Laura Bortolotti1 and Richard Bowtell1
1Physics, Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, United Kingdom
Synopsis
A fixed array of NMR field probes
has proved to be a valid tool for monitoring the effects of a wide range of
head movements in a 7T scanner. Here, we use this set up to detect significant
field changes in order to give early feedback on head motion. This information
could be used in deciding to stop a scan or to reacquire all or part of the
k-space data in order to avoid acquisition of motion corrupted image data. This
method can be implemented without requiring image sequence modification nor
rigid coupling of a motion marker to the head.
Introduction
Head movements during MRI cause artefacts in the resulting images. A
variety of Motion Correction (MoCo) techniques have been developed for correcting
these artefacts prospectively or retrospectively [1]. However, if the level of
movement is larger than the effectiveness range of the particular MoCo
technique, the image results will still be corrupted. Under these circumstances,
or when no MoCo is applied, it would be useful to have a system which could
indicate when head movement that is large enough to cause image artefact has
occurred. This would potentially allow a choice to be made to reacquire all, or
part of, the k-space data needed for image production. In this work, the
possibility of using NMR field probes to flag up significant motion has been
explored using measurements of extra-cranial magnetic field perturbations which
were concurrently acquired with optical monitoring of head pose. The proposed
approach allows the detection of significant head movement without requiring
image sequence modification [2] nor rigid coupling of a motion
marker to the volunteer’s head.Methods
A set of 16 NMR field probes (Skope) was positioned around the head in a
customised cylindrical holder [4] formed from four coaxial
rings (Figure 1). Extracranial magnetic field changes produced by a
range of different head movements executed by four subjects were measured using
these field probes at a sampling rate of 6.67 s-1, whilst head pose
was simultaneously monitored using a Moire’ Phase Tracking (MPT) system,
comprising an optical camera and an MPT marker attached to a dental mould. The
detection of significant motion is based on a comparison of the variance of the
field measurements during a short time interval to the average value of the
variance measured with the subject at rest in the scanner. The aim is to identify
a reference level (L) for the field variance at each probe (p) that indicates
an unacceptable level of head movement. Here, the level (L) is based on the
ratio (E) of the measured field variance at each time step ( σ2(Bp(Δt)) ) to the average value of the variance measured in the rest condition (R). Any
probe for which the level is exceeded is classified as indicating an invalid
state. The complete workflow of the method is shown in Figure 3. For
this analysis, Δt was set equal to ten time steps (corresponding to 1.5 s). Figure
4 shows a forecast of which probes are most likely to detect head movements
during the different conditions (“Rest”, “Feet wiggling”, “Head Shake” and
“Head Nod”) based on applying PCA to the variance of the data.Results
The plots in Figure 2 show
the extra-cranial changes in magnetic field for different head movements for one
subject. Larger changes are measured for larger head movements, as expected. Analysis
of the pattern of variation of field variances over the different probes shows
that the range of field variation depends on the probe position and type of
movement.
The
variation of the number of probes classified as invalid with time during the
different movement conditions is shown in Figure 5 for different values
of the threshold level. The concurrent
variation of the RMS displacement and rotation relative to the starting
position (measured using the MPT system) is also shown. These results indicate
that by varying the threshold level it is possible to adjust the level of
movement above which a significant number of probes are classified as invalid.
This would allow the level of motion at which a problem is flagged to be adjusted by
the operator.Discussion and Conclusion
This work has demonstrated that it is possible to develop a motion
detection tool for a 7T MRI scanner based on the use of NMR field probes. This
tool does not require image sequence modification [2] or use of markers that are
directly coupled to the head [4]. For these experiments the field probes were
sited in a bespoke holder, which took the place of the standard receiver coil
array, but in future the probes could be placed between the receiver array and
volume transmit coil. Finding the correct criteria to use in deciding when to stop
a scan automatically or to rescan some k-space data so as to avoid significant
motion artefacts will require further work. Possible criteria are based on adjusting
the level L and evaluating the number of probes that are invalid over a given period
of time. Acknowledgements
No acknowledgement found.References
[1] McLaren
review J. McLaren Prospective motion correction in brain imaging: A review 2013
March; 69, 621-63
[2]
Aranovitch A. et al, “Prospective motion correction with NMR markers using only
native sequence elements”, Magn Reson Med. 2018; 79:2046–2056
[3] Bortolotti
L. et al, “Non-contact measurement of head movements inside a 7 T Scanner using
a 16-channel field camera”, ISMRM2019, Abstract number: 4432.
[4]
Bortolotti L. et al, “Simulation of external magnetic field changes due to head
motion during 7 Tesla MRI scan”, ISMRM2019; Abstract number: 4431.
[5] Bortolotti
L. et al, “Using a field camera to monitor head movement in 7T scanner”, ESMRMB2019,
A-1346
[6]
Andre, Jalal B., et al. "Toward quantifying the prevalence, severity, and
cost associated with patient motion during clinical MR examinations."Journal
of the American College of Radiology” 12.7 (2015): 689-695.
[7] Simon Pascal Gross, Diss. ETH No. 23739,
Zurich, 2016.