Results
The bottom of Figure 1 illustrates that water velocities have
minimal effect on energy En, however, 4PC flow durations showed
significant increase in En relative to 1PC durations for both the GS and CS. The GS showed remarkably low energy for 1PC flow,
consistently less than 30% of the corresponding 1PC CS energy. Neither technique performed well given flow
during all phase cycles.
Figure 2 depicts a patient’s orbits in standard bSSFP,
GS-bSSFP, and CS-bSSFP axial images. The
GS exhibits relatively minimal vitreous darkening when compared with the CS and
standard bSSFP, with sharp margins of orbital structures.
Figure 3 depicts a patient’s IAC in a standard bSSFP image, and GS-bSSFP
and CS-bSSFP reconstructions. The GS
displays good contrast, cranial nerve demarcation and slightly better CSF
homogeneity when compared to the other techniques.
Figure 4 shows that the GS was assessed to have slightly less dark
artifact in the IAC than the CS and much less than standard bSSFP, with
p-values 0.001 and <0.001 respectively. The GS had less artifact than the CS
in 24% and the same in 76% of reads, and less artifact than bSSFP in 97% and
the same in 3% of reads; the GS was never rated to have more dark artifact in
the IAC than the CS or standard bSSFP.
Figure 5 depicts that the GS was assessed to have significantly
less dark artifact in the orbits than the CS and standard bSSFP
(p<0.001 for both). The GS had less artifact than the CS in 49% and the same
in 51% of reads, and less artifact than bSSFP in 81%, the same in 16%, and more
in 3% of reads (2/63); the GS was never rated to have more dark artifact in the
orbits than the CS.Discussion
Phantom and in vivo results indicate that the GS shows capacity
for flow artifact mitigation when motion occurs during only one phase cycle, while
motion during all phase cycles reduces the GS performance considerably. This
may be explained by the recent discovery that the GS is insensitive to motion’s
noise-like effects when those effects are oriented in a radial direction in the
complex plane with respect to a spoke in the GS cross-solution3. This logic breaks down when all four PCs are corrupts by motion, although further testing of variable flow duration is
necessary. Conclusion
The GS is a robust MRI technique that not only yields high
SNR efficiency and signal demodulation, but also demonstrates relative insensitivity to motion artifacts in a phantom and in vivo.Acknowledgements
NoneReferences
[1] Xiang QS, Hoff MN. Banding artifact removal for bSSFP imaging
with an elliptical signal model. Magn Reson Med. 2014;71:927–933.
[2] Hoff MN, Wilson GJ, Xiang XS, Andre JB. Artifact Correction in
Temporal Bone Imaging with GS-bSSFP. Proc ISMRM 2014; 22:1627.
[3] Hoff MN, Andre JB, Xiang QS. On the Resilience of GS-bSSFP to
Motion and other Noise-like Artifacts, Proc ISMRM 2015; 23:818.
[4] Hoff MN, Andre JB, Xiang QS. Combined Geometric and Algebraic
Solutions for Removal of bSSFP Banding Artifacts with Performance Comparisons,
Magn Reson Med. 2017;77:644–654.